5. Maintenance and Sterilization
The Helios Rotary Pen consists of four main components. The cartridge, which is
purchased separately, pre-sterilized and individually packaged, the handpiece, the
motor (contained in the upper portion of the handpiece), and the connecting wire. All of
the components, with the exception of the cartridge, which are intended for single-use,
MUST be cleaned and disinfected both before and after each individual use.
The handpiece and the connecting wire must be wiped down with a soft disinfecting
cloth such as Sani-Cloth, or DEFEND Plus Disinfecting wipes before and after each
use. All components of the machine must be covered with a protective covering or
sleeve during use to reduce the risk of infection.
The handpiece of the Helios Rotary Pen is made up of two parts. By holding both the
upper and lower portion of the machine, and twisting counter-clockwise, the handpiece
may be unscrewed, revealing the two portions of the handpiece. The lower portion may
be DRY HEAT autoclaved for sterilization. You cannot sterilize any part of the Helios
Rotary Pen by STEAM AUTOCLAVE! Please be advised that the aluminum body has
been anodized. Over time, if you continue to autoclave this portion of your machine, it
MAY cause the color to begin to fade. This is completely normal and is NOT covered
under the warranty.