To ensure safe operation, read carefully the entire or assembly, maintenance or specifications of your manual before installing, operating or servicing patient lifter, you should call Joerns Healthcare toll-free this equipment. at 800.826.0270. If anything is not completely understood, please For service and repair, remember your authorized contact your supplier for more details.
Page 3
Warning: Never use a damaged, torn or frayed sling. Caution: If trouble develops or you have questions regarding the hydraulic pump, call Joerns Healthcare Warning: Always place the sling around the user Customer Care at 1-800-826-0270 (be sure to provide according to the sling user instructions.
Joerns Lifting and Repositioning ® C-HLA-2 Lift Table of Contents Specifications Base ............Adjustable U Important Precautions ..........2 Maximum Weight Capacity ......400 lbs. Specifications ............4 Lift Range of Boom ........29" - 80" Assembly..............5 Width of Base ........... 24" - 39.5"...
Joerns Lifting and Repositioning ® C-HLA-2 Lift Assembly Recommended Tools: two 9/16" wrenches, 1/8" hex wrench, 5mm hex wrench Warning: Inside center attachment points on cradle are used only for Hoyer slings with chains or straps. The four outside attachment points are used for Hoyer 4-point slings.
Page 6
Joerns Lifting and Repositioning ® C-HLA-2 Lift 6. Cradle Attachment a. Remove nut and bolt from cradle. b. Hold cradle in position at the end of the boom and align the holes of the pin and cap with the holes in the boom end fork.
Joerns Lifting and Repositioning ® C-HLA-2 Lift Operation Warning: Lifters can tip over. Keep base widened for stability. NEVER lock wheel locks or block wheels when lifting someone. Wheel locks are provided for parking. The only exception is when lifting a patient from the floor - the locks should be engaged during a floor lift.
4. If trouble develops with the hydraulic pump, remove it and call your Joerns Healthcare Supplier to receive an RMA. Provide serial number (located on the base of the unit) and date purchased. Do not attempt to lift someone with a faulty hydraulic pump.
Joerns Lifting and Repositioning ® C-HLA-2 Lift Lifter Maintenance Checklist Only qualified persons should service and repair Hoyer lifters. Regular maintenance of lifters and accessories is necessary to assure proper and safe operation. Read and observe the following recommended instructions.
Joerns Lifting and Repositioning ® C-HLA-2 Lift Lifter Maintenance Checklist continued... Item Initially Before each use Monthly Lubrication Oil all pivot joints, including: Top of mast where boom attaches ✔ Top and bottom of actuator mounting ✔ Pump handle hinge ✔...
Use slings in good condition only. Destroy and discard old, unusable slings. Note: Do not use any other manufacturer’s sling with the C-HLA-2 Lifter. Note: Do not use Hoyer slings with any other manufacturer’s lift. Patient Size Guide...