Do you have a question about the 40L and is the answer not in the manual?
Questions and answers
Vasanth
May 12, 2025
Starting trouble in compressor once started then its okk
1 comments:
Mr. Anderson
May 20, 2025
Possible causes of starting trouble in the GNATUS 40L compressor include:
- Power plug disconnected: Connect the power plug to the outlet. - Main switch off: Turn on the main switch. - Power outage in the electrical network: Wait until power is restored. - Overload protector tripped: Wait 5 minutes for the compressor to cool and then reset the overload protector. If it trips again, contact technical assistance.
PRESENTATION OF MANUAL INSTRUCTIONS FOR USE Technical Name: Air compressor Model: 40, 65 and 150 Liters Brand: Gnatus Manufacturer / Distribuitor: Chiaperini industrial LTDA C.N.P.J. 59.064.766/0001-82 Avenida Professora Luiza Garcia Ribeiro, 130 Santa Rosa de Viterbo – SP – Brasil...
Page 3
INDEX 02 PRESENTATION OF MANUAL 04 IDENTIFICATION OF EQUIPMENT 06 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION 07 TECHNICAL SPECIFICATIONS 09 Dimensions 12 Packing symbols 12 Product symbols 13 INSTALLATION OF EQUIPMENT 14 EQUIPMENT OPERATION 14 Turning on the compressor 15 Compressed air outlet 15 Drainage 16 PRECAUTIONS, RESTRICTIONS AND WARNINGS...
IDENTIFICATION OF EQUIPMENT Dear Customer This manual is a general presentation of your product and it will give you important details to help you to solve possible problems. Please, read it and keep this with you. IDENTIFICATION OF PRODUCT Technical Name: Air compressor Model: 40, 65 and 150 Liters Indication of Equipment This equipment is for dental use use only.
IDENTIFICATION OF EQUIPMENT Principles and fundamentals applied to the product operation All components of the equipment are mounted over the air tank. One or more electric engines trigger the pistons´ system with rings made of non-metallic material resistant to abrasion and high temperatures that are able to operate without lubrication for a long period of time.
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION 01 - Air tank 07 - Safety valve 02 - Rubber feet 08 - Overload protector 03 - Air intake filter 09 - Register of the outlet pressure 04 - Electric engine 10 - Manometer of the outlet pressure 05 - On - off switch 11 - Outflow rate register...
65L 220V~: 6,5 A 150L 220V~: 13 A General power and energy consumption 40L 127V~: 1080 VA / 1,08 KW/h 40L 220V~: 1056 VA / 1,06 KW/h 65L 127V~: 1588 VA / 1,59 KW/h 65L 220V~: 1430 VA / 1,43 KW/h...
TECHNICAL SPECIFICATIONS Packing symbols It determines the maximum Packing to be transported and / quantity of boxes which can be or stored avoiding humidity, rains stacked during transportation and and wet floor storage “as per packaging” The packing must be stored and Packing to be transported and / or transported away from direct sun stored with the harrows up...
Gnatus techician. - This equipment shall only be able to be unpacked and installed by a Gnatus authorized technician, under penalty of losing the warranty, as only (s)he has the information, suitable tools and training required to execute this task.
INSTALLATION OF EQUIPMENT Voltage Rated current Circuit breaker 8,5 A 10 A 127V~ ±10% 220V~ ±10% 4,8 A 10 A Voltage Rated current Circuit breaker 12,5 A 16 A 127V~ ±10% 220V~ ±10% 6,5 A 10 A 150L Voltage Rated current Circuit breaker 13 A 16 A...
OPERATION OF EQUIPMENT Compressed air outlet The air pressure on the outlet can be adjusted through the register (09), check the setting on the pressure gauge (10). The air outflow rate can also be adjusted in the register (12). In tandem with the pressure and flow rate registers, there is a filter that removes moisture and impurities remaining in the air, the moisture that is removed from the outlet will get stored in the reservoir (11), to empty it press the pin underneath the reservoir.
Precaution and warnings “during the installation” of the equipment - This equipment can only be unpacked and installed by an Gnatus authorized technician. - Check that the socket in which the device will be connected has a ground connection. According to the ABNT standard, this is essential for the safe operation of the system.
- Do not allow patients and especially children get close to the compressor. - Gnatus air compressor is designed for use according to its air flow capacity, while using it, does not exceed their technical specifications. Gnatus assumes no liability for damages/ accidents resulting from improper use or non-compliance with the instructions described in this manual.
PRECAUTIONS, RESTRICTIONS AND WARNINGS Gnatus shall not be responsible for: - Use of the equipment differing from that for which it is intended. - Damages caused to the equipment, the professional and/or the patient by the incorrect installation and erroneous procedures of maintenance, differing from those described in these Instructions for use which come with the equipment or by the incorrect operation of it.
The equipment should be calibrated routinely, as per the legislation in force in the country. But never with a period exceeding 3 years. In order to protect your equipment, seek Gnatus technical assistance for periodic revisions of preventive maintenance. Filter Change: Keep the air filter(s) (B) clean and in good condition of usage.
Page 20
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION Corrective Maintenance Gnatus states that the supplying of the circuits’ diagram, Part lists or any other information that permits the technical assistance by the user, can be requested, since previously agreed between the buyer and Gnatus.
Among the care you have to take with your equipment, the most important is regarding of the spare parts replacement. To ensure the lifetime of your device, only replace original spare parts from Gnatus. They have the assurance of the standards and technical specifications required by the Gnatus representative.
Need help?
Do you have a question about the 40L and is the answer not in the manual?
Questions and answers
Starting trouble in compressor once started then its okk
Possible causes of starting trouble in the GNATUS 40L compressor include:
- Power plug disconnected: Connect the power plug to the outlet.
- Main switch off: Turn on the main switch.
- Power outage in the electrical network: Wait until power is restored.
- Overload protector tripped: Wait 5 minutes for the compressor to cool and then reset the overload protector. If it trips again, contact technical assistance.
This answer is automatically generated