DW E.S.T. (Elevated Skin Temperature) System

3. Installation Diagram and Considerations

3.1. Installation Diagram

The diagram below represents recommended installation settings.
1. Install the camera facing away from the entrance or windows.
2. Place the camera 6.5~16.4ft from the objects (10 feet recommended).
3. Place the camera at a 20° angle from the objects.
4. Make sure people are walking into the camera's FoV rather than straight towards the camera.
5. Only one person in the FoV at a time.
6. Place the Blackbody across from the camera, positioned in the corner of the camera's FoV.
7. For optimal readings, the Blackbody should be 9.8~14.7ft from the camera (10 feet
recommended).
8. Recommended room temperature is 68° to 86°F (20° to 30°C).
9. Set up a monitoring station next to the camera with a monitoring PC and personnel. See
'Camera Connection' for more information.
10. Always follow elevated temperature readings with additional screening such as a medical
thermometer and wellness questionnaire.
For DW-ESTS installations, where the camera is mounted with a
Blackbody, allow up to 5 minutes for the camera to power up and
calibrate properly.
For DW-ESTCAM installations, where the camera is mounted without
a Blackbody, allow up to 30 minutes for the camera to power up and
calibrate properly.