Rosemount Analytical NGA 2000 Manual

Rosemount Analytical NGA 2000 Manual

Non-dispersive infrared analyzer module

Advertisement

NGA 2000
N
-D
I
ON
ISPERSIVE
NFRARED
A
M
NALYZER
ODULE
Rosemount Analytical

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NGA 2000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Rosemount Analytical NGA 2000

  • Page 1 NGA 2000 ISPERSIVE NFRARED NALYZER ODULE Rosemount Analytical...
  • Page 2 E.I. duPont de Nemours and Co., Inc. ® Ty-Rap is a registered trademark of Thomas & Betts Corp. Manual Part Number 748332-D June 1997 Printed in U.S.A. Rosemount Analytical Inc. 4125 East La Palma Avenue Anaheim, California 92807-1802...
  • Page 3: Table Of Contents

    TYPICAL APPLICATIONS.................1-1 THEORY OF TECHNOLOGY ..............1-2 SAMPLE REQUIREMENTS ..............1-4 PURGE KITS .....................1-5 FEATURES....................1-5 2. I ECTION NSTALLATION UNPACKING....................2-1 ASSEMBLY....................2-1 LOCATION ....................2-2 GASES ......................2-2 2.4.1 Specifications ................2-2 2.4.2 Connections .................2-3 ELECTRICAL CONNECTIONS ..............2-5 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infrared Analyzer...
  • Page 4 4.11 OSCILLATOR TUNE/SOURCE BALANCE SHUTTER ADJUSTMENT..4-4 4-12 CLEANING CELLS ................... 4-8 4-13 CELL DESICCANT ................... 4-10 4.14 MODULATION CHECK ................4-10 ECTION REPLACEMENT PARTS REPLACEMENT PARTS ................5-1 A. I PPENDIX DENTIFICATION ATRIX NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 5 YLINDERS ARRANTY IELD ERVICE AND EPAIR ACILITIES IGURES NGA 2000 NDIR Analyzer Module (Typical - Actual configuration Vary)....................1-3 NDIR Technology ..................1-4 Analyzer Module Installation Into Instrument Platform.......2-1 Outline and Mounting Dimensions.............2-4 NDIR Back Panel..................2-5 NDIR Front Panel Electrical Connections ..........2-5 NDIR Wiring Diagram ................2-6...
  • Page 6 ONTENTS OTES NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 7: Preface

    UMMARY The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of this particular NGA 2000 module. Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it.
  • Page 8 General Precautions for Handling and Storing High Pressure Gas Cylinders at the rear of this manual. CAUTION: HAND INJURY HAZARD Dropping the front panel of the Platform while hand or fingers are inside either case handle can cause serious injury. NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 9 Platform is not properly supported. WARNING: OVER-VOLTAGE SPIKING If this Analyzer Module is used with a non-Rosemount Analytical power supply, adding Rosemount Analytical PN 903341 Current Protector in series with the 24V positive power line will prevent over-voltage spiking and resultant fuse blowing when powering up the instrument.
  • Page 10: Glossary

    A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector terminal for communication with external data acquisition devices and provides an input/output function. PERATOR NTERFACE The Display and Keyboard. NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 11 OWER UPPLY Any of a variety of components that provides conditioned power to other NGA 2000 components, from the Power Supply Board that plugs into the front of the Backplane in a stand-alone instrument to several larger ones that can power larger collections of modules and components.
  • Page 12: Specifications

    <2 microns ARTICULATES <40°C (104°F), no entrained liquid EWPOINT Gold plated Pyrex, sapphire, quartz, Irtran, FEP Teflon, Viton- ATERIALS IN ONTACT A, 316 stainless steel WITH AMPLE NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 13 Inside a Platform or custom-installed in a panel OUNTING 1600 m (1 mile) between Analyzer Module and Platform AXIMUM ENGTH OF LON C ABLE See the Preface Section of the Platform manual for specifications regarding Platform related components. 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infrared Analyzer...
  • Page 14: Customer Service, Technical Assistance And Field Service

    Technical Services Department at: Rosemount Analytical Inc. Phone: 1-714-986-7600 FAX: 1-714-577-8006 OCUMENTATION The following NGA 2000 Non-Dispersive Infrared Analyzer instruction materials are available. Contact Customer Service or the local representative to order. 748332 Instruction Manual (this document) NGA Non-Dispersive Infrared Analyzer...
  • Page 15: Compliances

    Nationally Recognized Testing Laboratory), for use in non-hazardous, indoor locations ® 97-C219 APPROVED Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe. NAMUR This product complies with the standard level of NAMUR EMC.
  • Page 16 REFACE OTES P-10 NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 17: Introduction

    1.1 OVERVIEW This manual describes the Non-Dispersive Infrared (NDIR) Analyzer Module of Rosemount Analytical's NGA 200 Series of gas analysis components. See Figure 1-1. The NDIR Analyzer Module is designed to continuously determine the concentration of oxygen in a flowing gaseous mixture. The concentration is expressed in one of three fashions: •...
  • Page 18: Theory Of Technology

    During analysis, a portion of the infrared radiation is absorbed by the component of interest in the sample. The quantity of infrared radiation that is absorbed is proportional to the component concentration. NGA Non-Dispersive Infraraed Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 19 DETECTOR TEMPERATURE CONTROL RTD DETECTOR COVER SIGNAL BOARD OSCILLATOR BOARD OSCILLATOR NETWORK INPUT TUNE ADJUST MODULE SHUTTER ADJUST ACCESS HOLES 1-1. NGA 2000 NDIR A IGURE NALYZER ODULE YPICAL CTUAL ONFIGURATION 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infraraed Analyzer...
  • Page 20: Sample Requirements

    This change, which is proportional to component concentration, is processed and expressed as the primary variable on the network. Other modules comprising the NGA 2000 unit then use this variable for a variety of purposes (e.g., expressing the gas concentration on the Front Panel Display or sending it to external data acquisition devices).
  • Page 21: Purge Kits

    The purge gas vents into the case, which has no outlet fitting for these types of purge gases. 1.6 FEATURES Among the features available in the NDIR Analyzer Module are: • Pressure compensation for barometric fluctuations (optional) • Flow sensing 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infraraed Analyzer...
  • Page 22 NTRODUCTION OTES NGA Non-Dispersive Infraraed Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 23: Unpacking

    Dropping the front panel of the Platform while hand or fingers are inside either handle can cause serious injury. ANALYZER MODULE GUIDES PIN SEATS DISENGAGED FRONT PANEL 2-1. A IGURE NALYZER ODULE NSTALLATION INTO NSTRUMENT LATFORM 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infrared Analyzer...
  • Page 24: Location

    PECIFICATIONS ALIBRATION ASES All applications require a zero standard gas to set the zero point on the display and external data acquisition devices. if the factory provided Calibration and Data Sheet NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 25: Connections

    (See Figure 2-3) Connect inlet and outlet lines for sample/zero/span and flowing reference (if applicable) to appropriately labeled fittings on the rear panel. All four connections are 1/4 inch ferrule-type compression fittings. 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infrared Analyzer...
  • Page 26: Outline And Mounting Dimensions

    3. ELECTRICAL INSTALLATION MUST BE IN COMPLIANCE WITH NATIONAL ELECTRICAL [mm] CODE (ANSI/NFPA 70) AND/OR ANY APPLICABLE NATIONAL OR LOCAL CODES. 2. MODULE IS NOT WEATHERPROOF. 1. APPROXIMATE WEIGHT: 24.2 LB (11.0 kg). 2-2 O IGURE UTLINE AND OUNTING IMENSIONS NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 27: Electrical Connections

    Note: Reference and purge gas connections are applicable only to certain applications. 2-3. NDIR B IGURE ANEL NETWORK 1 NETWORK 2 POWER FUSE 2-4. NDIR F IGURE RONT ANEL LECTRICAL ONNECTIONS 748332-D Rosemount Analytical June 1997 NGA Non-Dispersive Infrared Analyzer...
  • Page 28: Ndir Wiring Diagram

    BOARD DRIVER SENSOR °C RTD, TEMP CTRL DETECTOR BASE MOTOR ASSEMBLY LON/POWER BOARD REF SOURCE SOURCE ASSY E2 24V SAMPLE SOURCE E1 RTN CHASSIS GROUND 2-5. NDIR W IGURE IRING IAGRAM NGA Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 29: Section 3. Startup And Operation

    The Menu structure enables the user to access data and functions, and put information onto the network. The Main Menu (see Figure 3-2) is subdivided into three levels of control based 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 30: Help Displays

    ISPLAYS The Help structure is on-line "tutorial," context-sensitive and topic-interconnected, so that the user can practically operate NGA 2000 without need of an instruction manual. 3.3 STARTUP PROCEDURE Introduce zero gas into SAMPLE INLET and reference and source purge gas, if applicable, into their respective inlets.
  • Page 31: Run Mode Display

    23.2 % CO Analyzer XXXXXXXX Basic Controls Measurement range Numbers: Range upper limit: Range and functional control: Local Calibration… Status: Ready Home Escape Zero Span Info 3-3. B IGURE ASIC ONTROLS 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 32 The softkey marked “HOME” will always return you to this screen. Help menu system... Help on help... Keyboard controls... Editing controls... Home Escape 3-6. T IGURE YPICAL CREEN NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 33: Binding

    To remedy the problem, do the following: 1. Select the following from the Main Menu: Expert Controls and Setup, Analyzer Module Set Up, and Calibration Parameters. 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 34: Ndir Analyzer Module Alarms

    Power Supply, +5V WARNING PERCENT MOD Percent Modulation WARNING RAW SIGNAL Raw Signal WARNING SVFLOW Sample Bypass Flow WARNING SW ERROR Software Error FAILURE 3-1. NDIR A ABLE NALYZER ODULE LARMS NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 35: Linearization

    2500 3000 3500 4000 4500 5000 CONCENTRATION Display ppm 3-7. T IGURE YPICAL INEARIZATION URVE INEARIZER CONCENTRATION NORMALIZED READINGS, NORMALIZED (Axis Reversed) 3-8. O IGURE PERATOR ETERMINED INEARIZATION URVE ORMALIZED 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 36 A0 = 0.00000 A1 = 1.00000 A2 = 0.00000 A3 = 0.00000 A4 = 0.00000 To calculate linearization coefficients other than those installed at the factory, take a NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 37: Routine Operation

    7.6mm of mercury will result in a readout error of about 1% of reading). Therefore, if barometric pressure changes significantly, a recheck of calibration against a span gas is advised. Also, an optional Pressure Compensation Board is available that electronically compensates. 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 38 Calibration status: Ready Calibration status: Ready Error message for last zero: 24-08-94 Error message for last zero: 24-08-94 3-9. D IGURE ISPLAY CREENS 3-10 NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 39 Range 4 factors... Hardware zero offset: 0.0235 V Zero offset 500000 Zero compensation factor: Raw measurement signal: 0.5250 V Span compensation factor: Span factor: 500000 3-10. D IGURE ISPLAY CREENS 748332-D Rosemount Analytical June 1997 3-11 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 40 Fourth line’s parameter: Flow Span gas - range 4: 247.1 ppm May be displayed on the appropriate line of the single analyzer display screen. 3-11. D IGURE ISPLAY CREENS 3-12 NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 41: Display Screens (1 Of 5)

    Last span calibration date: 24-08-94 Last service date: 21-06-94 See #7 above List notes... Add service date! I/O module This function not applicable. Slot position: 3-12. D IGURE ISPLAY CREENS 748332-D Rosemount Analytical June 1997 3-13 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 42: Display Screens (2 Of 5)

    No error And: No error And: No error And: No error And: No error Edit to reset: Report Software error code (1 = no error): 3-13. D IGURE ISPLAY CREENS 3-14 NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 43: Section 4. Maintenance And Troubleshooting

    Also available for maintenance adjustment through the Front Panel display and keypad are the source current, preamplifier gain (referred to as Signal Gain on the sub-menu) and the modulation check. 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 44: Pcb Replacement

    4.4 MODULE FAN REPLACEMENT The Analyzer Module fan assembly is disassembled as shown in Figure 4-2. Before doing so, though, the entire optical bench must be removed. See Figure 4-4. NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 45: Chopper Motor Replacement

    4.8 FLOW SENSOR REPLACEMENT See Figure 4-4 for Flow Sensor location. To replace Flow Sensor, remove all connecting hardware and undo connections to the sample line. 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 46: Case Temperature Sensor Replacement

    (see Figure 3-13, in Primary Variable Parameters and Modulation Check menus), no adjustment is necessary. If not, see Figure 4-4 and 4-5 for Oscillator Tune and Source Balance Shutter adjustments, and do the following: NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical...
  • Page 47 Section 4.12.) 9. Rotate the shutter adjust screw clockwise until the display reads the value obtained in Step 8. 10. Reassemble the Analyzer Module and, if appropriate, the Platform. 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 48: Motor/Source Assembly

    AINTENANCE AND ROUBLESHOOTING Oscillator Tune Adjust Detector Assembly Motor/Source Assembly Shutter Shaft Power Transistor/Heatsink Flow Sensor 4-4. C , PCB A IGURE SSEMBLY XPLODED NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 49: Oscillator Tune, Source Balance Shutter Adjustments

    Source Balance Shutter Adjustment Screw Hole 4-5. O IGURE SCILLATOR OURCE ALANCE HUTTER DJUSTMENTS Detector Oscillator Board Resistive Temperature Detector Thermal Fuse Detector Heater 4-6. D IGURE ETECTOR LOCK XPLODED 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 50: Cleaning Cells

    ±5% of the factory settings (see Figure 3-13, in Primary Variable Parameters and Modulation Check menus). If they are not within that range, follow Steps 7 through 10 of Section 4.11. NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical...
  • Page 51: Cell Disassembly

    4mm TO 2.5 INCHES Desiccant Holder Shutter Block Assembly Spacer Sample Cell Desiccant Holders Motor/Source Detector Assembly Assembly LONG PATH CELLS Sample Cell Shutter Block Assembly Motor/Source Assembly 4.7. C IGURE ISASSEMBLY 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 52: Cell Desiccant

    ±10% of value on the Data Sheet. If it is out of this range, call the factory for further interpretation. If modulation checks are used for trending, variance from check to check should not exceed ±20% per month. 4-10 NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 53: Replacement Parts

    NL902931 Flow Sensor NL655943 Case Temperature Sensor NL655630 Power Supply Board NL655670 Pressure Compensation Board NL655588 Computer Analysis Board NL655610 Signal Board NL655893 NL655891 Chopper Motor NL655925 Source (Matched Pair) 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 54 EPLACEMENT ARTS OTES NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 55: Appendix A. Identification

    ZOO No Selection Code Linearized Non-Standard Range Selection Four Standard Ranges One Non-Standard Range Two Non-Standard Ranges Three Non-Standard Ranges Four Non-Standard Ranges Special Code No Selection Code No Selection Example 748332-D Rosemount Analytical June 1997 NGA 2000 Non-Dispersive Infrared Analyzer...
  • Page 56 A. NDIR I PPENDIX DENTIFICATION ATRIX OTES NGA 2000 Non-Dispersive Infrared Analyzer June 1997 Rosemount Analytical 748332-D...
  • Page 57 8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder. Rosemount Analytical Inc. 4125 E • A 92807-1802 • 714-986-7600 • FAX 714-577-8006...
  • Page 58 (blank)
  • Page 59 SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER. Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control. Rosemount Analytical Rosemount Analytical Inc. 4125 E • A 92807-1802 • 714-986-7600 • FAX 714-577-8006...
  • Page 60 (blank)
  • Page 61 Contact your local Rosemount Sales and Service office for service support. ACTORY For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact: Rosemount Analytical Inc. Process Analytical Division Customer Service Center 1-800-433-6076 ETURNING...
  • Page 62 (blank)