Rosemount Analytical 880A Instruction Manual

Rosemount Analytical 880A Instruction Manual

Non-dispersive infrared analyzer
Table of Contents

Advertisement

Quick Links

Rosemount Analytical 880A Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 880A and is the answer not in the manual?

Questions and answers

Summary of Contents for Rosemount Analytical 880A

  • Page 1 Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
  • Page 2 Rosemount Analytical 880A ODEL ISPERSIVE NFRARED NALYZER NSTRUCTION ANUAL 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 3 OTICE The information contained in this document is subject to change without notice. This manual is based on the production version of the Model 880A Non-Dispersive Infrared Analyzer. Hardware and/or software changes may have occurred since this printing. Teflon® and Viton® are registered trademarks of E.I. duPont de Nemours and Co., Inc.
  • Page 4: Table Of Contents

    CALIBRATION GAS REQUIREMENTS............2-4 SAMPLE HANDLING SYSTEM ..............2-4 LEAK TEST PROCEDURE...............2-5 SAMPLE FLOW RATE ................2-6 2.10 OPTION BOARDS ..................2-7 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 5 3.11.1 STATUS Display................ 3-17 3.12 CURRENT OUTPUT ................3-18 3.13 ZERO/SPAN CALIBRATION WITH CALIBRATION GAS CONTROL ..3-18 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 6 CHOPPER MOTOR ASSEMBLY..............6-5 6.5.1 Long Cell Configurations.............6-5 6.5.2 Short Cell Configurations ............6-5 MICRO BOARD REPLACEMENT ............6-6 DETECTOR REPLACEMENT ..............6-6 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 7 7. R ECTION EPLACEMENT ARTS CIRCUIT BOARD REPLACEMENT ............7-1 SELECTED REPLACEMENT PARTS ............. 7-1 7.2.1 Model 880A Common Parts ............7-2 7.2.2 Case Heater Temperature Control Assembly........ 7-2 7.2.3 Optical Bench ................7-2 HEET & S ENERAL RECAUTIONS FOR...
  • Page 8 Model 880A Detection System ..............1-1 Power Supply Board .................2-2 Case Heater Temperature Control Board..........2-3 Cable Gland Connections.................2-3 Rear View of Model 880A .................2-7 Calibration Gas Control and Alarm Connections ........2-8 Current Output Connections ..............2-9 Auto Zero/Span Board Connections ............2-10 Remote Input/Output Board Connections..........2-10 Model 880A Adjustments................3-1...
  • Page 9 624599 Schematic Diagram, Auto Zero/Span Board 655137 Schematic Diagram, Power Supply Board 655147 Pictorial Wiring Diagram, Model 880A Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 10: Preface

    To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the Model 880A NDIR Analyzer should be thoroughly familiar with and strictly follow the instructions in this manual. Save these instructions.
  • Page 11 If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling. Failure to observe these precautions could result in personal injury and/or damage to the product. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 12: Specifications

    Standard ANSI/NFPA 496-1989, Chapter 8. Consult factory for recommendations. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 13: Options Specifications

    30 VDC Max switching current: 0.5 A Eight optical couplers NPUTS Input Range: +5 VDC to +24 VDC Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 14: Customer Service, Technical Assistance And Field Service

    Technical Services Department at: Rosemount Analytical Inc. Phone: 1-714-986-7600 FAX: 1-714-577-8006 OCUMENTATION The following Model 880A Non-Dispersive Infrared Analyzer instruction materials are available. Contact Customer Service or the local representative to order. 748250 Instruction Manual (this document) 748250-L Rosemount Analytical...
  • Page 15: Compliances

    EMC Directive. If so, the product will carry approval insignia, like those shown here, on the product name rating plate. APPROVED 97-C209 Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe. This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
  • Page 16: Introduction

    NTRODUCTION 1.1 GENERAL DESCRIPTION The Model 880A Non-Dispersive Infrared Analyzer is designed to continuously determine the concentration of a particular component of interest in a flowing gaseous mixture. Within the analyzer, two equal energy infrared beams are directed through two parallel optical cells, a flow-through sample cell and a reference cell. The Luft detector continuously measures the difference in the amount of infrared energy absorbed within each of the two cells.
  • Page 17: Available Options

    NTRODUCTION 1.2 AVAILABLE OPTIONS Operation of the Model 880A can be enhanced with the choice of several options: LARMS STANDARD AND FAIL SAFE User-set dual alarms are available with configurable HI/LO designations and deadband. URRENT UTPUT For normal usage, the 0/4 to 20 mA current output can be set to represent 0 to 100% of fullscale, or a suppressed range of 25% or more of fullscale may be selected.
  • Page 18: Section 2. Unpacking And Installation

    Case Heater Temperature Control Board Figure 2-2 (see Figure 7-2, Case Heater Temperature Control Assembly) must be in the 230 VAC position. Power consumption is 350 watts. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 19: Electrical Connections

    3-wire flexible power cord, minimum 18 AWG (max. O.D. 0.480", min. O.D. 0.270") through hole labeled POWER, utilizing connector gland (P/N 899330) provided. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 20 C a b l e C a s e W a l l 2-3. C IGURE ABLE LAND ONNECTIONS 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 21: Recorder Connections

    • Maximum load impedance: 700 ohm 2.5 SAMPLE INLET/OUTLET CONNECTIONS Most applications of the Model 880A are intended for atmospheric pressure operation with venting to the atmosphere. High pressure and sub-atmospheric pressure applications require careful regulation of the sample pressure. Sample inlet and outlet connections are located on the rear panel.
  • Page 22: Leak Test Procedure

    Apply test liquid to cell or detectors and IR source windows only as a last resort. Should this occur, follow instructions in Section 6.1 to clean these components. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 23: Sample Flow Rate

    Assume that two cell volumes are required to flush any cell. Approximate flushing time for the Model 880A's 12-inch cell at atmospheric sampling pressure (i.e., the outlet of the cell venting to atmosphere) is approximately 12 seconds. Flushing time decreases with cell volume (see Table 2-1).
  • Page 24: Option Boards

    The Alarm, Auto Zero/Span, Calibration Gas Control and Remote Range Change Boards have jumper-selectable addresses (Figures 2-5, 2-7 and 2-8, Section 6.8 and Table 6-2). 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 25: Alarm Connections

    6-pin connector J2. The Calibration Gas Control consists of two form C contacts rated 3A-125/250 VAC or 5A-30 VDC (resistive). Run the cable through the cable gland and tighten the latter once the connector has been secured. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
  • Page 26: Auto Zero/Span Connections

    6 24 0 95 IS O L A T E D V T O I B D 2-6. C IGURE URRENT UTPUT ONNECTIONS 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 27 654416 IS O LA T E D R E M O T E C O N T R O L B D 2-8. R IGURE EMOTE NPUT UTPUT OARD ONNECTIONS 2-10 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 28: Ordering Option Kits

    NPACKING AND NSTALLATION 2.11 ORDERING OPTION KITS Options not ordered from the factory at the time of purchase of the Model 880A may be ordered as the following kits: • 624422 Isolated Remote Control Kit • 624423 Dual Alarm Kit •...
  • Page 29 NPACKING AND NSTALLATION OTES 2-12 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 30: Initial Tartup And Calibration

    Micro Board Gain Adjust Signal Board Motor/Source Detector Module Oscillator Tune 3-1 M 880A A IGURE ODEL DJUSTMENTS 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 31: Front Panel Indicators And Controls

    See Figure 3-3. 3.3.2 F UNCTION The Model 880A has twelve function keys (Figure 3-2). Each key must be pressed firmly to insure that the microprocessor recognizes the keystroke. The definitions for these keys are as follows: To activate the manual zero calibration of the analyzer.
  • Page 32: Model 880A Keypad

    SPAN and ZERO keys. [CALCULATING SPAN] or [CALCULATING ZERO] will then be momentarily displayed. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 33: User Programmable Keys

    For example, if the GAIN is frequently changed, access the RANGE sub-menu to access the GAIN display and press the F1 key. Press SHIFT/ENTER to return to the Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
  • Page 34: Run Mode Display

    The [WAIT-CALCULATING] message is displayed briefly after the user changes displayed data on some of the screens and then exits the screen. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 35: Accessing Mode Displays

    5. Repeat steps 1 through 4 to check the next function. 3.5 SECURITY CODE See Figure 3-5. The Model 880A is equipped with a security code feature, which is deactivated when the instrument is shipped from the factory. When the security feature is activated, only the STATUS and MODE function keys are active to access the STATUS and SECURITY displays.
  • Page 36: Logic Flow Chart

    Run Mode unless user enters Security Mode and REMOTE I/O enters password. 3-4. L IGURE OGIC HART 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 37: Range Parameters

    [RANGE: # CMP NNN →] allows RANGE 1, RANGE 2 or RANGE 3 to be selected with the ↑ or ↓: key. Of these three independent ranges, RANGE 3 should always be the least sensitive range (highest fullscale range). Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
  • Page 38: Component Of Interest

    In the [FS=XXXX ppm *] or [FS=XX.X % *] display, up to a four-digit fullscale value is entered in ppm (parts per million). 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 39: Calibration Gas

    (+5 V, +15 V, +12 V, and -15 V) to be displayed. It is recommended that the values for oscillator tune, detector signal and source current be recorded when the diagnostic display is first accessed. See Figure 3-7. 3-10 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 40: Zero Calibration

    To calibrate the zero setting for the analyzer for all three ranges, press ZERO and then press ENTER. 3.9 SPAN CALIBRATION The Model 880A Analyzer sets span for the selected range in a simple two-keystroke procedure. To span the analyzer, the operator simply presses SPAN and ENTER. When SPAN is pressed, display notes [SPAN - WAIT] and then [X.XX NN % MMM →...
  • Page 41: Linearization

    IAGNOSTIC 3.10 LINEARIZATION The Model 880A Analyzer can be operated in the linear and non-linear mode. Linearization can be toggled ON/OFF with the ↑ or ↓ key in the Linearization Mode Screen (see Figure 3-8) In the OFF position, linearization is disabled for all ranges. In the linear mode, the component of interest is measured in engineering units, either ppm (parts per million) or % (percent of composition).
  • Page 42: All Range

    GAIN and TIME CONSTANT from Range 3, regardless of the GAIN and TIME CONSTANT selected for Ranges 1 and 2. 748250-L Rosemount Analytical October 1999 3-13 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 43: Non-Standard Ranges And Coefficients

    The values of the non-significant digits do not affect the instrument's linearity. 3-14 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 44: Typical Application Linearization Curve, Linearizer Off

    INEARIZATION URVE INEARIZER CONCENTRATION, NORMALIZED READINGS, NORMALIZED (Axis Reversed) 3-10. U IGURE ETERMINED INEARIZATION URVE ORMALIZED 748250-L Rosemount Analytical October 1999 3-15 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 45: Alarm

    SHIFT ENTER will return analyzer to RUN MODE RUN MODE 3-11. A IGURE LARM 3-16 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 46: Status Display

    Alarm 2 (15 sec.) ALARM Alarm: MN (Manual) or AT (Automatic) 3-12. S IGURE TATUS ISPLAY 748250-L Rosemount Analytical October 1999 3-17 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 47: Current Output

    ALIBRATION 3.12 CURRENT OUTPUT Refer to Figure 3-13. The Model 880A Analyzer has a standard 0 to 20 or 4 to 20 mA current output with zero span suppression. This function can represent any suppressed range with at least a 25% span. For example. a valid range could be 0% to 25%, 28% to 61% or 33% to 100%.
  • Page 48: Auto Zero/Span

    At this time a zero and span calibration is made regardless of selection. If a zero delay is selected there will be an automatic two minute delay. 748250-L Rosemount Analytical October 1999 3-19 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 49 SHIFT will return analyzer to RUN MODE SHIFT ENTER 3-14. A IGURE 3-20 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 50 Two reference zero and/or span calibrations after initiation of the Calibration Drift function are required for calculation of the Calibration Drift value. 748250-L Rosemount Analytical October 1999 3-21 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 51: Remote Range Input/Output

    ALIBRATION 3.15 REMOTE RANGE INPUT/OUTPUT Refer to Figure 3-15. The Model 880A Analyzer has optional remote range change and ID capability. When the Remote Range Input/Output is switched to REMOTE in the RUN mode, the range indicator at the right corner of the display will be R# instead of L#.
  • Page 52 1 = High Voltage Pulse (5 to 24 VDC) 0 = Low 3-2. R I/O B ABLE EMOTE ANGE INARY AND ECIMAL ODING 748250-L Rosemount Analytical October 1999 3-23 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 53 NITIAL TARTUP AND ALIBRATION OTES 3-24 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 54: Theory

    (Sections 3.8 and 3.9). Next supply sample gas through the instrument. The Model 880A will now automatically and continuously analyze the sample stream. As a check of instrument performance, it is recommended that the operator keep a log of the zero/span status (percentage of pot value).
  • Page 55 Luft principle. It converts the difference in energy between sample and reference cells to a capacitance change. This capacitance change, proportional to component concentration, is processed and indicated on the display. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
  • Page 56: Error Code Summary

    The analyzer is then operable under values resident before modification. 5-1. E ABLE RROR UMMARY 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 57: Functional Block Diagrams

    ZERO DIGITALLY SPAN 1, 2, 3 CONTROLLED SOFTWARE DIGITALLY ZERO CONTROLLED 5-1. F IGURE UNCTIONAL LOCK IAGRAMS Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 58: Voltage Checks

    R3, located on the Signal Board, for the value obtained in Step 4. THIS VALUE SHOULD NEVER BE HIGHER THAN 7.5. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 59: Digital Gain Adjustment

    4. Using a screwdriver, rotate the shutter adjust screw until a minimum reading on the display is obtained. A typical reading is 0.2 to 0.5. Add 0.4 to this value. Use this Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
  • Page 60: Modulation System

    Decrease in inductance and/or capacitance in tank (10 MHz) circuit shifts resonance curve to right, decreasing carrier amplitude. 5-2. M IGURE ODULATION YSTEM 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 61: Source Alignment

    Heater - Check with an ohmmeter for continuity. The heater resistance is approximately 113 ohms at 25°C. Temperature Sensor - This is an RTD and should have approximately 550 ohms at 25°C. Check with ohmmeter for continuity. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 62: Erl Error Message

    Also, when the ERL message is flashing in the Run Mode screen, if the STATUS key is pressed, the message [EL-LIN.COEFF ERR] will be displayed briefly 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 63 ROUBLESHOOTING OTES Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 64: Routine Servicing

    9. After cleaning cells, examine o-rings at the end cap assembly and end cap/optical filter assembly and replace If damaged. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 65: Short Cell Configurations

    Do not use alcohol or other solvents. Allow to air dry. 10. To install the cells, replace the two retaining screws holding the end cap/optical filter assembly and sample cell assembly to the detector. Verify that the o-rings Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
  • Page 66: Cell Replacement

    End Cap/Optical Filters Assembly O-Rings Cell Assembly Detector Cover Optical Bench Plate 6-1. C IGURE EPLACEMENT 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 67: Cell Desiccant

    Sources are ceramic and can crack or break under excessive pressure. 6.4 SOURCE BALANCE SHUTTER ADJUSTMENT When the sources are replaced, follow the Source Balance Procedure in Section 5.6 to adjust the source balance shutter. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 68 624166 (Sapphire Windows) 624411 (Irtran Windows) Chopper Blade 620300 Chopper Cover 624167 6-2. M IGURE OTOR OURCE SSEMBLY 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 69: Micro Board Replacement

    5. Disconnect ribbon cables. 6. Remove oscillator board. Refer to Figure 6-3C. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 70: Removal - Short Cell Configurations

    2. Adjust source balance shutter and align source (see Sections 5.6 and 5.7). 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 71: Detector Replacement

    Oscillator Board 623998 Detector Thermal Fuse 898733 Temperature Sensor 622917 Detector Base 6-3. D IGURE ETECTOR EPLACEMENT Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 72: Electronic Circuitry

    At the moment the diaphragm reaches maximum distention, the curve reaches the position of Curve 3, Figure 5-3B. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 73: Radio Frequency Demodulator

    Additionally, the power supply board includes an adjustable source driver circuit, a chopper motor driver circuit, a proportional temperature controller circuit and a DC to AC converter for backlight. 6-10 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
  • Page 74: Adapter Board

    This board is used for both dual alarm and calibration gas control, depending on the position of the jumper in the jumper-selectable address. This is a peripheral circuit 748250-L Rosemount Analytical October 1999 6-11 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 75: Dual Alarm/Calibration Gas Control Board

    E5-E7, E9-E10 Auto Zero/Span -------------------- Isolated Remote I/O -------------------- 6-2. J ABLE UMPER ONFIGURATIONS FOR PTION OARDS 6-12 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 76: Replacement

    ARTS The following parts are recommended for routine maintenance and troubleshooting of the Model 880A Analyzer. If the troubleshooting procedures do not resolve the problem, contact your local Rosemount Analytical service office. A list of Rosemount Analytical Service Centers is located in the back of this manual. Figures 7-1, 7-2, and 7-3 show locations of components and assemblies.
  • Page 77: Model 880A Common Parts

    Detector Thermal Fuse 622917 Detector Temperature Sensor 620298 Detector Heater 652605 Chopper Motor 624442 Source (Matched Pair) Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 78: Power Supply Board

    Sensor, Case Heater Temperature Control Option (See Figure 7-2) Signal Board 624088 7-1. M 880A A IGURE ODEL SSEMBLY 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 79 Thermal Fuse 624433 Fan Assembly 622733 Heater Assembly 622732 7-2. C IGURE EATER EMPERATURE ONTROL SSEMBLY Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 80: Motor/Source Assembly

    (See Figure 6-3) Assembly O-Rings Cell Detector Heater 620298 Optical Bench Plate Detector Plate 7-3. O IGURE PTICAL ENCH 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 81 EPLACEMENT ARTS OTES Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 82 8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder. Rosemount Analytical Inc. 4125 E • A 92807-1802 • 714-986-7600 • FAX 714-577-8006...
  • Page 83 (blank) Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 84 SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER. Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control. Rosemount Analytical Rosemount Analytical Inc. 4125 E • A 92807-1802 • 714-986-7600 • FAX 714-577-8006...
  • Page 85 (blank) Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 86 Contact your local Rosemount Sales and Service office for service support. ACTORY For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact: Rosemount Analytical Inc. Process Analytical Division Customer Service Center 1-800-433-6076 ETURNING...
  • Page 87 (blank) Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 88 Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...

Table of Contents