Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 3
OTICE The information contained in this document is subject to change without notice. This manual is based on the production version of the Model 880A Non-Dispersive Infrared Analyzer. Hardware and/or software changes may have occurred since this printing. Teflon® and Viton® are registered trademarks of E.I. duPont de Nemours and Co., Inc.
CALIBRATION GAS REQUIREMENTS............2-4 SAMPLE HANDLING SYSTEM ..............2-4 LEAK TEST PROCEDURE...............2-5 SAMPLE FLOW RATE ................2-6 2.10 OPTION BOARDS ..................2-7 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 5
3.11.1 STATUS Display................ 3-17 3.12 CURRENT OUTPUT ................3-18 3.13 ZERO/SPAN CALIBRATION WITH CALIBRATION GAS CONTROL ..3-18 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 6
CHOPPER MOTOR ASSEMBLY..............6-5 6.5.1 Long Cell Configurations.............6-5 6.5.2 Short Cell Configurations ............6-5 MICRO BOARD REPLACEMENT ............6-6 DETECTOR REPLACEMENT ..............6-6 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 7
7. R ECTION EPLACEMENT ARTS CIRCUIT BOARD REPLACEMENT ............7-1 SELECTED REPLACEMENT PARTS ............. 7-1 7.2.1 Model 880A Common Parts ............7-2 7.2.2 Case Heater Temperature Control Assembly........ 7-2 7.2.3 Optical Bench ................7-2 HEET & S ENERAL RECAUTIONS FOR...
Page 8
Model 880A Detection System ..............1-1 Power Supply Board .................2-2 Case Heater Temperature Control Board..........2-3 Cable Gland Connections.................2-3 Rear View of Model 880A .................2-7 Calibration Gas Control and Alarm Connections ........2-8 Current Output Connections ..............2-9 Auto Zero/Span Board Connections ............2-10 Remote Input/Output Board Connections..........2-10 Model 880A Adjustments................3-1...
Page 9
624599 Schematic Diagram, Auto Zero/Span Board 655137 Schematic Diagram, Power Supply Board 655147 Pictorial Wiring Diagram, Model 880A Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the Model 880A NDIR Analyzer should be thoroughly familiar with and strictly follow the instructions in this manual. Save these instructions.
Page 11
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling. Failure to observe these precautions could result in personal injury and/or damage to the product. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
Technical Services Department at: Rosemount Analytical Inc. Phone: 1-714-986-7600 FAX: 1-714-577-8006 OCUMENTATION The following Model 880A Non-Dispersive Infrared Analyzer instruction materials are available. Contact Customer Service or the local representative to order. 748250 Instruction Manual (this document) 748250-L Rosemount Analytical...
EMC Directive. If so, the product will carry approval insignia, like those shown here, on the product name rating plate. APPROVED 97-C209 Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to harmonize the product requirements in Europe. This product complies with the standard level of NAMUR EMC. Recommendation (May 1993).
NTRODUCTION 1.1 GENERAL DESCRIPTION The Model 880A Non-Dispersive Infrared Analyzer is designed to continuously determine the concentration of a particular component of interest in a flowing gaseous mixture. Within the analyzer, two equal energy infrared beams are directed through two parallel optical cells, a flow-through sample cell and a reference cell. The Luft detector continuously measures the difference in the amount of infrared energy absorbed within each of the two cells.
NTRODUCTION 1.2 AVAILABLE OPTIONS Operation of the Model 880A can be enhanced with the choice of several options: LARMS STANDARD AND FAIL SAFE User-set dual alarms are available with configurable HI/LO designations and deadband. URRENT UTPUT For normal usage, the 0/4 to 20 mA current output can be set to represent 0 to 100% of fullscale, or a suppressed range of 25% or more of fullscale may be selected.
Case Heater Temperature Control Board Figure 2-2 (see Figure 7-2, Case Heater Temperature Control Assembly) must be in the 230 VAC position. Power consumption is 350 watts. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
3-wire flexible power cord, minimum 18 AWG (max. O.D. 0.480", min. O.D. 0.270") through hole labeled POWER, utilizing connector gland (P/N 899330) provided. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
Page 20
C a b l e C a s e W a l l 2-3. C IGURE ABLE LAND ONNECTIONS 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
• Maximum load impedance: 700 ohm 2.5 SAMPLE INLET/OUTLET CONNECTIONS Most applications of the Model 880A are intended for atmospheric pressure operation with venting to the atmosphere. High pressure and sub-atmospheric pressure applications require careful regulation of the sample pressure. Sample inlet and outlet connections are located on the rear panel.
Apply test liquid to cell or detectors and IR source windows only as a last resort. Should this occur, follow instructions in Section 6.1 to clean these components. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Assume that two cell volumes are required to flush any cell. Approximate flushing time for the Model 880A's 12-inch cell at atmospheric sampling pressure (i.e., the outlet of the cell venting to atmosphere) is approximately 12 seconds. Flushing time decreases with cell volume (see Table 2-1).
The Alarm, Auto Zero/Span, Calibration Gas Control and Remote Range Change Boards have jumper-selectable addresses (Figures 2-5, 2-7 and 2-8, Section 6.8 and Table 6-2). 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
6-pin connector J2. The Calibration Gas Control consists of two form C contacts rated 3A-125/250 VAC or 5A-30 VDC (resistive). Run the cable through the cable gland and tighten the latter once the connector has been secured. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
6 24 0 95 IS O L A T E D V T O I B D 2-6. C IGURE URRENT UTPUT ONNECTIONS 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 27
654416 IS O LA T E D R E M O T E C O N T R O L B D 2-8. R IGURE EMOTE NPUT UTPUT OARD ONNECTIONS 2-10 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
NPACKING AND NSTALLATION 2.11 ORDERING OPTION KITS Options not ordered from the factory at the time of purchase of the Model 880A may be ordered as the following kits: • 624422 Isolated Remote Control Kit • 624423 Dual Alarm Kit •...
Page 29
NPACKING AND NSTALLATION OTES 2-12 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
See Figure 3-3. 3.3.2 F UNCTION The Model 880A has twelve function keys (Figure 3-2). Each key must be pressed firmly to insure that the microprocessor recognizes the keystroke. The definitions for these keys are as follows: To activate the manual zero calibration of the analyzer.
SPAN and ZERO keys. [CALCULATING SPAN] or [CALCULATING ZERO] will then be momentarily displayed. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
For example, if the GAIN is frequently changed, access the RANGE sub-menu to access the GAIN display and press the F1 key. Press SHIFT/ENTER to return to the Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
The [WAIT-CALCULATING] message is displayed briefly after the user changes displayed data on some of the screens and then exits the screen. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
5. Repeat steps 1 through 4 to check the next function. 3.5 SECURITY CODE See Figure 3-5. The Model 880A is equipped with a security code feature, which is deactivated when the instrument is shipped from the factory. When the security feature is activated, only the STATUS and MODE function keys are active to access the STATUS and SECURITY displays.
[RANGE: # CMP NNN →] allows RANGE 1, RANGE 2 or RANGE 3 to be selected with the ↑ or ↓: key. Of these three independent ranges, RANGE 3 should always be the least sensitive range (highest fullscale range). Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
In the [FS=XXXX ppm *] or [FS=XX.X % *] display, up to a four-digit fullscale value is entered in ppm (parts per million). 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
(+5 V, +15 V, +12 V, and -15 V) to be displayed. It is recommended that the values for oscillator tune, detector signal and source current be recorded when the diagnostic display is first accessed. See Figure 3-7. 3-10 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
To calibrate the zero setting for the analyzer for all three ranges, press ZERO and then press ENTER. 3.9 SPAN CALIBRATION The Model 880A Analyzer sets span for the selected range in a simple two-keystroke procedure. To span the analyzer, the operator simply presses SPAN and ENTER. When SPAN is pressed, display notes [SPAN - WAIT] and then [X.XX NN % MMM →...
IAGNOSTIC 3.10 LINEARIZATION The Model 880A Analyzer can be operated in the linear and non-linear mode. Linearization can be toggled ON/OFF with the ↑ or ↓ key in the Linearization Mode Screen (see Figure 3-8) In the OFF position, linearization is disabled for all ranges. In the linear mode, the component of interest is measured in engineering units, either ppm (parts per million) or % (percent of composition).
GAIN and TIME CONSTANT from Range 3, regardless of the GAIN and TIME CONSTANT selected for Ranges 1 and 2. 748250-L Rosemount Analytical October 1999 3-13 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
The values of the non-significant digits do not affect the instrument's linearity. 3-14 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
SHIFT ENTER will return analyzer to RUN MODE RUN MODE 3-11. A IGURE LARM 3-16 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
ALIBRATION 3.12 CURRENT OUTPUT Refer to Figure 3-13. The Model 880A Analyzer has a standard 0 to 20 or 4 to 20 mA current output with zero span suppression. This function can represent any suppressed range with at least a 25% span. For example. a valid range could be 0% to 25%, 28% to 61% or 33% to 100%.
At this time a zero and span calibration is made regardless of selection. If a zero delay is selected there will be an automatic two minute delay. 748250-L Rosemount Analytical October 1999 3-19 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 49
SHIFT will return analyzer to RUN MODE SHIFT ENTER 3-14. A IGURE 3-20 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 50
Two reference zero and/or span calibrations after initiation of the Calibration Drift function are required for calculation of the Calibration Drift value. 748250-L Rosemount Analytical October 1999 3-21 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
ALIBRATION 3.15 REMOTE RANGE INPUT/OUTPUT Refer to Figure 3-15. The Model 880A Analyzer has optional remote range change and ID capability. When the Remote Range Input/Output is switched to REMOTE in the RUN mode, the range indicator at the right corner of the display will be R# instead of L#.
Page 52
1 = High Voltage Pulse (5 to 24 VDC) 0 = Low 3-2. R I/O B ABLE EMOTE ANGE INARY AND ECIMAL ODING 748250-L Rosemount Analytical October 1999 3-23 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 53
NITIAL TARTUP AND ALIBRATION OTES 3-24 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
(Sections 3.8 and 3.9). Next supply sample gas through the instrument. The Model 880A will now automatically and continuously analyze the sample stream. As a check of instrument performance, it is recommended that the operator keep a log of the zero/span status (percentage of pot value).
Page 55
Luft principle. It converts the difference in energy between sample and reference cells to a capacitance change. This capacitance change, proportional to component concentration, is processed and indicated on the display. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
The analyzer is then operable under values resident before modification. 5-1. E ABLE RROR UMMARY 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
R3, located on the Signal Board, for the value obtained in Step 4. THIS VALUE SHOULD NEVER BE HIGHER THAN 7.5. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
4. Using a screwdriver, rotate the shutter adjust screw until a minimum reading on the display is obtained. A typical reading is 0.2 to 0.5. Add 0.4 to this value. Use this Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
Heater - Check with an ohmmeter for continuity. The heater resistance is approximately 113 ohms at 25°C. Temperature Sensor - This is an RTD and should have approximately 550 ohms at 25°C. Check with ohmmeter for continuity. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
Also, when the ERL message is flashing in the Run Mode screen, if the STATUS key is pressed, the message [EL-LIN.COEFF ERR] will be displayed briefly 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 63
ROUBLESHOOTING OTES Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
9. After cleaning cells, examine o-rings at the end cap assembly and end cap/optical filter assembly and replace If damaged. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Do not use alcohol or other solvents. Allow to air dry. 10. To install the cells, replace the two retaining screws holding the end cap/optical filter assembly and sample cell assembly to the detector. Verify that the o-rings Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ...
Sources are ceramic and can crack or break under excessive pressure. 6.4 SOURCE BALANCE SHUTTER ADJUSTMENT When the sources are replaced, follow the Source Balance Procedure in Section 5.6 to adjust the source balance shutter. Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
At the moment the diaphragm reaches maximum distention, the curve reaches the position of Curve 3, Figure 5-3B. 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Additionally, the power supply board includes an adjustable source driver circuit, a chopper motor driver circuit, a proportional temperature controller circuit and a DC to AC converter for backlight. 6-10 Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L...
This board is used for both dual alarm and calibration gas control, depending on the position of the jumper in the jumper-selectable address. This is a peripheral circuit 748250-L Rosemount Analytical October 1999 6-11 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
ARTS The following parts are recommended for routine maintenance and troubleshooting of the Model 880A Analyzer. If the troubleshooting procedures do not resolve the problem, contact your local Rosemount Analytical service office. A list of Rosemount Analytical Service Centers is located in the back of this manual. Figures 7-1, 7-2, and 7-3 show locations of components and assemblies.
Sensor, Case Heater Temperature Control Option (See Figure 7-2) Signal Board 624088 7-1. M 880A A IGURE ODEL SSEMBLY 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 79
Thermal Fuse 624433 Fan Assembly 622733 Heater Assembly 622732 7-2. C IGURE EATER EMPERATURE ONTROL SSEMBLY Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
(See Figure 6-3) Assembly O-Rings Cell Detector Heater 620298 Optical Bench Plate Detector Plate 7-3. O IGURE PTICAL ENCH 748250-L Rosemount Analytical October 1999 Model 880A Non-Dispersive Infrared Analyzer Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 81
EPLACEMENT ARTS OTES Model 880A Non-Dispersive Infrared Analyzer October 1999 Rosemount Analytical 748250-L Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 82
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder. Rosemount Analytical Inc. 4125 E • A 92807-1802 • 714-986-7600 • FAX 714-577-8006...
Page 84
SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER. Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control. Rosemount Analytical Rosemount Analytical Inc. 4125 E • A 92807-1802 • 714-986-7600 • FAX 714-577-8006...
Page 86
Contact your local Rosemount Sales and Service office for service support. ACTORY For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact: Rosemount Analytical Inc. Process Analytical Division Customer Service Center 1-800-433-6076 ETURNING...
Page 88
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...