Sony DG10B Series Instruction Manual page 46

Digital gauge
Hide thumbs Also See for DG10B Series:
Table of Contents

Advertisement

Available languages

Available languages

When using a power supply with remote sensing function
Line driver
SIG
SIG
*SIG
AM26C31
Outputs: A/B phase
Point a
+V
+S
_S
0 V
When using a power supply without remote sensing function
Line driver
SIG
SIG
*SIG
AM26C31
Outputs: A/B phase
*
+V
+S
_S
0 V
* When using DG10BNM/DG10BPM/DG10BSM/DG25BNM/DG25BPM/
DG25BLM/DG25BSM, the main unit input should be at +5 V ± 5%.
14
(E)
50 m or less
AWG26 or thicker
twisted pair cable
30 m or less
AWG20 or thicker
twisted pair cable
AWG20 or more
Receiving device
Line receiver:
AM26C32 or
equivalent product
120
AM26C32
+V
+S
_S
0 V
Receiving device
Line receiver:
AM26C32 or
equivalent product
120
AM26C32
+V
0 V
DG10B/DG25B Series

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents