Suggested Settings For Use In 'Peak Picker' Mode - Ametek LAND UNO User Supplement

Target orbiting thermometer

Advertisement

Target Orbiting Thermometer
The target width required for accurate temperature measurement (>99.5% of
value) with the Target Orbiting Thermometer is as follows:
Thermometer
U1 600/1600 CV - TO
U1 600/1600 CS
U1 1100/2900 FV
U1 1100/2900 FS
U1 800/2600 CV
U1 800/2600 CS
U1 1500/4700 FV
U1 1500/4700 FS
U2 300/1100 CV
U2 300/1100 CS
U2 600/2000 FV
U2 600/2000 FS

3.1 Suggested settings for use in 'Peak Picker' mode

1)
For instructions on how to access the time function controls, refer to
Section 3 of the User Guide.
2)
Set the switch SW5 (Time function selector) to 2 i.e. Peak Picker.
3)
Set the switch SW4 (Decay Rate %/s) to 9. This is recommended as an
initial setting as it minimises the decay between peaks.
4)
After initial setup, adjust the setting of SW4 to suit the requirements of
your particular application.
Page 4
Field of View
Minimum target required
100:1
1.2 x target spot diameter
100:1
1.2 x target spot diameter
100:1
1.2 x target spot diameter
100:1
1.2 x target spot diameter
200:1
1.5 x target spot diameter
200:1
1.5 x target spot diameter
200:1
1.5 x target spot diameter
200:1
1.5 x target spot diameter
100:1
1.2 x target spot diameter
100:1
1.2 x target spot diameter
100:1
1.2 x target spot diameter
100:1
1.2 x target spot diameter
User Supplement