Additional blank copies of the Configuration Reference form can be found in Section 13. These blank forms are intended for use by the user to document user-defined changes from the factory default settings. Sensidyne Document No. 360-0152-01 (Rev F0)
Sensidyne Representative, or call 800-451-9444 or +1 727-530-3602 PROPRIETARY NOTICE This manual was prepared by Sensidyne, LP exclusively for the owner of the SensAlert ASI Universal Gas Transmitter. The material within this manual is the proprietary information of Sensidyne, LP and is to be used only to understand and operate the instrument.
DO NOT attempt to repair or modify instrument, except as specified in the Operation & Service Manual. If repair is needed, contact the Sensidyne Service Dept. to arrange for a Returned Material Authorization (RMA) (See Section 12 for details).
Page 11
Hazardous Area Certification. Repair or alteration of the product beyond the scope of these maintenance instructions, or by anyone other than an authorized SENSIDYNE® service technician, could cause the product to fail to perform as designed and persons who rely on this product for their safety could sustain severe personal injury or death.
Transmitter. The transmitter is capable of detecting the presence of potentially hazardous concentrations of a target gas. 1.1 Product Versions SensAlert ASI is a configurable instrument. The part numbering convention contains all necessary information to express the configuration: Sensidyne Document No. 360-0152-01 (Rev F0)
(part number 7013201-1). The integral LCD readout has been calibrated at the factory for direct reading in ppm, %vol, or %LEL. Field calibration consists of calibrating the sensor output to the applied calibration gas. Sensidyne Document No. 360-0152-01 (Rev F0)
An optional Relay Board can be installed that provides three additional relays for use with horns, strobes, and other external devices. 1.3.4 Communication Options An optional Communications Board can be installed for RS-485 with Modbus, or HART protocol. Sensidyne Document No. 360-0152-01 (Rev F0)
The rainshield is IP56 (Ingress Protection) rated. 1.4.6 Calibration Plug Assembly The calibration plug assembly secures into the sensor shield and replaces the rainshield while the transmitter is being calibrated. Sensidyne Document No. 360-0152-01 (Rev F0)
Page 16
SensAlert ASI – Advanced Safety Integrity Gas Transmitter Figure 2-1a SensAlert ASI Transmitter Standard Horizontal Mount Sensidyne Document No. 360-0152-01 (Rev F0)
Page 17
SensAlert ASI – Advanced Safety Integrity Gas Transmitter Figure 2-2b SensAlert ASI Transmitter Standard Vertical Mount Sensidyne Document No. 360-0152-01 (Rev F0)
Page 18
SensAlert ASI – Advanced Safety Integrity Gas Transmitter Figure 2-3c SensAlert ASI Transmitter for two option cards Sensidyne Document No. 360-0152-01 (Rev F0)
(Note: LEDs do not light up under any circumstances on 2-wire transmitters.) (9) This line is used to display important system messages and warnings. Sensidyne Document No. 360-0152-01 (Rev F0)
ONLY SENSORS LABELED AS PLUS SERIES WILL WORK IN THE SENSALERT ASI TRANSMITTER. A complete list of available sensors, sensor specifications, interferents, and calibration equipment can be found by contacting Sensidyne or on the sensor specification sheet. Sensors should be hot-swapped only when the Normal Operation Mode screen is displayed.
Page 21
“Loop-Fail” condition by providing an unvarying default output of < 3.0 mA (user adjustable 1-4 mA 3 wire Transmitter, 3-4 mA 2wire Transmitter). This is shown as “MISSING SENSOR” on the display. Sensidyne Document No. 360-0152-01 (Rev F0)
Expert consultation may be necessary to determine the most appropriate location for optimum detecting. In all circumstances, the plant safety officer or other appropriate personnel should be consulted before installation. Sensidyne Document No. 360-0152-01 (Rev F0)
Page 23
• physical properties of the target gas • air convection in the area due to ventilation or ambient conditions • operational environment (temperature, humidity, wind, etc.) • presence of interferent gases Sensidyne Document No. 360-0152-01 (Rev F0)
6) Cap the wires and replace the condulet cover if you are not going to wire the transmitter at this time. (You need not tighten the hex head screw at this time.) Sensidyne Document No. 360-0152-01 (Rev F0)
3) Locate the power supply board attached to the display board. The power terminal block is located on the underside of the power supply board. Once the board stack has been removed wire the transmitter according to the instructions found in Section 2.4. Sensidyne Document No. 360-0152-01 (Rev F0)
NOTE The relay terminal block is designated as “Relay 1” and is the factory-set fault relay. Once the board stack has been removed wire the transmitter according to the instructions found in Section 2.4. Sensidyne Document No. 360-0152-01 (Rev F0)
3) Locate the power supply board attached to the display board. The power terminal block is located on the underside of the power supply board. Once the board stack has been removed wire the transmitter according to the instructions found in Section 2.4. Sensidyne Document No. 360-0152-01 (Rev F0)
The relay terminal block [E-8] is designated as “Relay 1” and is the factory-set fault relay. Once the board stack has been removed, wire the transmitter according to the instructions found in Section 2.4. Sensidyne Document No. 360-0152-01 (Rev F0)
RFI/EMI immunity (dependent on the transmitter type). Either two or three wires are required to connect a transmitter to a Sensidyne controller or a user supplied 18–30 VDC power supply and readout device. Any supplemental readout device used must have a termination resistance of 250 ohms or less (100 ohms or less for 2-wire transmitter).
The table below assumes cable inductance of 0.2 µH/ft (or less) and cable capacitance of 60 pf/ft or less. The wire length specified below is between Safety Barrier and Transmitter. If you cable does not meet these specifications, please contact Sensidyne. Maximum Wire Length –Three wire I.S. Transmitters...
1) Get confirmation from the safety officer, or appropriate personnel, that the area is declassified. 2) Verify that the conduit and the transmitter are properly connected. 3) Verify that the input and output wires are not energized. Sensidyne Document No. 360-0152-01 (Rev F0)
Page 32
8) Replace / close the transmitter cover. For transmitters with metallic condulet covers, tighten the hex head screw to secure the condulet cover. 9) Go to Section 2.4.4 to perform Start Up. Figure 4 - 700-0100-01 2-Wire Power Supply Sensidyne Document No. 360-0152-01 (Rev F0)
9) Replace / close the transmitter cover. For transmitters with metallic condulet covers, tighten the hex head screw to secure the condulet cover. 10) Go to Section 2.4.4 to perform Start Up. Figure 5 - 700-0101-01 3-Wire Power Supply Sensidyne Document No. 360-0152-01 (Rev F0)
4) Replace /close the transmitter cover. For transmitters with metallic condulet covers, tighten the hex head screw to secure the condulet cover. 5) Go to Section 2.4.4 to perform Start Up. Figure 6 - 700-0102-01 3-Wire Power Supply Sensidyne Document No. 360-0152-01 (Rev F0)
4) Replace /close the transmitter cover. For transmitters with metallic condulet covers, tighten the hex head screw to secure the condulet cover. 5) Go to Section 2.4.4 to perform Start Up. Figure 7 - 700-0114-01 Relay PCA Sensidyne Document No. 360-0152-01 (Rev F0)
WARNING The 3-Wire I.S. Transmitters (S2S-XXXX-XX) are designed for use exclusively with the Sensidyne Intrinsic Safety Barrier (P/N add replacement barrier pn). Failure to use this barrier will void all product warranties and may jeopardize intrinsic safety protections and result in permanent damage to the transmitter.
6) Allow the transmitter to stabilize at least one (1) hour before zeroing. During stabilization, the display reading should change toward “0” (or 20.9 %vol for ambient Oxygen sensors). After stabilization has been completed go to Section 3.1 and perform the zeroing procedure (even if the display shows “0”). Sensidyne Document No. 360-0152-01 (Rev F0)
1) Select ACK on the Normal Operation Display to bring up the Main Menu. 2). Select ACK to bring up the Calibration Mode menu. > Calibration Mode Maintenance Mode Data Review Test-On-Demand System Configuration Lost Password Sensidyne Document No. 360-0152-01 (Rev F0)
Page 39
If zeroing is not successful for any reason this screen will not appear. If zero fails, try zeroing a second time or replacing the sensor. If zeroing fails repeatedly contact the Sensidyne Service Department. Sensidyne Document No. 360-0152-01 (Rev F0)
3.2.3 Span Calibration Procedure Your Sensidyne gas detector is tested and calibrated at the factory prior to shipment. However, this product must be calibrated prior to initial use and at regular intervals in accordance with this User Manual to ensure proper operation.
Page 41
After the calibration equipment has been set up, continue with calibration as shown below: 1) Select ACK with the magnetic wand to bring up the Main Menu. Note: To cancel calibration at any time select <<. Sensidyne Document No. 360-0152-01 (Rev F0)
Page 42
5) Use the arrows to adjust the calibration gas level. Select ACK to set the new gas level. The unit returns to the Calibration Mode menu after a few seconds. Set Cal Gas Conc 25 PPM Sensidyne Document No. 360-0152-01 (Rev F0)
Page 43
60 second period). The gas Cal In Progress concentration used for calibration is that value displayed when Cal To: 50 PPM stability is reached. Level: 50 PPM Select << To Cancel Sensidyne Document No. 360-0152-01 (Rev F0)
1) From the Main Menu use the ▼ arrow to scroll to “System Configuration.” Select ACK to bring up the System Configuration Menu. Main Menu Calibration Mode Maintenance Mode Data Review Test On Demand > System Configuration Lost Password Sensidyne Document No. 360-0152-01 (Rev F0)
The screen below it typically appears if an Infrared Select Cal Gas Combustible sensor is installed (see Note below). Select ACK Propane Has Been Selected to save the new cal gas. Select ACK to Save Select << To Cancel Sensidyne Document No. 360-0152-01 (Rev F0)
Use the ▼ arrow to scroll to “Propane” and select Hexane Butane ACK. Custom K Factor Note K Factors are preset for the following gases: None, Hydrogen, Propane, Hexane, Butane, Custom K factor Sensidyne Document No. 360-0152-01 (Rev F0)
K-Factors are not applicable when an Infrared Combustible Ethane sensor is installed and Methane is used as the calibration gas. Note K Factors are preset for the following gases: None, Custom K Factor, Acetone, Butadiene, Butane, Ethane, Ethanol, Ethyl Sensidyne Document No. 360-0152-01 (Rev F0)
Catalytic Bead Combustible sensor is installed and is Select ACK to Save calibrated with Propane. An expanded list of K-Factors can be Select << to Cancel found in the Sensor Data Sheets for the Catalytic Bead Combustible and Infrared Combustible sensors. Sensidyne Document No. 360-0152-01 (Rev F0)
This indicates that there are additional menu items that cannot be displayed in the initial screen. Use the ▼ arrow to continue scrolling down the list until “>” is next to “TOD Mode Adjustment.” Sensidyne Document No. 360-0152-01 (Rev F0)
Page 50
(e.g., “Test Date/Time,” “Cell Intensity,” etc.). Select ACK to bring up that menu item. TOD Mode Adjustment > Auto Mode Enable Test Date/Time Days Between Tests Cell Intensity Output Indicators Sensidyne Document No. 360-0152-01 (Rev F0)
(i.e., both the system and test times change to 17:01, 17:02, 17:03, etc.). If you then enable the Auto Mode a test will be immediately performed, and will be repeated at intervals set in the Days Between Tests Menu. Sensidyne Document No. 360-0152-01 (Rev F0)
Test Date/Time environmental conditions (e.g., high winds). Days Between Tests > Cell Intensity Access the TOD Mode Adjustment Menu as shown in Output Indicators Section 3.4.1. Sensidyne Document No. 360-0152-01 (Rev F0)
2) A screen appears showing the current output indicator, in this case “None.” To change the output indicator use the ▼ arrow to scroll to “4-20 Current Loop.” Select ACK to save. Output Indicators > None 4-20 Current Loop Sensidyne Document No. 360-0152-01 (Rev F0)
1) The screen shows that Alarm 1 is enabled. To disable the alarm use the ▼ arrow to scroll to “Enabled” and select ACK. Alarm 1 Relays Assigned Add Relay Delete Relay > Enabled Ascending Setpoint Release Offset Sensidyne Document No. 360-0152-01 (Rev F0)
1) Use the ▼ arrow to scroll to “TWA Alarm.” Select ACK to bring Alarm Functions up the TWA Alarm screen. Alarm 1 Alarm 2 Alarm 3 > TWA Alarm Select ACK to Change Sensidyne Document No. 360-0152-01 (Rev F0)
Page 59
3) For this example use the ▼ arrow to scroll to “8 Hours.” Select ACK to change the time period. TWA Alarm Average Time Adjust 15 Minutes 30 Minutes 60 Minutes 2 Hours 4 Hours > 8 Hours Sensidyne Document No. 360-0152-01 (Rev F0)
Down = Decrease Adjust Current Delay 4.3.1.5 The Adjust Current Delay menu is used to change the fault current duration for the specific fault. In this example, when a Sensor End of Life fault occurs Sensidyne Document No. 360-0152-01 (Rev F0)
This is indicated when the Select ACK to Save LED associated with the “latched” relay remains lit. The relay must be manually deactivated (reset) by selecting << from the Normal Operation Display screen (see screen at right). Sensidyne Document No. 360-0152-01 (Rev F0)
Time Delay Select ACK to Change 2) A screen appears showing that-Relay 1 is now “Norm De- energized.” Select ACK to save the change. Relay 1 - Fail Norm De-energized Select ACK to Save Sensidyne Document No. 360-0152-01 (Rev F0)
Access the Warn Current – Relay 5 Menu as shown in Section 4.1. The following screen will be displayed. 4.5.1 Latching/Non-Latching 1) The screen shows that the Warn Current-Relay 5 is currently Non-Latching. To change the status select ACK. Warn Current-Relay5 Latching Select ACK to Save Sensidyne Document No. 360-0152-01 (Rev F0)
Non-Latching Disabled 1) To adjust the low current time use the ▼ arrow to scroll to “Low > Low Current Time Adj Current Time Adj” and select ACK. Select ACK to Change Sensidyne Document No. 360-0152-01 (Rev F0)
Page 67
1 second, while the maximum time delay is 240 seconds. When finished select ACK to save the new Warn Current-Relay5 Low Output Current value. Delay 020 Seconds UP = Increase DOWN = Decrease Select ACK to Save Sensidyne Document No. 360-0152-01 (Rev F0)
“Cal In Progress” (stabilizing) 5.2.1.2.4 “Cal In Progress” (stable) 5.2.1.2.5 “Calibration Complete” (either “OK” or “FAIL”) 5.2.1.3 Set Cal Gas Conc. 5.2.1.3.1 Screen for changing concentration 5.2.2 Maintenance Mode 5.2.2.1.1 “Transmitter will be isolated for 10 minutes” Sensidyne Document No. 360-0152-01 (Rev F0)
-- Comm Installed / No Comm Installed 5.2.4 Test On Demand 5.2.4.1 4-20 mA Not Active 5.2.4.1.1 -- “In Progress” screen, then result 5.2.4.2 4-20 mA Active 5.2.4.2.1 -- “In Progress” screen, then result Sensidyne Document No. 360-0152-01 (Rev F0)
Ascending ◄► Descending 5.2.5.2.1.2.4 -- Select “ACK” to change status, then select “ACK” again to save 5.2.5.2.1.2.5 Setpoint -- Use ▲ and ▼ to change value, then select “ACK” to save 5.2.5.2.1.2.6 Release Offset Sensidyne Document No. 360-0152-01 (Rev F0)
Page 73
-- Use ▲ and ▼ to change value, then select “ACK” to save 5.2.5.2.1.4.6 Average Time Adjust 5.2.5.2.1.4.6.1 15 Minutes 5.2.5.2.1.4.6.2 30 Minutes 5.2.5.2.1.4.6.3 60 Minutes 5.2.5.2.1.4.6.4 2 Hours 5.2.5.2.1.4.6.5 4 Hours 5.2.5.2.1.4.6.6 8 Hours 5.2.5.2.2 Fault Functions Sensidyne Document No. 360-0152-01 (Rev F0)
Page 74
Enabled ◄► Disabled 5.2.5.2.2.3.3 -- Select “ACK” to change status, then select “ACK” again to save 5.2.5.2.2.3.4 Adjust Fault Current -- Use ▲and ▼to change value, then select “ACK” to save 5.2.5.2.2.4 Sensor End of Life Sensidyne Document No. 360-0152-01 (Rev F0)
Page 75
-- Select “ACK” to delete relay(s), then “<<” when done 5.2.5.2.2.6.2.1 Relay 1 – Fail 5.2.5.2.2.6.2.2 Relay 2 5.2.5.2.2.6.2.3 Relay 3 5.2.5.2.2.6.2.4 Relay 4 Enabled ◄► Disabled 5.2.5.2.2.6.3 -- Select “ACK” to change status, then select “ACK” again to save Sensidyne Document No. 360-0152-01 (Rev F0)
Page 76
Relay 3 5.2.5.2.2.9.2.4 Relay 4 5.2.5.2.2.9.3 Adjust Fault Current -- Use ▲and ▼to change value, then select “ACK” to save 5.2.5.2.2.9.4 Adjust Current Delay --Use ▲and ▼to change value, then select “ACK” to save Sensidyne Document No. 360-0152-01 (Rev F0)
Page 77
-- Use ▲ and ▼ to change value, then select “ACK” to save 5.2.5.2.6 Relay 4 Latching ◄► Non-Latching 5.2.5.2.6.1 -- Select “ACK” to change status, then select “ACK” again to save Norm Energized ◄► Norm De-energized 5.2.5.2.6.2 Sensidyne Document No. 360-0152-01 (Rev F0)
Select K Factor (IR CMB Sensor) 5.2.5.7.9 If Cal Gas is Methane 5.2.5.7.9.1 None 5.2.5.7.10 If Cal Gas is Propane 5.2.5.7.10.1 None 5.2.5.7.10.2 Custom K Factor 5.2.5.7.10.3 Acetone 5.2.5.7.10.4 Butadiene 5.2.5.7.10.5 Butane 5.2.5.7.10.6 Ethane 5.2.5.7.10.7 Ethanol Sensidyne Document No. 360-0152-01 (Rev F0)
Select “ACK” to restore system to Factory Defaults 5.2.5.10 Set Transmitter Tag 5.2.5.10.1 Enter New Transmitter Tag 5.2.5.10.2 Accept New Transmitter Tag 5.2.5.11 Screen Contract 5.2.5.11.1 Adjust Screen Contract Up (Increase) or Down (Decrease) 5.2.6 Lost Password Sensidyne Document No. 360-0152-01 (Rev F0)
Select OK To Continue is not within acceptable limits the display will show “Zero Complete” and “Fail” and the previously stored zero reading will not be altered. See Section 3.1 for Select << To Cancel more information. Sensidyne Document No. 360-0152-01 (Rev F0)
Relays, the relays will not change state while in Maintenance Mode. This is to Maintenance Mode allow for periodic maintenance on the unit without setting off alarms. Transmitter will be isolated for 10 minutes Select OK To Continue Select << To Exit Sensidyne Document No. 360-0152-01 (Rev F0)
Sensor Temp C 23.2 stamp. All Sensor Data values are reset with a successful sensor Max Temp C 25.3 calibration. 11/20/07 14:35:10 Min Temp C 19.7 11/15/07 05:46:20 Sensidyne Document No. 360-0152-01 (Rev F0)
The Fault Current Delay review screen displays the number of seconds of each hour that the Fault Current will be active for an active fault. Fault Current Delay Sensor EOL Out Cur Track TOD Fail TOD End of Life Sensidyne Document No. 360-0152-01 (Rev F0)
TOD test, the results of the test (Pass/Fail), the Last 01/24/06 09:07 number of tests that have been performed by the cell, and the total amount Passed @ 74 PPM of testing time (in seconds). Tests 2 Sec Sensidyne Document No. 360-0152-01 (Rev F0)
Sensor interface, sensor, and communications board if Display Version 1.17 installed. Also the present Data Format the Display is using. Head Version 1.09 Sensor Version 1.04 Data Format MM/DD/YY No Comm Installed Sensidyne Document No. 360-0152-01 (Rev F0)
TOD cell functions, setting 4/20 mA Adjustment combustible sensor parameters, and setting a password. Adjust Date/Time Communication Setup TOD Mode Adjustment --more-- System Configuration --more-- TOD Mode Adjustment Sensor Adjustment Set Password Reset Defaults Set Transmitter Tag Sensidyne Document No. 360-0152-01 (Rev F0)
(Alarm 1, Alarm 2, Alarm 3), as well as the TWA Alarm. Instructions for using this menu are found in Section 4.2. Alarm Functions > Alarm 1 Alarm 2 Alarm 3 TWA Alarm Select OK to Change Sensidyne Document No. 360-0152-01 (Rev F0)
Page 89
Fault Functions are described in detail in Section 4.3. Fault Functions Fault Functions > Head Fail --more-- Missing Sensor Maintenance Mode Sensor Fail TOD Fail TOD End Of Life Sensor End Of Life Calibration Due Output Current Track Calibration Mode --more-- Sensidyne Document No. 360-0152-01 (Rev F0)
Page 90
Sensor Fail The Sensor Fail fault is activated when the output of the sensor shows that it is not operating properly. Sensor Fail Relays Assigned > Add Relay Delete Relay Disabled Adj Fault Current Sensidyne Document No. 360-0152-01 (Rev F0)
Page 91
Disabled Adj Fault Current Maintenance Mode The Calibration Mode fault is activated when the transmitter is put into the Maintenance Mode. Maintenance Mode Relays Assigned > Add Relay Delete Relay Disabled Adj Fault Current Sensidyne Document No. 360-0152-01 (Rev F0)
Page 92
It is based on time Calibration Due elapsed from previous calibration date. Relays Assigned > Adjust Cal Due Date Add Relay Delete Relay Enable Adj Fault Current Sensidyne Document No. 360-0152-01 (Rev F0)
& time, screens are presented in the following sequence: Month, Date, Year, Hour, Minute. Note: The month and date screens are reversed when > MM/DD/YY Format the “DD/MM/YY” format is selected. DD/MM/YY Format 04/26/06 16:33:37 Sensidyne Document No. 360-0152-01 (Rev F0)
HART Comm Modbus Comm No Comm Installed HART Comm Modbus Comm No Comm Installed No User Adjustments > Modbus Address Through this Interface No Communication Baud Rate Use Current Loop Board Installed Parity Stop Bits Sensidyne Document No. 360-0152-01 (Rev F0)
Factors are used depending on whether a Catalytic Bead Combustible or Infrared Combustible sensor is installed, and whether Methane or Propane is the calibration gas. Instructions for making sensor adjustments for combustible sensors can be found in Section 3.3. Sensidyne Document No. 360-0152-01 (Rev F0)
When you have entered the last digit of the new password select ACK to set the new password. Enter New Password 123450 6 Digits 0 - 9 Select << To Exit Sensidyne Document No. 360-0152-01 (Rev F0)
Selected location is at the left edge of the display selecting << one more time will cancel any changes to the Tag line. Selecting when the selected location is at the right edge of the display will advance to the next screen. Sensidyne Document No. 360-0152-01 (Rev F0)
The ▲key will increase the contrast. The ▼key will decrease the Adjust Screen Contrast contrast. The counter will show an increase in number to reflect the change in contrast. XXXXX Counts UP=Increase Down=Decrease ACK To Accept Tag Select << to Cancel Sensidyne Document No. 360-0152-01 (Rev F0)
SensAlert ASI – Advanced Safety Integrity Gas Transmitter 6 Product Numbers & Parts List 6.1 Transmitters Product Numbers & Parts List Sensidyne Document No. 360-0152-01 (Rev F0)
823-0243-22 ....SA+ O3 2ppm SENSOR 823-0245-21 ....SA+ ETO 10ppm SENSOR 823-0247-21 ....SA+ COCL2 1ppm PHOSGENE SENSOR 823-0249-51 ....SA+ IR ACETYLENE 0-50%LEL SENSOR For a complete list of available sensors and ranges contact the factory. Sensidyne Document No. 360-0152-01 (Rev F0)
Page 103
009824-17 ...... Sulfur Dioxide [SO2], 50 ppm in Nitrogen (58L) Recommended practice for determining calibration frequency. Sensidyne equipment is tested and calibrated prior to shipment. After installation and stabilization of the gas monitor, qualified personnel should verify calibration by applying zero and span gases. This procedure should be performed at commissioning, then repeated 30 and 60 days afterwards, with deviations in zero and span recorded.
7.5 Proof Interval Calibration of sensor - IR sensor 1 year interval, other sensors 6 months. Refer to Section 3.1 Zeroing The Sensor and Section 3.2 Span Calibration for instructions to perform these procedures. Sensidyne Document No. 360-0152-01 (Rev F0)
Intrinsic Safety ............. NEC and CEC Class I Div 1, Grps A, B, C, D; Class II Grps E, F, G; Class III T4 ATEX II 1 G Ex ia IIC T4 Ga CE compliant, RoHS exempt Sensidyne Document No. 360-0152-01 (Rev F0)
Make certain sensor head unit is properly connected to the power supply board. Sensor board or interface is defective. Contact Sensidyne for RMA. Transmitter remains with “APPLY GAS NOW” on display. Calibration gas is present. Verify gas is correct type for sensor calibration.
Page 109
Earth Ground at Transmitter and Controller must be the same. Connect conductor between Transmitter Safety Ground and Controller ground, or use galvanic isolation between transmitter and controller. Termination Resistance too high Verify Termination Resistance is 100 ohms or less. Sensidyne Document No. 360-0152-01 (Rev F0)
II 3 (1) G Ex nA nC [ia Ga] IIC T4 Gc • The most restrictive of Ta = -29°C to +75°C and Sensor limit FM09ATEX0064X Input Ratings: 12-30VDC 500mA 6W Output Ratings: 240VAC 3A 50/60Hz Sensidyne Document No. 360-0152-01 (Rev F0)
II 1 G Ex ia IIC T4 Ga • The most restrictive of Ta = -29°C to +75°C and Sensor limit FM09ATEX0064X IECEx FMG 12.0024X Ex ia IIC T4 Ga Ta = -29°C to +75°C Entity Parameters: Vmax (Ui) = 18VDC Imax (Ii) = 600mA Li = 0 Ci = 0.12uF Pi = 3.3W Sensidyne Document No. 360-0152-01 (Rev F0)
Please indicate if a price quotation is required before authorization of the repair cost, understanding that this invokes extra cost and handling delay. Sensidyne’s repair policy is to perform all needed repairs to restore the instrument to its full operating condition.
Need help?
Do you have a question about the SensAlert ASI S2 Series and is the answer not in the manual?
Questions and answers