Page 1
All rights, including rights created by patent D-65205 Wiesbaden grant or registration of a utility model or design, are Phone: +49 (0)611 97393-0 reserved. Heitronics GmbH reserves the right to make Fax: +49 (0)611 97393-26 modifications and improvements in their design without E-Mail: info@HEITRONICS.com...
Page 2
Please pay attention to the instruction given in MAINTENANCE AND CALIBRATION. HEITRONICS production meets the latest technical development thereby applying sophis- ticated components. Nevertheless, functional errors may occur in exceptional cases. The failure of an instrument may cause a measuring value being issued which seems to be correct, but is wrong.
Page 3
KT 19 II Series Designation of model: Diese Erklärung wird abgegeben durch This declaration is issued by HEITRONICS HEITRONICS Infrarotmesstechnik GmbH HEITRONICS HEITRONICS Kreuzberger Ring 40 65205 Wiesbaden, Germany Hiermit wird bestätigt, dass die Produkte gemäß den Richtlinien des Rates zur Angleichung der Rechtsvorschriften der Mitgliedsstaaten über die...
Page 4
PREFACE The remarkable features of the HEITRONICS devices are a design suitable for specific application as well as an uncomplicated operation. Nevertheless, it is recommended to read the entire instruction manual before putting the instruments into operation. Please observe the safety instructions as well as the particularly emphasized annotations when reading the pertinent instruction to the operation.
KT 19.61 II / KT 19.62 II ................3-21 KT 19.63 II ....................3-23 KT 19.69 II ....................3-25 KT 19.81 II / KT 19.82 II ................3-27 KT 19.83 II / KT 19.85 II ................3-29 CONTENTS KT 19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 17/03/18e...
Page 6
Check routines ..................5-55 Error detection via interface ............... 5-55 Error detection via display ..............5-55 Error detection via analog output ............5-56 Monitoring the instrument function during operation ......5-56 CONTENTS KT 19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 19/06/19e...
Page 7
Assignment for 12-pin connector ..............3-5 Assignment for 7-pin connector ..............3-6 "Temperature resolution" ................3-7ff “Communication" ..................5-3 Alarm contacts ..................5-15 Minimum temperatures ................6-1 WARRANTY CONDITIONS SERVICE ADDRESS CONTENTS KT 19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 19/06/19e...
Accessories ......................................................Code ..........................Further data ..................................................Hereby it is confirmed that the Infrared Radiation Thermometer specified above conforms to the data indicated in the specifications. Tester: Wiesbaden, DATA SHEET KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 04/01/15e...
For the application of the Infrared Radiation Thermometer under high ambient tem- perature, water-cooling, air purge fittings and vacuum tight lenses can be delivered as accessories. GENERAL KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 04/02/14e...
Thermometer type is also suitable for measuring thick foils. The Infrared Radiation Thermometer KT19.82 II is applied for the standard measur- ing range of 8 to 14 µm thereby guaranteeing a good signal resolution. GENERAL KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 05/06/18e...
300 °C onwards, the Infrared Radiation Thermometer type KT19.02 II is suitable for measuring temperatures from 200 °C onwards. Temperatures from 500 °C respectively 700 °C onwards are preferably measured by means of HEITRONICS Series CT18 or KT18RD. Infrared Radiation Thermometers for measuring P LASTICS These types of Infrared Radiation Thermometers are employed for the measuring of surface temperatures of plastics within a temperature range of 0 to 600 °C.
Page 12
KT19.69 II has been especially developed for determination of combustion tempera- tures from 400 °C onwards. Using KT19.41 II, objects can be measured through gases and flames from 400 °C onwards. GENERAL KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 05/06/18e...
Ø 3 mm @ 1 m distance 55.0 °C laser Focus cross lines adapted to field of view < 1 mW 650 nm 55.0 °C laser with centering circle in the focus distance TECHNICAL SPECIFICATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 15/03/18e...
Technical Information Protective Cooling Jacket The HEITRONICS Infrared Radiation Thermometers can be operated at an environmental temperature of up to 60 °C/ 70 °C without any additional coolants. If the environmental temperatures are higher than that mentioned above, cooling fittings can be used.
Page 16
25°C 1 ,2 1 ,4 1 ,6 Water flow rate in liter/min (1 l/min = 16 gal./h) Parameter: Water temperature 15/20/25 °C Max. recommended water pressure: 6 bar (87 psi) TECHNICAL SPECIFICATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 13/02/14e...
− data ground black brown alarm contact 2 (high) green (1): Hardware-Programming see data sheet page 1-1 (2): Configurable from Firmware 3.93 12-pin plug connector 12-pin socket-outlet TECHNICAL SPECIFICATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 14/01/15e...
► for current output 0 to 20 mA ≤ 550 Ohm ► for current output 4 to 20 mA ≤ 550 Ohm For the signal evaluation, differential inputs are recommended. TECHNICAL SPECIFICATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 14/01/15e...
Page 19
0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 20
** = Valeur > 20 °C. This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 21
0.35 1.10 10 s 0.05 0.10 0.20 0.75 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 22
** = Valeur > 20 °C. This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-10 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 23
0.15 0.35 1.20 0.05 0.15 0.35 1.20 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-11 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 24
** = Valeur > 20 °C. This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-12 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 25
1.75 10 s 0.05 0.15 0.35 1.20 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-13 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 26
** = Valeur > 20 °C. This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-14 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 27
0.10 0.25 0.75 0.05 0.05 0.10 0.40 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-15 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 28
** = Valeur > 20 °C. This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-16 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 29
0.75 10 s 0.05 0.05 0.15 0.50 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-17 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 30
** = Valeur > 20 °C. This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-18 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14def...
Page 31
0,15 0,35 10 s 0,05 0,25 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-19 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 32
This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-20 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 33
0.10 0.35 0.05 0.05 0.10 0.30 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-21 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 34
This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-22 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 35
10 s 0.05 0.05 0.10 0.30 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-23 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 36
This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-24 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 37
10 s 0.05 0.05 0.10 0.30 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-25 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 38
This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-26 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 39
0.05 0.15 0.05 0.05 0.05 0.10 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-27 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-28 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 41
0.05 0.10 0.05 0.10 0.20 0.65 ** = Wert / Value > 20 °C. Diese Einstellung wird nicht empfohlen. / This setting is not recommended. 3-29 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
Page 42
This setting is not recommended. Cette configuration n'est pas recommandée. Radiation temperature Température du radiateur Response time Temps de réponse Detector type Type de détecteur 3-30 TECHNISCHE DATEN / TECHNICAL SPECIFICATIONS KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 11/01/14 def...
3 onto the objective and fasten it, using the four screws 4. Attach cover 6 and move the cable protection 5 such that 5 fits into the cover 6. Use the screws 7 to tighten cover 6. START-UP KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 07/02/14e...
Page 44
• Screw Adapter B2 with flange DIN 2573 NW65 onto the mating flange • Connect Adapter B7 to Adapter B2 using the quick release fastener • Connect KT19 II with Adapter B4 to Adapter B7 using the quick release fastener The purging air is connected to adapter B7 through flexible tubing.
4.2.3 Connecting and Running the Nitrogen Purge (Option) On units of the KT19 II / TRT series equipped with nitrogen purge, there is a hose con- necter between the two plugs. The nitrogen is connected with a slight overpressure of max.
The measuring field marker also appears in focus. ATTENTION In this case, however, it only indicates the centre of the target diameter and not its size. START-UP KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 09/04/18e...
It is not possible to show the minimum target diameter with this option. 4.4.4 Controlling anti-glare filter in optical view finder Setting to continuous operation (IN) or automatic control by means of KT19 II. Setting FILter IN↵ AUTo...
/CTS ). Some RS232 drivers use DTR instead of RTS for the handshake. In this case CTS of Infrared Radiation Themometer KT19 II must be connected to DTR of the computer. The transfer parameters baud rate, number of data bits, parity, stop bits and delimiter are set by means of the operating keys at the Infrared Radiation Thermometer KT19 II.
Infrared Radiation Thermometer KT19 II identifies the ASCII control characters carriage return (0D) and line feed (0A) as end of instruction. As soon as the end of instruction has been recognized, KT19 II will reset the ready-to-receive state (RTS) until the command is evaluated. The answer of the Infrared Radiation Thermometer KT19 II is (0D).
ON or OFF temperature Object Emissivity temperature Laser Menue status ON or OFF Infrared Radiation Thermometer KT19 II – Rear view display 5.2.2. Function of keyboard Left display Right display function function To the left To the right Laser ON / OFF Confirm input ▲...
Operating Mode Viewfinder * LCD Contrast * Default Config Baud rate Cal.-Factor Standard Display Brightness Scanner Data bits Stop bits Parity Handshake * Optional, if included in the instrument OPERATION AND APPLICATION KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Changes within the menu require entering the code which is valid until exiting the menu. Code see DATA SHEET Example: Code setting: 0001 Select digit Confirm code Change value OPERATION AND APPLICATION KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Parameters of the instruments include model, size of detector, serial number and current temperature range. Press several times or keep Call up menu pressing to return to meas- urement Select menu item Confirm OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
In each case formula 1 = + + says that all received radiation is the sum of emitted radiation (S reflected portion () and transmitted portion (). = ( ) ∙ = ( ) ∙ = ( OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 56
Further options (see ty (if activated) Option Intern/Extern/Manual) Note: In setting of reflectiveness (transmis- Press several times or keep sion/reflection) emissivity may not be changed. pressing to return to measure- ment OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 57
Display and setting of reflection and transmission values (if Further options (see activated) Option Intern/Extern/Manual) Press several times or keep Code setting see 5.2.4 pressing to return to measure- ment OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 58
Display and setting of reflection Further options (see and transmission values (if Option Intern/Extern/Manual) activated) Press several times or keep pressing to return to measure- Code setting see 5.2.4 ment 5-10 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 59
The temperature is measured close to the detector. Press several times or keep Select menu item pressing to return to measure- ment Use selection Show internal temperature 5-11 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 60
Code setting see 5.2.4 Select digit Change value Change range Save value Press several times or keep pressing to return to measure- ment Same procedure valid for High 5-12 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 61
Use selection Change value Showing manual temperature Save value Press several times or keep Code setting see 5.2.4 pressing to return to measure- ment Change value 5-13 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Call up menu Select value Confirm Select menu item Press several times or keep Confirm pressing to return to measure- ment Change value Code setting see 5.2.4 5-14 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Commands for the configuration of memory functions and for the configuration of the analog output are to be found in HEITRONICS document “Universal Protocol”. Stored data can be adapted optimally to the corresponding application. At first memories may be reset by external activation of the digital input (trigger) and prepared for new materials or a new measuring cycle.
Page 64
Select menu item Confirm Further options Press several times or keep pressing to return to measure- ment Same procedure valid for Max, Min, Diff Select menu item Show selection 5-16 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 65
Time constant of the instrument may be set to auto-reset function only as large as the meas- urement is not distorted by integration! The extended configuration of the analog output allows switching the maximum or minimum value memory to the analog output. 5-17 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 66
Wert ändern Code setting see 5.2.4 Select menu item Confirm Select digit Change value Save value Press several times or keep pressing to return to measure- ment 5-18 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 67
Figure 1: Measurement Figure 2: Signal curve 5-19 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 68
This feature can be selected for only one of the memory types. Select digit Select menu item Change value Use selection Save value Change value Same procedure valid for Reset Mode Time 5-20 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 69
Minimum value memory as well as maximum value memory allow self-reset. Selection is possible via self- reset menu. Select menu item Select menu item Use selection Confirm Code setting see 5.2.4 Select menu item Confirm 5-21 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 70
The memory approaches the current value with the set decay rate. Select digit Select menu item Change value Confirm Save value Press several times or keep pressing to return to measure- ment 5-22 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
5.2.9. Alarms The Infrared Radiation Thermometer KT19 II provides two temperature alarms for temperature control or monitoring. 5.2.9.1. Change of alarm threshold Alarm thresholds Low and High may be configured in in menu ‚Alarm‘. Low alarm is activated when falling below alarm threshold, high alarm is activated when exceeding alarm threshold.
Page 72
Press several times or keep pressing to return to measure- ment Code setting see 5.2.4 Same procedure for High ON / OFF Select menu item Confirm 5-24 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 73
High Alarm (2) Pin C and L Option: Digital input instead of Low Alarm 1 Digital input Pin D and F High Alarm (2) Pin C and L 5-25 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 74
Alarm setting is displayed in the first row of the display. In case of alarm, the status is displayed as long as the warning is active. Hysteresis of the alarm is about 2 5-26 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
1200 °C. The same applies when turning on the laser. The anti-glare filter may be turned on perma- nently via this menu. Press several times or keep Call up menu pressing to return to measure- ment Select menu item Turn on filter resp. Automatic (Toggle) 5-27 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Use selection Code setting see 5.2.4 Press several times or keep pressing to return to measure- ment Select menu item Confirm Select menu item Change setting 5-28 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 77
Adopt selection Code setting see 5.2.4 Press several times or keep pressing to return to measure- ment Select menu item Confirm Select menu item Change setting 5-29 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 78
Adopt selection Code setting see 5.2.4 Press several times or keep pressing to return to measure- ment Select menu item Confirm Select menu item Change setting 5-30 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Call up menu Select menu item Use selection Code setting see 5.2.4 Select menu item Confirm Press several times or keep pressing to return to measure- ment 5-31 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
The analog input may be used for continuously set process values. The configuration allows reading in ambient temperature or emissivity for compensation purposes. If required the analog input may be deac- tivated. 5-32 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 81
For transmitting the measured value of a contact thermometer to a 0 to 10 V signal a recommended mod- ule is necessary See Options and Accessories . 5-33 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 82
The figure shows the connection of a signal transducer for PT100 in three-wire technique to an Infrared Radiation Thermometer KT19 II. signal transducer KT19 II 5-34 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 83
Press several times or keep Call up menu pressing to return to measure- ment Code setting see 5.2.4 Select menu item Confirm Select menu item Use selection 5-35 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Select menu item Adopt selection Code setting see 5.2.4 Select menu item Confirm Press several times or keep pressing to return to measure- ment Select menu item Confirm 5-36 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 85
Confirm Select digit Change value Wert speichern Press several times or keep pressing to return to measure- ment Select menu item Same procedure valid for Temp. High Confirm 5-37 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 86
Select menu item Adopt selection Code setting see 5.2.4 Select menu item Confirm Press several times or keep pressing to return to measure- ment Select menu item Confirm 5-38 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 87
Code setting see 5.2.4 Select menu item Confirm Press several times or keep pressing to return to measure- ment Same procedure valid for Min Select menu item Confirm 5-39 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 88
Code setting see 5.2.4 Select menu item Select digit Confirm Change value Save value Press several times or keep pressing to return to measure- ment Select menu item Confirm 5-40 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Example: 0 V resets the maximum value memory to actual value. Reset of the digital outputs If configured, the digital outputs can be reset to inactive state via digital input. Example: 0 V resets the digital output. 5-41 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 90
Confirm Code setting see 5.2.4 Select menu item Confirm Press several times or keep pressing to return to measure- ment Same procedure valid for further alternatives 5-42 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
5.2.16. Configuration of digital outputs The Infrared Radiation Thermometer KT19 II may be preset at factory for 1 or 2 digital outputs. Digital ouput 2 is only available alternatively to digital input. Both outputs allow the activation condition, reset con- dition, inverse logic or non-inverse logic.
Page 92
Note: Digital output 2 is available only if digital input is deactivated. Call up menu Select menu item Confirm Same procedure valid for digital output 2. Select menu item Confirm 5-44 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 93
Select menu item Adopt selection Confirm Code setting see 5.2.4 Press several times or keep pressing to return to measure- ment Same procedure valid for f.e. Lo/Hi Alarm 5-45 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 94
Select digit Select menu item Change value Confirm Save value Press several times or keep pressing to return to measure- ment Select menu item Use selection Change value 5-46 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Page 95
High to Low if the configured event occurs. Select menu item Select menu item Use selection Confirm Code setting see 5.2.4 Press several times or keep pressing to return to measure- ment 5-47 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
5.2.17. Configuring serial interface For connecting KT19 II to the PC, the configuration of the instrument interface needs to correspond to the configuration of the PC interface. This setting is changed in menu “serial interface”. Call up menu Select value...
Select digit Change value Save value Press several times or keep pressing to return to measure- ment Select menu item Confirm Change value Code setting see 5.2.4 5-49 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Scanner: 19200 Baud, 8 data bits, 1 stop bit, no parity, handshake ON. In standard mode the interface is adjusted to the preset settings of the KT19 II. Call up menu...
In order to improve the legibility of the menu the contrast of the graphic display may be adjusted. Call up menu Adjust contrast Confirm Select menu item Confirm 5-51 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
In order to improve the visibility of the temperature shown in the viewfinder, the brightness of the display may be adjusted. Call up menu Determine brightness Confirm Select menu item Confirm 5-52 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
Call up menu Select menu item Confirm Code setting see 5.2.4 Press several times or keep pressing to return to measure- ment Select menu item Confirm 5-53 OPERATION AND APPLICATION KT19 II 95582716 HEITRONICS Infrarot Messtechnik GmbH 20/06/19e...
If the Infrared Radiation Thermometer is communicating via serial interface and errors occur, they are au- tomatically output via the latter as text. If KT19 II interface is programmed for repetitive temperature output, an error message is output instead of a temperature value.
≤ 0.07 mA (refer to data sheet), exceeding the temperature range outputs ≥ 21 mA. Monitoring power consumption Infrared Radiation Thermometers of KT19 II Series draw a current of 90 mA and 150 mA at 24 V during operation. Values out of range indicate malfunction.
MAINTENANCE AND CALIBRATION General information The design of HEITRONICS Infrared Radiation Thermometers provides for a long period of service without the necessity of special maintenance. It is advisable to check the display accuracy approximately every three months. Measuring errors may arise, for instance, ow- ing to pollution of the lens.
Page 105
KT 19.69 II 580 to 600 KT 19.81 II 80 to 100 KT 19.82 II 80 to 100 KT 19.83 II 80 to 100 KT 19.85 II 80 to 100 MAINTENANCE AND CALIBRATION KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 04/02/14e...
Page 106
Spectral emissivity of Infrared Radiation Thermometers and transmission curves of various synthetic materials........7-5 Spectral emissivity ε, transmissivity ε and reflection ρ of glass ......7-6 Dimensions KT19 II ................... 7-7 Dimensions KT19II – with cooling and protective housing ....... 7-8 Target Diameter ....................7-9...
Fig. 10a Spectral emissivity of various materials Fig. 10b Total emissivity of some materials at 20 °C (Spectral range 8 ... 14 µm) ILLUSTRATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 06/02/14e...
Page 108
6 Copper, slightly oxidized 7 Graphite 8 Sheet iron, rolled 9 Aluminium, heavily oxidized 10 Cast iron, polished 11 Aluminium, technically pure T/°C b) Total emissivity of metals Fig. 11 Emissivity of metals ILLUSTRATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 06/02/14e...
Page 109
6.8 µm ± 0.15 µm Emissivity ε as function of the thickness d of various foils Within the wavelength range of 8.05 µm ± 0.15 µm Fig. 12 Emissivity of foils ILLUSTRATIONS KT19 II 95582717 HEITRONICS Infrarot Messtechnik GmbH 06/02/14e...
HEITRONICS is not responsible. Warranty does not include batteries. The device must be returned to HEITRONICS in the original packaging, carriage paid. Costs for taxes, fees and customs duties are to be paid by sender. Shipping damage is borne by the sender.
Page 124
For details about our regional representatives, please, refer to our website www.heitronics.com. Réseau des ventes Vous pouvez trouver les coordonnées de nos représentants régionaux sur notre site Web www.heitronics.eu. Organizacion de la venta Informaciónes referente a nuestros regionales colaboradores de venta se encuentran en nuestra pagina web www.heitronics.eu.
Need help?