Calibration; Preliminary Control Settings; Initial Starting Procedure - Goldstar OS-7020 Service Manual

Table of Contents

Advertisement

8-4. Preliminary control settings
Set the instrument controls as follows, when starting the calibration
procedures.
Controls
Settings
Controls
Settings
FOCUS
Midrange
TRACE
ROTATION As desired
- VERT -
V. POSITION
Mid r-ange , pushed
in
SWP VAR
Fully clockwise
V. VARIABLE
CAL, fully clockwise
INPUT COUPLING
GND.
- TRIG -
(AC-GND-DC)
VOLTS/DIV
5 mV/DIV
TRIG. COUPLE·
AC
V.MODE
CH1
LEVEL
Midrange,
Normal
CH2INV
· Normal, Pushed in
SLOPE
+
Normal
- HORZ -
H. POSITION
Midrange
MODE
AUTO
-
PULL X10 MAG
~or~al,. Pushed in
TIME/DIV
1 ms
TRIG. SOURCE
CH1
8-5.
!nitial starting procedure.
1.
Rotate the INTENsity control to the midrange and set the POWER
switch to ON.
2.
Wait a few
seconds for the cathode ray tube (CRT) to warm up.
A trace should appear on the CRT.
3.
If tz-acedteappeaz-s , increase (clockwise) the INTENsity control
setting
until the trace
is easily
observed, or roughly check/adjust
the DC balance to get a trace as same as
@ •
4.
Adjust the FOCUS control for the. best focused display.
5.
Readjust the POSITION controls if necessary, to center the trace.