Page 2
For example, choosing “0.1/cm” under Cell Constant would be B, “PT1000 RTD” for Temperature Compensation would be 1, “10 feet (3m) S856” under Cable Length would be FC, “Tinned Leads” under Cable Termination Connector would be 3, and “Sensorex-Branded” under Branding would be A.
Page 3
® Cell Constants 0.1cm and 1.0cm CS300-C CS300 with Cable and Tee CS300-B Calibration Follow calibration instructions in your conductivity transmitter or controller manual. Suggested calibration is at least two points. One point is in air (0uS) and the other is typically >75% of the selected range.
Page 4
CS300 Process Graphite Contacting Conductivity Sensor Sensor Installation FIG 1 Mechanical Glue (FC875/S) or thread (FC875/T) into your cooling tower water loop. Install sensor into the tee by aligning the notch of the sensor with the FC875 tab. See FIG 1.