Temperature Probe Failure
System Failure
WARNING:
Do not use the warmer if the system failure alarm is activated. Remove the unit from service
and refer to qualified personnel for repair.
If an electrical failure is detected, the system failure automatically shuts off the heater and triggers a
two tone audio alarm that can not be silenced.
Action: make note of the error message and remove unit from service.
Scale Alarms
1-20
Panda iRes Warmer
Screen
Screen
M1110737 006
Step
Description
1
This "Temp Probe Failure" alarm activates
when the difference in the two thermistors in
the skin temperature probe is greater than
0.5°C for over 6 minutes.
Action: replace probe.
Description
Step
1
Weight on scale above maximum
This alarm means that the weight on the
scale is above 8 kilograms (17.6 lb).
Action: check for other objects on scale.