Baby Cold-Check Temp Probe
Check Baby
1-18
Panda iRes Warmer
Screen
Screen
M1110737 006
Step
Description
1
Baby Mode
The "Baby Cold- Check Temp Probe"
alarm activates when the baby's skin
temperature is greater than 1.0°C below
the Set Temperature. Silencing the alarm
suspends the low priority audio alarm
for twelve (12) minutes. When the baby's
temperature is greater than 2.0°C below
the Set Temperature a high priority alarm
activates. Silencing the alarm suspends the
high priority audio alarm for five (5) minutes.
The low priority alarm setting can be
narrowed to 0.5°C using the "Setup" screen
and "Pt Temp Alarm" keys.
Manual Mode
The alarm can be activated in the Manual
Mode using the "Setup" and "Man Temp
Alarm" key. (See Chapter 3 for detailed
description).
Action: monitor the baby carefully and
check the attachment of the probe on the
baby's skin.
Description
Step
1
Baby Mode
The low priority "Check Baby" alarm
activates, and an intermittent tone sounds,
after the radiant heater has been running at
100% power for more than twelve minutes.
Manual Mode
The low priority "Check Baby" alarm
activates, and an intermittent tone sounds
after the warmer has been operating at
greater than 25% heater power* has been
exceeded for more than twelve minutes.
Action: check the attachment of the skin
probe to the baby and monitor the baby's
temperature closely.
*Default power setting is 25%, but this can be adjusted
on the Service Screen. Refer to the Service Manual.