INTRODUCTION
The Gilson model HM-380D Dead-Weight Direct Shear
Machine is a motorized unit for direct and direct/residu-
al shear testing of soil samples and uses a 10:1 dead-
weight beam loading system to apply vertical confining
pressures. These compact, self-contained units are
built for harsh laboratory environments and require lit-
tle floor space.
This Direct Shear Unit has a strain rate of 0.0001 –
0.3 in/min (0.0025 – 7.62 mm/min) with a maximum
shear displacement of 0.8in (20.3mm) and the travel
speed is set with the thumbwheels on the left side of
the unit. The residual shear force is determined by a
load cell while real time load settings and peak load are
displayed on the digital readout. Vertical load capacity
is 1,411lbs (640kg) and the maximum horizontal shear
force is 1,500lbf (6.67kN). Consolidation and shear
displacement are measured with Displacement Trans-
ducers and displayed on a digital readout.
Shear boxes and weight sets are required and sold
separately.
FEATURES
• Total shear capacity of 1,500lbf (6.67kN)
• Stepper motor controls strain rates to +/- 1% from
0.0001 – 0.3in/min (0.0025 – 7.62mm/min)
• Teflon coated anodized aluminum water chamber
with drainage port included.
• Meets requirements for ASTM D3080 and AASHTO
T 236
REQUIRED ACCESSORIES (purchased separately)
Direct shear boxes are sold separately and are avail-
able in a wide range of inch and metric sizes with round
or square shapes. See all options at www.globalgilson.
com and in the chart.
PHONE: 800-444-1508
740-548-7298
HM-380D Digital Direct Shear
Machine, Dead-Weight
P.O. Box 200, Lewis Center, Ohio 43035-0200
E-mail: customerservice@gilsonco.com
OPERATING MANUAL
Dead-Weight Direct Shear Machine with Linear Variable
Displacement Transducers, Four-channel Data Readout and
Shear Box shown (Shear Box sold separately)
Direct Shear Boxes & Accessories
Description
Shear Box
Round Direct Shear Boxes
50mm (1.97in) Diameter
HMA-733M
2.0in (50.8mm) Diameter
HMA-733
60mm (2.36in) Diameter
HMA-734M
2.42in (61.4mm) Diameter
HMA-734
2.5in (63.5mm) Diameter
HMA-735
100mm (3.94in) Diameter
HMA-737
4in (101.6mm) Diameter
HMA-736
Square Direct Shear Boxes
50mm (1.97in) Square
HMA-733MS
2.0in (50.8mm) Square
HMA-733S
60mm (2.36in) Square
HMA-734MS
2.42in (61.4mm) Square
HMA-734SA
2.5in (63.5mm) Square
HMA-735S
100mm (3.94in) Square
HMA-737S
4in (101.6mm) Square
HMA-736S
Product Web Page:
www.globalgilson.com
1
Porous
Cutters
Extruders
Stones
HMA-743M
HMA-763M
GSA-208
HMA-743
HMA-763
GSA-210
HMA-744M
HMA-764M
GSA-213
HMA-744
HMA-764
GSA-217
HMA-745
HMA-765
GSA-219
HMA-747
HMA-767
GSA-231
HMA-746
HMA-766
GSA-232
HMA-743MS
HMA-763MS
GSA-208S
HMA-743S
HMA-763S
GSA-210S
HMA-744MS
HMA-764MS
GSA-213S
HMA-744SA
HMA-764SA
GSA-217S
HMA-745S
HMA-765S
GSA-219S
HMA-747S
HMA-767S
GSA-231S
HMA-746S
HMA-766S
GSA-232S
Rev: 07/2020
FAX: 800-255-5314
740-548-5314

Advertisement

loading

Summary of Contents for Gilson HM-380D

  • Page 1 HM-380D Digital Direct Shear Machine, Dead-Weight INTRODUCTION The Gilson model HM-380D Dead-Weight Direct Shear Machine is a motorized unit for direct and direct/residu- al shear testing of soil samples and uses a 10:1 dead- weight beam loading system to apply vertical confining pressures.
  • Page 2 Weight Sets are required to apply the consolidation load and are available in both Pound and Kilogram sets. Pound Weight Sets Included Weight Sets Total Load at 10:1 0.852lb (1/8tsf) 1.704lb (1/4tsf) 3.409lb (1/2tsf) 6.818lb (1tsf) 13.635lb (2tsf) 27.270lb (4tsf) Model Total Mass Beam Ratio¹...
  • Page 3 Once the rate is set and the test is started, the switch gap between the cross arm and loading ball. Align setting is ignored until the stop switch is pressed or a the cross arm pin so it is above the loading ball. limit switch is tripped.
  • Page 4 8. At the test completion, switch the DIRECTION tog- gle to the right towards the HOME light. Press the stop button when Zero load is reached on the load cell, channel 1 of the readout box. 8A. If Residual Shear values are needed, these can be determined during this time.