Technical Specifications

Minimum Resolution (ppm Model)
Standard running mode:
TAC mode:
Tube mode:
Minimum Resolution (ppb Model)
Standard running mode:
TAC mode:
Tube mode:
Maximum Reading
Standard running mode:
Tube mode:
Response Time
Accuracy
Linearity
Intrinsically Safe Approvals
Battery Life
Li-ion:
Alkaline 3 x AA:
Lamps
Data Logging
Communication
Calibration
Unrivalled Gas
Detection.
up to 20,000 ppm or 20,000 mg/m
200 ppm or 639 mg/m
o
130 seconds at 20
Progressive indication of benzene breakthrough is displayed
in real time.
+/- 10% display reading +/- one-digit benzene
+/- 5% display reading +/-one digit
II 1G Ex ia IIC T4 Ga
Tamb = - 15C ≤Ta ≤+45C (with Lithium ion Battery Pack)
Tamb = - 15C ≤Ta ≤+45C (with Alkaline Battery Pack)
ITS09ATEX26890X IECEx ITS 10.0036X
3193491 conforms to UL Std. 913, 61010-1
Certified to CAN/CSA Std. C22.2 No. 61010-1
up to 24 hours
charge time 6.5 hours
typically 8.5 hours life
10.0 eV Krypton PID lamp
> 120,000 data log points including date and time stamp
Direct USB 1.1
2 - and 3-point calibration (via calibration kit accessory)
Page 70 of 73
TIGER Select Instrument User Manual V3.2
0.1 ppm
0.01 ppm
0.001 ppm
0.001 ppm
0.001 ppm
0.001 ppm
3
(gas dependent)
3
benzene
C (variable).
ionscience.com

Advertisement

Table of Contents