Unico PowerSpin C8604 Operation Manual

4 place horizontal spin centrifuge

Advertisement

Quick Links

PowerSpin™ C8604
4 Place Horizontal Spin Centrifuge
OPERATION MANUAL
182 Ridge Road, Suite E, Dayton, NJ 08810, USA
800-588-9776 Fax: 732-274-1151 www.unicosci.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the PowerSpin C8604 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Unico PowerSpin C8604

  • Page 1 PowerSpin™ C8604 4 Place Horizontal Spin Centrifuge OPERATION MANUAL 182 Ridge Road, Suite E, Dayton, NJ 08810, USA 800-588-9776 Fax: 732-274-1151 www.unicosci.com...
  • Page 2 • Power on/off switch located on back of unit Construction: The enclosure of the PowerSpin C8604 centrifuge is made of easy to clean and disinfect high impact plastic. The internal rotor chamber is made of stainless steel. A safety device is installed to prevent the lid from opening while the centrifuge is running.
  • Page 3 Unpack the shipping carton carefully and verify that the shipment is complete. It should include the following items: PowerSpin C8604 centrifuge 1 unit Power cord 1 pc.
  • Page 4 Caution: Make sure the voltage of the power source matches the voltage input of centrifuge. Control Panel and Key Pad Controls The PowerSpin C8604 centrifuge control panel appears as per the diagram below: RCF: xxxxx Status: Running START Spd: xxxx...
  • Page 5 Note : Make sure the samples are balanced in the rotor! Stop Button: Press to stop the centrifuge at anytime. Open Button: Press the Open button to open the lid. The Open button will only open the lid when the rotor is not spinning. Emergency Lid-opening: The lid features an electronically controlled safety system to prevent it from being opened accidentally while the centrifuge is spinning.
  • Page 6 Safety: Your PowerSpin C8604 centrifuge is a carefully designed and built instrument. Please take note of the following safety precautions. Lid with zero rpm lid latch (electronically operated): NEVER ATTEMPT TO OPEN THE LID UNTIL THE ROTOR HAS COME TO...
  • Page 7 OSHA, etc. Centrifuge Service: If you feel that your centrifuge needs factory service, please contact the UNICO customer service department for assistance. Simple problems may be resolved over the telephone. If the instrument needs to be returned to UNICO, an R.G.A. (return goods authorization) number will be issued.
  • Page 8 Caution DISCONNECT THE POWER CORD BEFORE REPLACING FUSE. Follow the procedures below to replace the fuse: a) Unplug the power cord. Turn the on/off switch to “off” b) Remove the power cord from the power inlet on the back of the centrifuge. c) Locate the fuse holder.
  • Page 9 Note: This warranty will not extend to any repairs or modifications made to the instrument by some party other than the manufacturer or a party authorized by Unico. Also, this warranty shall be in effect only upon the notice of the defect to the manufacturer or it’s authorized distributor, or dealer, within five (5) days after the occurrence of said defect No other warranty of any kind is made expresses or implied.
  • Page 10 PowerSpin MX C8606, C8604, C8624 and M24 Trouble Shooting Problem Cause Solution Error 1 Initialization failure: Main control board related. 1. Check all connectors. (Initialization failure) 1. Loose connection(s) 2. Replace TR1 transistor 2. Defective transistor (TR1) (p/n on main board BATA06-600B) 3.
  • Page 11 UNICO PowerSpin C8604 Centrifuge ™ Parts and Nomenclature Horizontal Rotor Rotor Spindle Power indicator light Emergency opening Manual Stop, Start override & Open hole keys RPM, RCF Speed and and timer Timer display Setting window keys...