5. Technical data

Power supply
Voltage
Frequency
Power consumption
Cable dimension
according to IEC
Sensor connection
Recommended cable
type
Controller
connection
Recommended cable
type
16
230/115 VAC ± 10 % or 24 VDC
50/60 Hz
Max. 2.5 W
2
0.2 - 4 mm
UL: 24-12 AWG
Screened
Cross-section: 0.25 - 1 mm
Max. cable length: 3 m (10 feet)
Screened, twisted-pair
2-wire, 4-20 mA
Cross-section: 0.26 - 1 mm
2
/ AWG: 24-18
2
/ AWG: 23-17