Page 1
Percent Weighing – Quick Reference Guide – AD Series Percent Weighing (Percent) Percent Weighing is used to compare the relative weight of an unknown item to that of a previously stored sample. An “E” indicator will appear on the screen to indicate the scale is in Percent Weighing mode. 1. To enter Percent Weighing Mode, press the “F” key. Select “percent” by pressing the “T” key. PERCENT 2. Place an empty weighing container on the pan. The container’s weight will be displayed. Wait for the weight to stabilize and press the “T” key to tare. 100.000 ------ 3. The scale will display the weight of the previously stored sample and the command “SAMMPLE” sequentially. To ...
Page 2
Percent Weighing – Quick Reference Guide – AD Series SAPL ON 100.000 Note: The reference sample must be equal to or greater than 100d. SAPL LO Attempting to set a sample lower than 100d will result in a “ ” error. READY 4. After setting the sample weight, the command “ ” will be displayed. Remove the sample and place the object on the pan. READY 5. The scale will express the weight of the object as a percentage of the ...