4. Snap the handpiece cord clips onto the handpiece cable to secure the tubing.
5. Install cassette onto the pump rotor on the drive console.
2.3
Operating Instructions
These shaver handpieces can only be operated using a footswitch. For the
operating instructions for these handpieces, please refer to the surgical
console instruction manual.
To control the suction flow rate, move the suction port valve towards the front
of the handpiece to increase flow and move the suction port valve towards the
back of the handpiece to restrict or stop flow. To attain maximum
performance while resecting tissue or bone, the suction port valve must be in
the fully open position.
2.4
Preoperative Functional Test
Prior to each use, perform the following preoperative functional test:
1. Insert a bur or blade into the handpiece. Gently pull on the cutting
accessory to ensure it is properly seated.
2. Verify that the console properly recognizes the handpiece.
3. Press the footswitch to ensure that it functions properly.
4. Run the handpiece for less than 5 seconds at a speed less than 1500 rpm
to observe any abnormal noises, vibrations or heat rise.
5. If any operating difficulties occur, return the handpiece for service.