Electrical fast
transients/burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF
immunity
IEC 61000-4-3
Conducted RF
immunity
ENGLISH
DC input +/- 2kV, 100
kHz PRF
N/A
N/A
30 A/m, 50 or 60 Hz
tested in X, Y and Z
axes.
10 V/m @ 2450 (spot
frequency) MHz
6Vrms
150 kHz - 80 Mhz
DC input +/- 2kV, 100
kHz PRF
30 A/m, 50 or 60 Hz
10 V/m @ 2450 (spot
frequency) MHz
6Vrms
150 kHz - 80 Mhz
should be at least
30%.
Note: Some ESD
events might cause
the unexpected reset
of the Physilog6
sensor device. If after
the unexpected
reset, the device
seems blocked,
please perform a
manual reset (see
section 6.1)
USB cable must be
the one provided by
the manufacturer
Power frequency
magnetic fields
should be at levels
characteristic of a
typical commercial
or hospital
environment.
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Physilog6
sensor device,
including the USB
cable, than the
recommended