VIDEO OTOSCOPE v4.0 20/05/2020 hearScope is a trademark of the hearX Group IP...
Page 2
Congratulations with the purchase of your hearScope smartphone and/or web app video otoscope. Please read this guide carefully to ensure the correct setup and safe operation. hearScope™ - Ground Floor, Building 2, Ashlea Gardens Office Park, 180 Garsfontein Road, Ashlea Gardens, Pretoria, 0081, South Africa For more information visit: www.hearxgroup.com...
WHAT IS HEARSCOPE hearScope consists of a handheld video-otoscope with a light source and ear tips of various sizes that connects to an Android smartphone, laptop or personal computer (PC). The device is supplied with a protective, reclosable case. The intended use of hearScope is to visually inspect the outer ear canal and tympanic membrane under magnification.
SETUP INSTRUCTIONS Download and open the hearScope application. Connect your hearScope to your smartphone (use the micro USB or USB-C adapter), or to your laptop/PC (plug in the USB cable in the USB port of the laptop/PC) and grant your smartphone, laptop or PC permission to access the...
Page 5
Please refer to the in-app tutorial for operating instructions. Open your hearScope application on your On the Home Page go to Help and then select Orientate your hearScope right side up by smartphone, laptop or PC. Tutorial. ensuring the text on the hearScope device...
Page 6
The light intensity can be adjusted by Point the hearScope at an object about 25mm Select an ear tip and fit it onto the hearScope turning the wheel attached to the cable, it from the front of the scope and adjust the...
Page 7
‘Done’. Images can be edited, Do not insert the hearScope too deep, as this your left hand over your head to the right ear. voice/text notes can be added and an image may cause damage to the eardrum.
Page 8
SETUP INSTRUCTIONS All images can be edited, as well as deleted. Saved images can be viewed by accessing the ‘Gallery’.
Page 9
The length of the tongue depressor is flexible effortlessly. This enables the use of hearScope™ as a dermascope. based on user preference. hearScope™ has LED lighting at the tip to use it as a laryngoscope and assists with throat inspections.
CONTRAINDICATIONS hearScope should not be used on individuals with visible obstructions of the outer ear canal. In individuals with fluid draining from the outer ear canal, hearScope should be used with caution, with strict infection control protocols in place. hearScope should not be used for any purpose other than visually inspecting the outer ear canal and tympanic membrane, as well as to inspect throats or skin.
Ensure that the hearScope is disconnected from the smartphone or laptop/PC when not in use. Unplug the cable from the smartphone or laptop/PC to turn off the light to preserve the LED lamp life. Store the hearScope in the provided protective case at all times to avoid damage. The device and ear tips...
Take care to ensure all components of the hearScope are dry before the next use. Warning: Do not use Chlorhexidine-based cleaning solutions on the hearScope as it may cause deafness. TROUBLESHOOTING For troubleshooting tips and answers to frequently asked questions, please visit www.hearxgroup.com/hearscope...
Faults on the hearScope must be reported to the Manufacturer within 60 days of discovering the fault for the warranty to be applicable. This warranty is non-transferable and does not cover damage due to negligence, accidents, abuse, misapplication, modification and repairs not made by the manufacturer or authorised service groups.