Page 1
GSI Pello ™ UseR MANUAl part Number D-0114341 Rev. C Setting The Clinical Standard www.grason-stadler.com Grason-Stadler, 10395 West 70th Street, Eden Prairie, Minnesota 55344 800-700-2282 • 952-278-4402 • fax 952-278-4401 • e-mail info@grason-stadler.com...
Indication for Use The GSI Pello is intended to be used for the identification and etiology of hearing loss in patients of any age. It is intended to be used by an audiologist, ENT, hearing healthcare professional, or trained technician in a hospital, clinic, healthcare facility or other suitable quiet environment as defined in ANSI S3.1 or...
Audiometric Standards The Pello is designed to meet or exceed the following standards: Audiometer Standard Requirements - Type 2 1. ANSI S3.6 (2010) Specification for Audiometers (Type 2) 2. IEC 60645-1 (2012) Electroacoustics - Audiological Equipment - Pure- Tone Audiometers Type 2 3.
To avoid the risk of electric shock, this equipment must only be connected to supply mains with protective earth. Do not block access to the power switch. Do not position the GSI Pello ME EQUIPMENT so that it is difficult to operate the disconnection device.
The CAUTION label identifies conditions or practices that could result in damage to the equipment. NOTE: Notes help identify areas of possible confusion and avoid potential problems during system operation. Status/Error Messages Please try another selection: Indicates an incorrect selection. This could include actions such as incompatible transducers, incompatible routing, or no calibration data stored for the selected transducers.
Elimination of Ambient Noise The GSI Pello may be installed in a single room environment or as part of a two room suite. Excessive noise in the test environment, such as that produced by conversation, office equipment, or printers, reduces test validity because it tends to mask the test signals.
Page 12
Maximum Ambient Noise Test Tone Freq. (Hz) 750 1000 1500 2000 3000 4000 6000 8000 Test Room level Max 29.0 17.5 14.5 16.5 21.5 21.5 23.0 28.5 29.5 33.0 38.5 dB SPL, Ears covered Max dB SPL, 23.0 13.5 Ears not covered Notes: Maximum permissible 1/3 octave band level.
Page 13
Sound Attenuation Sound Attenuation for Earphones per ISO 4869-1 Frequency Attenuation (Hz) DD45 with IP30 inserts DD450 HDA 300 MX41/AR or (dB) (dB) (dB) PH51 Cushion (dB) 33.5 14.5 12.5 34.5 12.7 34.5 22.5 1000 35.0 28.5 12.8 1250 1500 1600 2000 33.0...
ANY EQUIPMENT CONNECTED TO THE GSI INSTRUMENT AND USED IN THE PATIENT VICINITY MUST BE POWERED BY AN ISOLATED POWER SOURCE TO MAINTAIN THE ELECTRICAL SAFETY OF THE OVERALL SYSTEM.
Always turn OFF system power before connecting or disconnecting system components to help guard against personal injury. Warning - Electric Shock Hazards Do not open the case of the GSI Instrument. Do not remove any GSI instrument covers. Refer servicing to qualified personnel.
Follow all safety standards set by each place of employment. Shutdown Procedure To turn off the GSI Pello, use the power switch on the right side of the device. Recycling / Disposal Many local laws and regulations require special procedures to recycle or dispose...
Page 17
boards, electronic components, wiring and other elements of electronic devices. Follow all local laws and regulations for the proper disposal of batteries and any other parts of this system. Below is the contact address for proper return or disposal of electronic wastes relating to Grason-Stadler products in Europe and other localities.
Description Conforms to European Medical Device Directive 93/94/EEC. Classified under the Medical Device Directive (93/42/EEC) as a Class IIb device. GSI Part number and model Symbol for "SERIAL NUMBER." Return to Authorized Representative, Special disposal required. Medical Equipment Classified by Intertek Testing Services NA Inc.
Page 19
Symbol Description Patient response switch. Follow Instructions for Use. China RoHS compliance.
Audiometric Symbols The Pello can support different symbol sets to accommodate the conventions in different countries. The country symbol sets that are supported include: • Australia • China • Hong Kong • • The Pello Configuration Application (Config App) allows the selection of the desired symbol set.
Page 21
GSI Pello Symbol Sets Australia China Hong Kong (NR) masked (NR) (NR) masked (NR) Forehead (NR) Forehead masked (NR) (NR) masked (NR) SF Aided (NR) Cochlear (NR) (NR) (NR) Tinnitus (NR) (NR)
The GSI Pello continues the tradition of excellence in audiometry by maintaining the Grason-Stadler legacy of fast, efficient, and familiar navigation. The one- button, one-function front panel of the Pello is recognized worldwide as the Gold Standard of user-friendly design, allowing audiologists to test with confidence.
Unpacking Carefully remove the GSI Pello from its shipping container. If the instrument appears to have suffered any damage, notify the carrier immediately so that a proper claim can be made. Be certain to save all packing material so that the claim adjuster can inspect it as well.
Pello + GSI-AMTAS + Special Test + Speech Plus + High Frequency (DD450 8504261 in place of DD45) NOTE: Part numbers may change periodically. Please see the current GSI price/parts list for current part numbers. * Is considered applied part according to IEC/ES 60601-1...
Page 25
8508410 GSI-AMTAS + Speech Plus Automated Audiometry, Quick SIN, BKB- Upgrade Kit SIN, AutoScore/Play, AZ-Bio GSI –AMTAS + Special Test + 8508411 Automated Audiometry, Pediatric Noise, Speech Plus Upgrade Kit Remote Keyboard, ABLB, TEN Test, SISI, Tone Decay, Quick SIN, BKB-SIN,...
Chapter 3: Connectors, Controls and Indicators Rear Panel The connectors on the rear panel of the GSI Pello are shown in the following diagram. The label and jacks are visible by turning the instrument around on a flat, stable surface.
Page 27
Connection Description Graphic 6.35 mm stereo jack R10 Talkback Microphone NOTE: Microphone inputs are between .25 mV and 5 Input mV for a 0 dB reading on a VU indicator; the input impedance is 3,200 ohm. R11 Ext. A and B 3.5 mm stereo jack Input jacks for optional digital music player or CD player input...
Monitor speaker USB Port The Pello is equipped with four (4) USB A and one (1) USB B ports. It is possible to connect external devices such as mouse, keyboard, or external printer to be used with the audiometer. Additionally, a memory stick may be inserted into a USB port for updating software, adding additional sound files, license updates, printing to PDF or exporting diagnostic log files.
Conforms to European Medical Device Directive 93/42/EEC. Manufacture Date (year and month will be inserted below). China RoHS symbol for products compliance. B Patient Applied Part according to IEC 60601-1. GSI Part number and model Serial Number Return to authorized representative, special disposal required. Manufacturer.
Chapter 4: Front Panel Controls The controls on the front panel of the GSI Pello are shown below. Power The green LED, located in the upper right portion of the front panel, is illuminated when mains power is supplied to the GSI Pello. This indicates that the power switch is in the on position.
Channel 1 and Channel 2. The GSI Pello resumes the test status when the pushbutton is released. The light pipe around the Talk Forward button will be illuminated when enabled.
The printer type (HP or PDF) may be set up in the Configure, Set Up menu on the Pello or using the Config App Software. The HP color printer must be PCL 5E, PCL 3, or PCL 3 GUI compatible. If PDF is the selected printer, a memory stick or USB drive must be inserted into a USB ports on the Pello prior to printing test results.
Stimulus Channel 1 Tone — The Tone pushbutton allows the selection of a pure tone stimulus for air/bone conduction testing with the choice of four transducer types. Mic — The Mic pushbutton provides input capability from the test microphone for monitored live-voice testing with the choice of four transducer types. Int/Ext A, Int/Ext B —...
The selection of any stimulus will deselect a previously selected stimulus on the opposite channel if the stimuli are not compatible. Refer to the following table for the stimuli compatibilities listing: Valid Stimuli Combinations Channel 1 Stimulus Tone Ext A Ext B Channel 2 NB Noise...
Attenuators (HL Controls) Channel 1 and Channel 2 The GSI Pello contains two independent HL rotary controls for Channel 1 test signal and Channel 2 masking level control with a range of -10 dB HL to 120 dBHL. Maximum dB HL values apply to the mid-frequencies with earphones only.
Navigation Controls The four navigation buttons and the middle select button may be used to make selections from the on-screen menus as well as navigate through the internal .Wav files for speech testing. The center button when pressed enables the selection.
Patient - This button displays a screen that allows the examiner to create a new session, enter patient demographics, select a patient from the patient list, import a patient list from GSI Suite, select an examiner and transfer and delete one or all sessions.
Plus and High Frequency options are located in Chapter 9. Monitor The GSI Pello comes standard with an LCD display. The LCD is hinged to the Pello and is used to display all of the testing information from the instrument.
Page 41
Interlock – When interlock is active, a padlock icon will appear. Tracking – When tracking is selected, a railroad track icon will appear. Data Transfer – When there is an active connection between the Pello and an external computer, communication will be indicated by the black arrows.
Tone Test Type - Audiogram Tone Test - Audiogram Display Title Bar On the left side of the title bar, “Patient…” will appear if a list has been entered. The patient name, if selected, will be displayed. In the center of the title bar, the test type (Pure Tone) will be displayed.
Page 43
Channel 1 and Channel 2 The Channel 1 and 2 windows display the current output for each channel. The sound wave symbol indicates that a stimulus is being presented. This sound wave will be present as long as the present bar is depressed, will flash to indicate a pulsed stimulus, and will be steady if “interrupt”...
Page 44
Audiogram View Selecting the Audiogram viewing mode displays the audiometric data in graphic format. The user may determine the layout of the audiogram graphs (Right/Left, Left/Right, or combined into a single graph). Press and hold the Test Type Tone button for two seconds to change the graph view.
Page 45
The Aided menu has 3 options that toggle each time the button is selected. • Blank • Aided (HA) • Cochlear Implant (CI) When the box is HA or CI, the aided or cochlear implant symbol will appear on the audiogram. NOTE: When HA or CI is selected the transducer will automatically change to speakers as the selected transducer and FM as the signal type.
Tone Test Type - Status Tone Test Status Display The Status (table) display for the Tone Test Type displays the data in a tabular format. The columns indicate the frequency, dB HL level and the effective masking level (dB EM). The display contains the same elements as the...
audiogram display. You may navigate the list of frequencies using the mouse or the frequency keys. Speech Test Type - Status Speech Test Status Display Title Bar On the left side of the title bar, the patient name, if entered, will be displayed. In the center of the title bar, the test type (Speech) will be displayed.
Page 48
Channel 1 and Channel 2 The sound wave icon and the VU meter indicate when a stimulus is being presented. The sound wave icon and VU meter will remain active through the duration of the stimulus. The level of the stimulus will be displayed in the color of the ear that has been selected for each channel (red for right, blue for left and black for binaural).
Page 49
and the masking level. There are two tables and each table can hold up to 6 tests per ear. Words/Sentences for Presentation The lower part of the Speech display shows the words from the selected word list. The words on the list may be presented by selecting the word with the mouse, by using the navigation buttons on the device (and the Word Nav option from the Navigation Menu) to highlight the word and pressing the present button.
Page 50
The favorite’s list selection at the top of the word list dialog refers to a favorites list per test type that may be set up using the Config App. When the test type is selected, the first wordlist in the favorite’s list will automatically appear in the test screen.
Page 51
Ch2 Mic The Ch2 Mic button enables the microphone in channel 2 allowing for the presentation of the same speech stimuli via ‘live’ voice at different levels in each ear. Display Pure Tone Audiogram The audiogram checkbox enables the display of the pure tone air conduction audiogram of the test ear in place of half of the Speech test Results table.
Stenger Test Results In addition to entering comments, the dialog has the options for recording Stenger test results (both pure tone and speech). The Stenger buttons on the comments dialog may be toggled to indicate a positive or negative test result. Off indicates that the test was not performed.
The “More” test type button calls up a menu of special tests that are included with the Special Test and/or Speech Plus modules. If the Pello is not licensed for these options, no additional menu items appear when the More Test Type button is pressed.
Inspect all cords for fraying and damage. If there is any damage to any cord, do not use the Pello. If speech testing with recorded voice from an external source is to be performed, check that the CD or digital music player device is connected and operating properly.
Config App. Tone Test Type Button Pressing the Tone Test Type button prepares the Pello for pure tone air and bone conduction testing. Each selection on the blue Navigation Menu is specific to Pure Tone Testing. It is possible to utilize headphones ( DD45), insert earphones (IP30), bone vibrator (B81), High Frequency Headphones (D450) and Sound Field speakers from this test type.
Page 56
Integrated Word Files When Speech Test Type is selected, the Pello defaults to internal .Wav files or what is specified in the Config App. These digitized lists may be presented for consistent recorded speech testing. For manual presentation: •...
Routine Test Procedures The following procedures are in compliance with the current ANSI and ISO recommendations for Manual Pure Tone Threshold Audiometry. Patient Instructions Preparing the subject for test: 1. Put the subject at ease. 2. Make sure the subject understands the task. 3.
• Repeat the sections on Patient Familiarization and Threshold Determination for each tone setting in the following order: 1,000 Hz, 2,000 Hz, 4,000 Hz, 8,000 Hz. Retest 1,000 Hz followed by 500 Hz and 250 Hz. If there is a difference of 20 dB or greater between octaves, test the inter-octave frequencies, i.e.
Stenger test The Stenger Test is used to evaluate patients with unilateral hearing loss that are suspected of malingering. Speech stimuli or pure tones are presented 10 dB above threshold in the better ear and 10 dB below in the reported threshold in the ear exhibiting the hearing loss.
These settings are downloaded from the application software on the PC to the Pello. It is recommended that a copy of the custom configuration is saved as a back-up. This will allow the custom configuration to be loaded quickly onto multiple Pello audiometers.
Log Tab In the event of a repeatable error, the log window allows the examiner to upload or email a file from the Pello to the computer. This file “retraces your steps” (button pushes) for the purposes of troubleshooting. Audiometry...
The GSI factory does not include sample norm values to be used on the Speech Audiogram screen. Each facility should enter its own values, if desired. If Display on the Pello box is checked, the normative curves will appear on the Speech Audiogram screen based on the transducer being used.
More information may be found on the Software and Manuals USB flash drive in the GSI Instrument Services folder that was included in the original shipment of the Pello or contact your GSI representative. GSI Suite GSI Suite Audiometric Data Management software (Rev.
With extended use, all transducer cords tend to fray internally at the connectors. To evaluate the cord status, turn on the GSI Pello. Set the HL to a comfortably audible level. Place the transducer on your head. Activate both Interrupt buttons.
and should be replaced. In any case, the audiometer should be scheduled for immediate maintenance. Speech Level Check To check the speech level with recorded speech, select the Speech test type button. Place the earphones on a person with normal hearing and present a word list at 40 dB.
Cleaning and Disinfecting Agents According to the recommendations from the CDC, audiometric equipment is considered to be non-critical medical equipment and typically requires cleansing followed by low to intermediate level disinfecting, depending on the nature of the contamination. Cleaning should be done with a mild soapy detergent (such as dishwashing liquid) and a damp cloth or an Endozime Sponge followed by an application of EPA-registered hospital disinfectant.
The Range selection determines the displayed audiogram frequency range and available test frequencies. It will appear in the navigation menu if the High Frequency license is enabled on the Pello. The three audiogram options are Standard, High and Full range: •...
Page 68
Single Hz Frequency Resolution Tone Test Status Display – Single Hz Resolution The right and left outlined areas show the list navigation icons. The top (darker icon) moves +/- 1000 Hz and the other arrow moves to the next/previous page. High Frequency Test Procedure High frequency testing (above 8000 Hz) may be completed by using the Range Selection button on the Navigation Menu to select the high frequency range...
Remote Keyboard The Pello may be controlled with an external keyboard. Many of the operations of the front panel keys on the instrument may be performed using keyboard shortcuts. This functionality is included with the Special Test license. The following table shows the mapping of the keyboard keys to the instrument.
Page 70
TEN Test and Tone Decay. Use the on-board navigation buttons or an external mouse to select the desired special test. If the Pello is not licensed for these options, no special test menu items appear when the More Test Type is pressed.
ABLB In ABLB (Alternate Binaural Loudness Balance) test, the tone is presented alternately between the two ears. The level of the tone stays the same in one ear (i.e. fixed ear) and is varied up / down in the other ear (i.e. variable ear). The top section ABLB test display has common elements found on the previously described screens.
Page 72
• Select the ear to serve as the reference ear, typically the ear with the better hearing sensitivity. This ear will receive the tone at a fixed hearing level. • Select ABLB from the More Test Menu. • Set the level of the tone for each channel to 20 dB above the threshold of each corresponding ear.
The Navigation Menu has an option for the dB Step (continuous HL level) and an option for the SISI step (level increment). The results are not transferred to GSI Suite via the data transfer and therefore should be entered as a comment.
Page 74
Test Procedure - SISI (Short Increment Sensitivity Index) Test The SISI test is used to detect small level changes in a steady-state signal in patients with disorders of the cochlea. The SISI tests a patient’s ability to detect a 1 dB level change in a pure tone stimulus presented at 20 dB SL. A SISI consists of 20 target increments (200 msec at 1, 2, or 5 dB) presented every 5 seconds and can be completed for a number of frequencies.
TEN Test The TEN test involves measuring the threshold for detecting a sinusoidal tone presented in a special background noise called “threshold-equalising noise” (TEN). The methods used to conduct the test are similar to those used for masking in conventional pure-tone audiometry, except that the signal threshold is measured in the presence of ipsilateral, continuous background noise and a 2-dB final step size is used to measure thresholds.
Page 76
Test Procedure - TEN Test Purpose of the TEN Test is to identify cochlear dead regions. This is useful for several purposes including the following: • Counseling about the benefit of hearing aids. • Assisting in hearing aid selection or cochlear implant candidacy. •...
Navigation Menu will stop the timer after the defined number of minutes is reached on the timer. The results are not transferred to GSI Suite via the data transfer and therefore should be entered as a comment.
Page 78
Test Procedure – Tone Decay Carhart Tone Decay Test (1957) Patients with retro cochlear pathology of the eighth nerve exhibit a rapid “abnormal auditory adaption” or “temporary threshold drift” in response to a continuous pure tone presentation. Presentation Level • Establish the patient’s hearing threshold for the test ear using earphones or insert phones using a pulsed tone.
Speech Plus license that include: BKB-SIN and QuickSIN. Use the on- board navigation buttons or an external mouse to select the desired speech test. These tests would not appear in the More menu unless the Pello is licensed for Speech Plus.
BKB-SIN The BKB-SIN is a speech-in-noise test that uses BKB (Bamford-Kowal-Bench) sentences, recorded in four-talker babble. The BKB-SIN can be used to estimate SNR loss in children and adults for whom the QuickSIN test is too difficult. The BKB-SIN display has the Title bar and the Channel 1 and 2 Output sections that are similar to what has been described for the speech displays.
Page 81
BKB-SIN Test Results The BKB-SIN Test Results Table displays the information for the tests that have been stored. The data is separated by ear and group. The results include the SNR 50 and the SNR Loss. For details on the scoring see the BKB-SIN manual.
Page 82
Test Procedure- BKB-SIN For a detailed description of the BKB-SIN test the user is referred to the BKB- SIN manual provided on the Pello USB flash drive. The BKB-SIN Test uses the Bamford-Kowal-Bench sentences (Bench and Bamford, 1979; Bench, Kowal and Bamford, 1979) spoken by a male talker in four-talker babble (Auditec of St.
Page 83
Test Instructions Child “You will hear a man talking to you through the earphones (or loudspeaker). He is going to say “Ready” and then he'll say a sentence. Repeat the sentence the man says. You will hear other talkers in the background. Don't pay any attention to them;...
QuickSIN The QuickSIN is a speech-in-noise test that quickly measures the ability to understand speech in noise. The QuickSIN is comprised of sentences recorded in four-talker babble. The QuickSIN display has the Title bar and the Channel 1 and 2 Output sections that are similar to what has been described for the speech displays.
Page 85
QuickSIN Test Results The QuickSIN Test Results Table displays the information for the tests that have been stored. The data is separated by ear and group. The results include the SNR 50 and the SNR Loss. For details on the scoring see the QuickSIN manual. QuickSIN Sentences and Score The lower section of the display contains the QuickSIN sentences.
Page 86
applied clinically in a variety of ways by adjusting the presentation level or the presentation mode. Test Procedure - QuickSIN The primary complaint of hearing-impaired persons is difficulty in background noise. The measurement of SNR loss (signal-to-noise ratio loss) is important because speech understanding in noise cannot be reliably predicted from the pure tone audiogram (Killion &...
• Select additional lists for testing if necessary • To interpret the SNR loss score see table below. DEGREE OF SNR EXPECTED IMPROVEMENT WITH DIRECTIONAL LOSS LOSS 0-3 dB Normal / near normal May hear better than normals hear in noise 3-7 dB Mild SNR loss May hear almost as well as normals hear in noise...
Appendix 1: Specifications Dimensions and W x D x H: 14.8 inches x 10.5 inches x 13.8 inches (LCD raised) Weight 37.5 cm x 26.7 cm x 35.1 cm Height with LCD lowered: 4.0 inches 10.2 cm Weight: 8.18 pounds 3.7 kg Shipping Weight: 20 pounds...
Page 89
Paired Inserts: -10 dB HL to 95 dB HL Channel 2 Masking Level Range Narrowband Noise (Calibrated in effective masking) Maximum dB HL is 15 dB below tone Speech Noise (calibrated in effective masking): Air Conduction -10 dB HL to 95 dB HL (DD45) Bone Conduction -10 dB HL to 50 dB HL (mastoid) -10 dB HL to 40 dB HL (forehead) Sound Field:...
Page 90
obtain the List Pair score. Results are compared to normative data to obtain the SNR Loss. Special Tests (User Defined) Lombard test Pure Tone Stenger Speech Stenger Communications and Talk Forward: Permits the tester to speak through the test microphone into the Monitoring selected transducer at approximately the level set by the front panel controls.
Appendix 2: Calibration Reference & Maximum Levels The Pello is supplied from the factory calibrated for the transducers that were purchased with it. The exception is the speakers, as those must be calibrated in the environment where they will be used. The calibration data supplied from the factory is only valid for GSI supplied transducers and cannot be applied to non- GSI supplied transducers.
Earphones - Pure Tone max HL Transducer DD45 DD45 TDH50 TDH50 HDA200/DD450 HDA300 10 Ω 10 Ω 60 Ω 60 Ω 23 Ω 23 Ω Impedance Coupler 318-3 318-1 318-3 318-1 318-1 318-1 Signal Max HL Max HL Max HL Max HL Max HL Max HL...
Bone Vibrators - Pure Tone max HL Transducer 50 Ω 50 Ω 10 Ω 10 Ω Impedance Coupler Mastoid Forehead Mastoid Forehead Signal Max HL Max HL Max HL Max HL Tone 125 Hz Tone 160 Hz Tone 200 Hz Tone 250 Hz Tone 315 Hz Tone 400 Hz...
Bone Vibrators - NB noise max HL Transducer 50 Ω 50 Ω 10 Ω 10 Ω Impedance Coupler Mastoid Forehead Mastoid Forehead Max HL Max HL Max HL Max HL NB 125 Hz NB 160 Hz NB 200 Hz NB 250 Hz NB 315 Hz NB 400 Hz NB 500 Hz...
Free Field Speakers – ANSI RETSPL and Max HL ANSI Free Field ANSI S3.6-2010 Max HL Binaural Basic GSI High GSI High Performance ISO 389-7 Speakers Performance Speakers with external Speakers Amplifier 0° 45° 90° 45° 45° 45° RETSPL RETSPL...
Page 112
Appendix 3: PIN Assignments Socket Connector Pin 1 Pin 2 Pin 3 Mains Live Neutral Earth Headphone x 2; Insert x 2; Bone Ground Signal Talkback; Patient Response Ground DC Bias Signal Ext. CD Ground Talk Forward Ground DC Bias Signal Monitor Ground...
Install and operate the GSI Pello according to the EMC information presented on this page and the next 4 pages. The GSI Pello has been tested for EMC emissions and immunity as a standalone instrument. Do not use the GSI Pello adjacent to or stacked with other electronic equipment.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The GSI Pello is intended for use in the electromagnetic environment specified below. The customer or the user of the GSI Pello should assure that it is used in such an environment. Emissions Test...
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The GSI Pello is intended for use in the electromagnetic environment specified below. The customer or the user of the Pello should assure that it is used in such an environment.
Note: UT is the a.c. mains voltage prior to application of the test level. Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The GSI Pello is intended for use in the electromagnetic environment specified below. The customer or the user of the Pello should assure that it is used in such an environment.
Page 117
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Pello is used exceeds the applicable RF compliance level above, the Pello should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Pello.
Appendix 4: Reference Materials Moore, B.C.J. (2004). Dead regions in the Cochlea: Conceptual Foundations, Diagnosis, and Clinical Applications. J Ear and Hearing 2004; 25; 98-116. USA. Lippincott Williams & Wilkins. Moore, B.C.J., Glasberg, B.R., Stone, M.A. (2004). New Version of the TEN Test With Calibrations in dB HL. J Ear and Hearing 2004;...
Need help?
Do you have a question about the PELLO and is the answer not in the manual?
Questions and answers
Can I connect the device with phone or audio device
Yes, the GSI Pello can connect to a digital music player by using a calibration track to adjust the volume level.
This answer is automatically generated
Can I import my logo to the audiogram and then print?