Summary of Contents for PAS Systems International Alcovisor MARK V

  • Page 1 MARK V Alcovisor®  Advanced Settings Manual  PAS Systems International, Inc. 215 Southport Dr. Suite 400 Morrisville, NC 27560 www.pasintl.com (800) 660-SNIF · (540) 372-3431 Copyright© 2010 by PAS Systems International Inc., All Rights Reserved...
  • Page 3 Introduction This manual contains the passwords and instructions on operating the advanced settings of the MARK V along with the proper procedures to be used for instrument calibration. Passwords Default Passwords: • Calibration: 119119 • Advanced Settings: 112112 The default passwords for advanced settings and calibration can be changed. To change the passwords, please refer to the “Password”...
  • Page 4 Advanced Settings To access the Advanced Settings menu: 1. Choose the Settings Icon 2. Select “Advanced” 3. Enter the password 112112 Advanced Settings: 1. Blow: This icon under advanced settings allows the user to change the blow time (in seconds) and blow pressure (1 to 6, with 6 being the hardest). These are set to optimal levels already and it is not recommended to change them.
  • Page 5 PAS recommends all MARK V units be calibrated annually or whenever an accuracy test shows that BAC readings are no longer within tolerance limits. Because precise calibration is crucial, it can be performed only by PAS Systems International or by persons who have been properly trained for alcohol detector calibration. (Calibration of the MARK V must be performed using a NHTSA-approved wet bath simulator or certified dry gas standard.) Successful completion of calibration should be recorded...
  • Page 6 More examples are in the following table: ‰ & g/L Mg/100mL 0.20 20.0 0.02 0.50 50.0 0.05 0.80 80.0 0.080 1.00 100.0 0.10 1.50 150.0 0.15 4. Select to confirm. 5. The next screen to appear is the calibration screen. The first line shown is the concentration of the calibration solution/cylinder.
  • Page 7 Maintenance Repairs of the MARK V may only be performed by PAS Systems International, Inc. or an authorized service technician. Only original MARK V parts may be used. To keep the instrument clean, periodically use a mild disinfectant and a soft cloth on the outside of the case.
  • Page 8 MARK V Alcovisor® PAS Systems International, Inc. 215 Southport Dr Suite 400 Morrisville, NC 27560 http://www.pasintl.com Tel: (540) 372-3431 800-660-SNIF Printed 09/13...