11. Troubleshooting MRC LTD. are greatly appreciated. 12. Warranty MRC Incubator Shaker Series is model designed for shaking and heat samples according 13. Packing List to the research and experiment needs in modern biology. State-of-the-art embedded microprocessor system is utilized in the equipment for its powerful data processing ability, outstanding stability, excellent interference resistance, and high precision control ability of the temperature and frequency of the instrument.
1. Unpacking Guide 1. Unpacking Guide 09. Please use a power supply specified on the nameplate on the right side of the Note instrument. 10. The maximum load of an individual unit is 15 kg. When three units are stacked, the 01.
2. Overview 2. Overview 2.2 Features 2.1 Description LOM-7450 are a stackable incubator shaker models with cooling function. The instrument ● Large view window Minimizes the need to open the lid. can be used individually or stacked on top of each other to save floor space. Flip-down ●...
Page 5
2. Overview 2. Overview 2.3 Specifications Illuminance Levels for models with lighting function: Illuminance (Lx) Illuminance Level Natural Light in Chamber LOM-7450 LOM-7450-CO2 LOM-7450-L Model 3000 Temperature 4℃~60℃ Settings Range 6000 Temperature 9000 Ambient temperature minus 15 °C to 60 °C (lowest 4 °C) Control Range 11000 Environmental Condition...
3. Preparation 4. Safety Instructions 3.1 Placement Conditions The instrument must be placed on a firm and level surface. The total weight of the Please make sure to follow the safety instructions! instrument when fully loaded must be taken into consideration. There must be suffient space left around the instrument.
5. Installation 5. Installation 3. Adjust the secondary flow control valve to point the pointer to 0.1 Note – 0.2MPa (as shown in figure 2). It may cause the pipe connect the Lifting the instrument by hand should be prohibited. Please use a lift or a forklift or other incubator and the inlet port to fall off and the leak of the CO gas if the lifting apparatus to raise or move the instrument.
6. Operational Instruction 7. Structure ● ● The instrument should be palced in a clean and dry environment. Any heat source or Product Front View heat dissipation outlet and direct sunshine should be avoided. ● Adjust the instrument footings to level the instrument and make sure the instrument can operate smoothly.
Turn on the power switch at the right side of the instrument after the instrument is properly placed. “Thanks for choosing MRC LTD. products” will be shown on the display window. The function selection menu will be displayed after 2 seconds.
Page 10
9. Settings 9. Settings The instrument will enter the next segment of program to run. When the cursor is at “Stop and Return” :37.0℃ Temp. Press the “ ” button to stop the operation and return to the main interface. :37.0℃...
Page 11
9. Settings 9. Settings Press the “ ” button to choose a function and enter the “ ” button to enter the 2. Choose “Programming” to enter the program page. corresponding operation page. Set Programs: a) Basic settings Total(1-6): 4 Main Interface Speeding Alarm: ON/OFF...
Page 12
9. Settings 9. Settings Press the “ ” button to choose the parameter that needs to be modified; press the Press the “ ” button to choose the parameter that needs to be modified or “Main “ ” or “ ”...
10. Alarm System 9. Settings the temperature parameters may be inaccurate and other unpredictable problems Alarm prompt: such as inprecise temperature control or temperature out of control may occur ● Overspeed alarm interface during normal operation. d) CO Settings Speeding Alarm Press the “...
10. Alarm System 11. Troubleshooting ● concentration high alarm (when the CO concentration is 1% higher than the set Troubleshooting value) Error Cause Solution LCD Display Sound : 06.0% Speed difference more than 5 rpm Contact local sales/service Humidity: Overspeed Intermittent- Contact local sales/service Electrical malfunction...
Page 15
11. Troubleshooting Instructions Plug not connected No power Power outlet out of power Code Power switch off Display window Change to a new fuse with the same specification off Platform not shaking Burned fuse Position: at the powr Position: inside the outlet electrical cabinet Press the confirm button to recover display...
Packing List Name Note Stackable Incubator Shaker Power Cord Liquid Collection Tray AC250V/10A Fuse Cross Screwdriver Platform handle backup M8 Star Nut User Manual Accessories Certificate Wrench For Model LOM-7450-CO2 only Filter For Model LOM-7450-CO2 only Filter Protection Package For Model LOM-7450-CO2 only Gas Pressure Regulator For Model LOM-7450-CO2 only Stainless Steel Hood...