InSight V-IA Veterinary Immunoassay Analyser uses Immunofluorescence technology to provide accurate, quantitative laboratory results. For in vitro diagnostic use only. For veterinary use only. InSight V-IA Veterinary Immunoassay Analyser is suitable for use in veterinary laboratories. Technical Specifications I.3.1 Main Parameters ◆...
I.3.2 Performance Specifications ◆ Repeatability: CV≤10% ◆ Stability: σ≤±8% ◆ Linear Correlation: r≥0.97 ◆ Accuracy: Δn≤±15% Contents Accessories Quantity Remark Power Adaptor Included Instruction for Use Included Ethernet Cable Included...
2. Check the analyser and accessories to see if they are in good condition. Notice: If there are any problems, please contact Woodley Equipment Company. III.1.2 Analyser placement 1) The analyser should be placed in a clean and ventilated room with temperature between 10°C ~ 30°C, relative humidity of less than 70%, away from direct sunlight.
6) Select standard test or instant test (refer to Section IV.2). 7) Place the test cartridge with specimen (follow cartridge operating instructions) onto the cartridge holder and run prepared test. Notice: ⚫ Do not touch the cartridge holder when it’s moving. ⚫...
IV Software Introduction IV.1 Main Interface Figure 4.1 As shown in Figure 4.1, there is a home [Menu] key at the bottom of all screens. Click the [Menu] key and the screen in Figure 4.1 will display. From left to right, the screen will display [Test], [Batch Testing], [History Record], [Parameter] and [Setting].
IV.2 Testing Interface Figure 4.2.1 1. Click [Test] option on the main home screen and Figure 4.2.1 will display. 2. Insert ID chip in the ID Port on the side of the analyser and select “Read ID Card” before using a new lot of test cartridges. 3.
IV.3 Batch Testing Figure 4.3.1 1. The screen for batch testing is shown in Figure 4.3.1. The user can select the sample type and test item and add or delete the item to be tested. 2. Select the test item to determine the time that is displayed in the interface [Time]. 3.
1. In the [Results] screen, users can view previous test results. 2. After each test has completed, the system will automatically save the results to the analyser memory. 3. Adjust the dates to search for a sample. Select [OK]. 4. Select [Upload] in Figure 4.4.1 to upload the selected records or all records to the PNS.
Page 14
Figure 4.6.5 1. In the [Setting] screen, Institution Information, Test Setting, LIS Setting, System Setting and Software Version can be viewed. 2. In [Institution Information] of Figure 4.6.1, the user can add the clinic name and address. 3. In Figure 4.6.2, the sample code, sample ID length and alignment can be set in [Test Setting] screen.
Quality Control Quality control can be carried out by testing the InSight V-IA quality control cartridges. Use InSight V-IA Veterinary Immunoassay Analyser to determine the concentration of the test cartridges. There are 3 control level cartridges – low, mid and high.
Page 16
Notices: ⚫ Turn off the power and unplug the analyser before moving it. ⚫ When moving the analyser, try to avoid vibration. ⚫ Desktops supporting the analyser should be able to hold at least 2.5kg. ⚫ The analyser should be placed carefully, with at least 5cm space all around to ensure good air circulation.
Page 17
Woodley Equipment Company. ⚫ Other than service personnel from Woodley Equipment Company and service personnel authorised by Woodley Equipment Company, other users are not permitted to remove, modify or repair the analyser. Any violation will invalidate the analyser warranty.
Page 18
Power frequency GB/T 17626.8 3A/m,50/60Hz magnetic field 1 cycle 0%; Voltage sag, 5 cycle 40%; GB/T 17626.11 interrupt 25 cycle 70%; 250 cycle 5% Performance judgement: A. Normal performance within the standard limiting value. B. Function or performance is temporarily reduced or lost but can be self-recovered. C.
Page 19
VI.3.5 Limitation Requirements for Toxic and Hazardous Substances This analyser meets the limitation requirements of toxic and hazardous substances in SJ/T11363-2006 Regulation. Table 3 Classification of Electronic Information Products Classification Definition homogeneous materials constitute electronic information EIP-A products EIP-B Metal coating of all parts of electronic information products Small parts or materials in electronic information products which cannot be further split in the existing conditions, EIP-C...
Let it reset itself and repeat Abnormal sound during stuck the test. testing Please contact Woodley Mechanical motion failure Equipment Company. Power interruption Restart the analyser and retest. Analyser stops during testing Communication failure Restart the analyser and retest. Please contact Woodley Equipment Company.
/ new cartridge. contaminated Barcode not The analyser recognised Analyser cannot read the Contact Woodley Equipment Technical Support barcode The reagent cartridge is not Make sure the reagent cartridge is inserted into Operation inserted to the the innermost contact slot...
Page 22
Verify that the sample type is correct. The reagent If the reagent cartridge or sample is taken out of Operation cartridge is the refrigerator, please leave to warm to room contaminated temperature before testing. Confirm the buffer tube is from the correct kit. Wrong buffer tube Operation Check buffer volume is correct.
Page 23
Ensure thermal printer paper used and installed Install paper paper used correctly Analyser Printer failure Contact Woodley Equipment Technical Support Data was not Click √ the square inside the box to the left of the Please select Operation selected when data...
Page 24
Contact Woodley Equipment Technical Support Analyser QC expired Repeat with fresh QC QC stored Check QC storage conditions meet Analyser QC out of range incorrectly requirements. Repeat test with fresh QC. Stop using the analyser and contact Woodley Analyser Analyser issue Equipment Technical Support...
Micro SD card Transportation Conditions VIX.1 Transportation 1. The InSight V-IA should be transported in the original packaging. 2. Avoid severe vibration during loading and transportation. 3. Keep away from damp. 4. Do not transport with flammable and corrosive substances.
Page 26
Old Station Park Buildings St. John Street Horwich Bolton BL6 7NY, UK Tel: +44 (0) 1204 669033 Fax: +44 (0) 1204 669034 Email: sales@woodleyequipment.com Web: www.woodleyequipment.com...