7. Auto Ramp in the mA, mV, and V functions
In the mA, mV, and V function, auto ramp function is available to the users. To
start the ramp function:
1. Press the SHIFT button to enter the SHIFT mode.
2. Press the NUMBER 7 key to start
3. Ramp function increment from 0% to 100%, then decrements from 100%
to 0% repeatedly. The resolution of each step is 1% (with respect the
range selected) at an interval of 0.08 seconds. So it takes 8 seconds to
ramp from 0 to 100%.
4. To temporarily stop the ramp function, press the NUMBER 7 key, and the
output will stay at the value when the NUMBER 7 key is pressed. These
features will facilitate the checking of trigger point of tested device, such as
valve controller.
5. To restart again, press the NUMBER 7 button again.
6. To return to the regular output mode, press the SHIFT button again when
ramp function is stopped.