3. Hz, Frequency Output

1. Turn the power on, then plug the test leads into the output connectors of
calibrator accordingly (Black to black, red to red). Attach alligator clip if
necessary.
2. Move the sliding switch to Hz position
3. Press the keypad (excluding the decimal point) to enter the value of Hz
directly.
4. Using the test leads or alligator clips, touch or clip on the terminals to be
calibrated.
5. Because not all the frequencies are available between 126 to 62500Hz,
calibrator will automatically adjust and display the users input value to a
frequency, which is available and always larger or equal to users input.
1
2
4-20mA
0-20mA
4
5
mV 1V
°
C
°
F
12V
7
8
%
. _
%
0
%
PROCESS
CALIBRATOR
Note: The resolution of Hz function is 1 Hz. For the range 1-125Hz, all the
frequencies in between are available. But for the range 126 - 62500Hz, not all
the frequencies are available (total 604 frequencies available). Please refer to
section III Electrical Specifications for available frequencies.
-
+