LIFEVET O OXYGEN ANALYSER USER MANUAL Art. No. 213195 TELEPHONE +49 7461 96580 0 www.eickemeyer.com...
Page 2
USER MANUAL LIFEVET O OXYGEN ANALYSER GENERAL The LifeVet O Oxygen Analyser is a battery-powered device suitable for monitoring the percentage of oxygen in an anaesthetic or in free-air. The LifeVet O uses a longlife galvanic cell and associated micro-power electronics to measure and display the oxygen level. The unit is powered by a single 1.5 volt alkaline battery and is designed to run for approximately one year, based on a usage rate of 8 hours per day.
Page 3
USER CONTROLS There are only 2 buttons on the LifeVet O unit. The left hand button is labelled 'CALIBRATE' and the right hand button is labelled 'ON / OFF' . For routine use only the ON / OFF button need be used to turn the unit on and off.
Page 4
5) The screen changes to a flashing 100 6) Flow 100 % oxygen through the LifeVet O connector for 60 seconds 7) After 60 seconds press the CALIBRATE button for 1 second 8) The screen changes to a flashing CC, indicating Calibration Complete 9) Press the CALIBRATE button one more time to leave the calibration routine RESTORING FACTORY DEFAULTS If the calibration routine is not performed correctly, or the values obtained after a calibration routine are not correct then the default factory settings can be loaded.
Page 5
Solution: The internal AA battery is low. Replace the battery. Problem: There is nothing on the display and no response to the ON / OFF button. Solution: Fit a new Long-life AA battery. If this does not resolve the problem, contact EICKEMEYER ®...
Page 6
GERMANY EICKEMEYER KG Eltastraße 8 78532 Tuttlingen T +49 7461 96 580 0 F +49 7461 96 580 90 E info@eickemeyer.de www.eickemeyer.de SWITZERLAND DENMARK ITALY EICKEMEYER ApS EICKEMEYER S.R.L. EICKEMEYER AG Sandgrube 29 Lysbjergvej 6, Hammelev Via G. Verdi, 8...
Need help?
Do you have a question about the LIFEVET O2 and is the answer not in the manual?
Questions and answers