Calibration Description - VXI VT1433B User Manual

8-channel 196 ksamples/s digitizer plus dsp
Table of Contents

Advertisement

Calibration Description

The Cal connector on the front panel of the standard VT1433B can be configured
(in software) as either an input or an output. It can be set to any of four
settings:
q
DC - The VT1433B outputs a DC calibration signal from the millivolt range up to
15 volts.
q
AC - The VT1433B outputs a signal from an Arbitrary Source option (in the same
module or a different VT1433B module in the system.)
q
Ground - The connector is shunted to ground for a zero-volt reference.
q
Open Circuit - In this mode the connector becomes an input which can receive a
calibration signal up to ±15 volts.
The VT1433B is calibrated at the factory and the calibration placed in EPROM
memory for use at each power-up. In addition an auto-zero function is provided.
VT1433A Substrate
DC Voltage
Reference
VXI SUMBUS
CALIN
CALOUT
Figure 5-6: Calibration block diagram
VT1433B User's Guide
Module Description
Front panel connector
Input Circuitry
to ADC
from
DC_BUS
Source
from
Output
Tachometer
trigger
monitor
5-15

Advertisement

Table of Contents
loading

Table of Contents