Sanxin PC5 Operation Manual

Ph/conductivity/tds/salinity/temp tester

Advertisement

Quick Links

PC5 pH/Conductivity Tester
pH/Conductivity/TDS/Salinity/Temp.
Operation Manual
Shanghai San-Xin Instrumentation Inc.

Advertisement

loading

Summary of Contents for Sanxin PC5

  • Page 1 PC5 pH/Conductivity Tester pH/Conductivity/TDS/Salinity/Temp. Operation Manual Shanghai San-Xin Instrumentation Inc.
  • Page 2: Battery Installation

    PC5 pH/Cond. Tester Operation Manual 1. Battery Installation Please install batteries according to the following steps. Please note polarity: “+” (anode) is upward; “-” (cathode) is downward. ① Pull the battery cap up ② Push the battery cover along to the direction of arrow ③...
  • Page 3 3. pH Calibration 3.1. Before first use or unused for a long time, fill certain 3M KCl solution in the calibration bottle, soak probe for at least 15 minutes. Pay attention that, soaking solution little higher that pH bulb will be enough. Users can make this solution by themselves. Preparation: dissolve 25g KCl in 100 ml of pure water.
  • Page 4 5.2. The tester has self-diagnostic function, information as following chart: Self-diagnostic information Icons Checking Wrong pH buffer solution or the recognition 1. Check whether pH buffer solution is correct. range of calibration solution exceeds standard. 2. Check whether the probe is damaged. Press key when measuring value is not Press...
  • Page 5 Calibration Indication Icon Calibration Standards Measuring Range 84μS/cm 0 ~ 200μS/cm 1413μS/cm 200 ~ 2000μS/cm 12.88mS/cm 2 ~ 20mS/cm 8.2. The tester has been calibrated in factory. Generally, user can use the tester directly or users can test conductivity standard solutions first. If the error is big, then calibration is needed. 8.3.
  • Page 6 9. Parameter Setting 9.1. Parameter setting method When turned off, long press to enter parameter setting→Short press to switch P1-P2… P8→Short Press parameter flickering→Short press to choose parameter→Short press to confirm parameter→Long press return to measurement mode. 9.2. Setting chart Code Factory Default Symbol...
  • Page 7 e) TDS factor (P5): Adjust TDS factor between 0.40~1.00. Factory default setting is 0.71. f) Salinity and temperature unit (P6 and P7): . Factory default setting is ppt and ℃ ℉ ℃ Salinity unit: ppt and mg/L, Temperature unit: g) Factory default setting (P8): Select “Yes”...
  • Page 8 11. Other Specifications Ultra clear STN LCD. 180°viewing angle Blue: Measurement. Green: Calibration. Red: Alarm Reading lock HOLD Replace batteries when icon blinks Low battery reminder Automatic power-off 8 minutes after last key press Water proof rating IP67 floats on water Power DC3V, AAA batteries×4 Battery life...