Incidental Fm Test; Maximum Rf Output Test - HP 608E Operating And Service Manual

Vhf signal generator
Hide thumbs Also See for 608E:
Table of Contents

Advertisement

Model 608E/ F
5-16. INCIDENTAL
FM
TEST.
a. Connect Signal generator in test setup illustrated
in Figure 5-4.
b. Set FREQUENCY RANGE switch to range A.
c. Adjust
frequency control for MEGACYCLES
dial setting of 10 MHz.
d. Adjust RF OUTPUT andAMPLTRIMMER
con-
trols for a +7 dBm (ATTENUATOR CALIBRATED)
indication on RF OUTPUT meter.
e. Set MODULATION switch to
INT
1000
posi-
tion.
Section
V
5-17.
MAXIMUM
RF OUTPUT TEST.
a. Connect signal generator in test setup illustrated
in Figure 5-5.
b. Set FREQUENCY RANGE switch to range
A.
c. Adjust RF OUTPUT control to the
full
clockwise
position.
d. Adjust attenuator control to the full clockwise
position.
f. Adjust MODULATION control for a 50% indica-
tion on PERCENT MODULATION meter.
g. Adjust attenuator control for output of 0 dEmo
h. Readjust frequency control for a 50 kHz indica·
tion on frequency meter on 100 kHz range.
i.
Measure peak-to-peak amplitude of signal dis-
played on oscilloscope.
Measurement should not ex-
ceed 2.0 mY peak-to-peak (200
Hz
peak-to-peak
maximum deviation).
60BE/F
SIGNAL
GENERATOR
o
0
8491,0,
10-OB
ATTENUo\TOfl
478A
T~ERMISTOR
MOUNT
432:"
"'
...
METER
j.
Repeat steps h.
::.;-.d
i.
:.t the FREQUENCY
RANGE and MEGACYCLES dial settings
listed
with
applicable oscUloscope indications in the following
chart.
Figure 5-5. RF Output Test Setup
OSCILLOSCOPE INDICATION
FREQ. RANGE
FREQ
INCIDENTAL FM
MAXIMUM DEY
B
30
MHz
6 mY pk-pk
600
Hz pk-pk
C
70
MHz
14 mY pk-pk
1400
Hz
pk-pk
D
150
MHz
20 mY pk-pk
2000
Hz
pk-pk
E
300
MHz
20 mY pk-pk
2000
Hz
pk-pk
E
400
MHz
20 mV pk-pk
2000
Hz
pk-pk
E
420
MHz
20 mV pk-pk
2000
Hz
pk-pk
E
450
MHz
(608E)
20 mY pk-pk
2000
Hz
pk-pk
E
455 MHz (608F)
20 mY pk-pk
2000
Hz
pk-pk
E
480
MHz
(608E)
20 mV pk-pk
2000
Hz
pk-pk
5-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for: