I-Tech WonJin Power Q-6000 User Manual

Compressive limbs therapy

Advertisement

Quick Links

USER MANUAL
MNPG61 Rev. 2 Edition 11/04/11
Compressive Limbs Therapy model
Power
Q-6000

Advertisement

Table of Contents
loading

Summary of Contents for I-Tech WonJin Power Q-6000

  • Page 1 USER MANUAL MNPG61 Rev. 2 Edition 11/04/11 Compressive Limbs Therapy model Power Q-6000...
  • Page 2 I.A.C.E.R. Srl Via S. Pertini 24/A – 30030 Martellago (VE) ITALY Tel. 041.5401356 • Fax 041.5402684 e-mail: iacer@iacer.it http://www.iacer.ve.it...
  • Page 3: Table Of Contents

    Summary Summary Compressive Limbs Therapy Machine body and accessories Accessories size Specifications Technical Specifications Purpose Working instructions Troubleshooting Warnings Warranty Assistance Spare parts MDD 93/42/EEC Compliance • Electromagnetic Interference tested • Agency for Technology • MD Manufacturer Reg. n° 1636 and Standards •...
  • Page 4: Compressive Limbs Therapy

    Compressive Limbs Therapy Compressive limbs therapy is able to promote a right venous circulation by decreasing muscle tension caused by stress or chronic and acute pains: thanks to its pump action it favorites venous return by increasing tissues blood circulation and finally their physiological renewal.
  • Page 5: Machine Body And Accessories

    Machine body and accessories Manometer Pressure adjustment switch Start/Stop Air discharge output Speed Level 1 - 80 sec. – 0,8 Rpm On/Off Timer Level 2 - 39 sec. – 1,8 Rpm (Incr./Decr.) Level 3 - 26 sec. – 2,0 Rpm Level 4 - 18 sec.
  • Page 6: Accessories Size

    Power Q-6000 is equipped with: • machine body • double legging (1) • 2 derivations connection tubes for leggings(2) • 2 pressure plantars(3) • abdominal band (4) • connection tube for abdominal band (5) • armband (6) • connection tube for armband (7) •...
  • Page 7: Specifications

    Specifications Power Q-6000 has the following specifications: • Class IIa equipment (Directive 93/42/EEC, annexed IX, rule 9); • Class I, applied part type BF (Classif. CEI EN 60601-1); • Equipment not protected against liquids penetration; • Equipment and accessories not subjected to sterilization; •...
  • Page 8 Insert the pressure plantars under the feet (available only with double leggings); iii) When wearing the leggings pay attention to position the tubes connection cables in the internal side of each legging; iv) Connect 6 tubes to the 6 plugs positioned on each legging, paying attention to connect the longer tube to the higher position connector of legging (thigh);...
  • Page 9 Wear the armband and connect the tubes to the armband plugs, paying attention to connect the longer tube to the higher position connector on armband; After connecting the tubes to the armband connect the tubes to the air discharge output. 3) Pressure regulator (selector on machine body) Air pressure in compression chambers can be adjusted from 0 to maximum pressure (around 217 mmHg ±...
  • Page 10: Troubleshooting

    Troubleshooting Problem Solution • Power Q-6000 Check the connection to the domestic plug doesn’t switch on • Check the ON/OFF button • Check the power supply status (no sign of damages) • Power Q-6000 emits Check if there are any damages and/or holes a strange sound like in the tubes and connectors air leakage...
  • Page 11: Warnings

    Warnings The following warnings guarantee user safety and prevention against damages and/or lesions. Please read it carefully before using Power Q-6000 Warning Caution In case of not following the In case of not following the instructions the user may instructions the user may suffer from injuries or device suffer from serious injuries damages...
  • Page 12 Do not exceed 20 minutes in the first Do not use the device if you feel pain or treatments. We suggest to start with bad feelings. treatments of 5/10 minutes Attention do not spill liquids or any Use only in consenting adults substances on the device.
  • Page 13 Caution Do not move when the device is working. Do not expose the device to the sun and do not storage it next to heat sources Do not place any objects on the device or If you not use the device for a long time or on its parts (connector tubes, applicators, if you use the device in cold places, a etc.)
  • Page 14: Assistance

    Maintenance, carriage and storage If it is used in accordance with the instructions of the user manual, the device does not need a particular regular maintenance. Carriage precautions It doesn’t need any particular carriage precautions: however we recommend to put away Power Q-6000 and its accessories in their own box after every treatment.
  • Page 15: Spare Parts

    Spare parts Contact the distributor or the fabricant for original spare parts at following address: I.A.C.E.R. S.r.l. Via S. Pertini, 24/a • 30030 Martellago (VE) Tel. +39 0415401356 • Fax +39 0415402684 www.iacer.it • iacer@iacer.it To preserve product warranty, functionality and product safety we recommend to use only original spare parts.