Page 2
About this quick guide: Service Manual Page 2...
Page 3
Table of contents ABOUT THIS QUICK GUIDE: 2 Service Manual Page 3...
Page 4
Secure the bond tester off and before switching it on again: Service Manual Page 4...
Page 5
Mounting the protective cage Remove the head cable. Open the 4 fastening screws on the test head Careful removal of the test head Service Manual Page 5...
Page 6
Remove the axis cable. Open the 4 fastening screws Service Manual Page 6...
Page 7
Degrease and clean the spindle and linear guides with cleanroom cloths and industrial alcohol or spirit. Then re-lubricate the spindle and the linear guides with a machine grease suitable for cleanroom use. (e.g.: ISOFLEX Klüber NBU 15) Available under the order number VK-DIV-B00007 Service Manual Page 7...
Page 14
Degrease and clean the spindle and the linear guides with cleanroom cloths and industrial alcohol or spirit. Then re-lubricate the spindle and the linear guides with a machine grease suitable for cleanroom use. (e.g.: ISOFLEX Klüber NBU 15) Available under the order number VK-DIV-B00007 Service Manual Page 14...
Page 15
After assembling the machine, you can reconnect the supply cable and switch on the machine. Check all functions of the machine before starting real pull tests. Service Manual Page 15...
Page 17