GasLab CM-1000 Operating Instructions

Pro multi gas sampling data logger

Advertisement

Quick Links

GasLab Pro Multi Gas Sampling Data Logger
Operating Instructions
CM-1000
Product Overview
Thank you for selecting the CM-1000 multi gas sampling data logger. The CM-1000 is designed to
simultaneously measure multiple gas concentrations through sampling methods. With long-term data storage,
the user can assess the previous records. When the CM-1000 measures the gas concentration reaching the
alarm setting or higher, the data display and alarm functions are activated. The CM-1000 is equipped with
an RS485 interface for connecting to a computer for remote monitoring. The Multi Gas Sampling Data
Logger can be used in a wide-range of industries and applications.
Features:
V □ Options for CO
measurement range: 1%, 5%, 10%, 20% 30%, and 100%
2
V □ Measure CO
, CO, O
, CH
, RH, AMB, DP, ALTI
2
2
4
V □ Large LCD for easy reading of gas concentration and temperature and humidity.
V □ Audible alarm
V □ Data logging with SD card
V □ T hrough the RS485 interface connection, multiple CM-1000 units can be connected to the back-end
computer system as a security monitoring application.
V □ Built-in LCD back light for easy reading in the dark.
V □ This device is supplied with Li- ion 18650 3.7V rechargeable batteries, can be used for a long time.
A. Power Key
B. Temperature & Humidity Sensor
D. Rubber Boot
E. Up Key
F. Left Key
H. Down Key
I. Mode Key
J. Enter Key
L. USB Socket
M. SD Card Slot
N. RS485 Jack
P. Luer Cap
Q. Screw Position
R. Battery Cover
Key description:
Enter the setting menu.
MODE (○
)
I
Up (○
E
) / Down (○
H
) / Left (○
F
) / Right (○
G
Change parameter values or select options.
)
Execute a command
Enter (○
)
J
Start the pump or data logger.
PUMP (○
)
K
LCD display symbol description:
Symbol
Meaning
CO
concentration, parts per
2
million (ppm)
CO concentration, parts per
million (ppm)
O
concentration, % (percent)
2
CH
concentration, parts per
4
million (ppm)
Relative humidity
Ambient temperature
Dew point
Atmospheric pressure
Operation Instructions
1. Power on:
Switch the Power Key (○
2. Measurement:
After power on, the device starts to measure and update the data every second.
Press Enter key (○
J
display a red frame. Press Left (○
selected parameter or sensor displays the red background color, press Right key (○
and then press Up (○
Zone 1 sensor display: CO
Zone 2 sensor display: O
Zone 3 sensor display: CO, CH
Zone 1 parameter display and setting
Item
(Display Features and Modes)
AL L / AL H
C. LCD
Note: AL L and AL H are expressed as a percentage. Example: 0.04 = 400ppm
G. Right Key
K. Pump Key
Zone 3 parameter unit selection:
Sensor
O. Analog / Voltage Output Port
DP
AMB
ALTI
3. Display all sensor readings
Press and hold Enter key (○
screen.
Description
The current CO
concentration
2
The current CO concentration
The current O
concentration
2
The current CH
concentration
4. Setting
4
Press and hold Mode key (○
Relative humidity
page. Press Up (○
E
Ambient temperature
(○
) to confirm it. Press and hold Mode key (○
J
The current dew point
Compensate the pressure changes with
The first page:
appropriate altitude of location when
Date
measure.
Time
Time weighted average (8 hours)
CO
, 8-hr time weighted average
2
Short‐Term Exposure Limit
CO
, 15 min. weighted average
2
(15 min. weighted average)
Maximum value
Maximum value of gas measurement
Minimum value
Minimum value of gas measurement
High/low alarm
Customize the range of color Bar
Battery capacity
Battery capacity without segment display.
Real-time display
12/24-hour time display
Need to calibrate the sensor or exceed the
Warning
pump life time.
The pump icon rotates to indicate that the
Pump
pump has been started.
Micro SD card working status (When the
icon flashes, it means that the Micro SD
Micro SD card
card is broken or full; when the icon remains
stationary, it indicates normal operation.)
) on.
A
) to select zone. The selection order is: Zone 1Zone 2Zone 3. The selected zone will
) / Right (○
) key to select the parameter or sensor to be adjusted. The
F
G
) to proceed the setting
G
) / Down (○
) key to edit the parameter value.
E
H
, O
, CO, CH
2
2
4
, CO, CH
, RH
2
4
, AMB, DP, ALTI
4
Select 1
Select 2
Select 3
Select 4
MAX
MIN
STEL
TWA
Press Up (○
) / Down (○
) key to
E
H
edit the parameter value.
Unit 1
Unit 2
Unit 3
DP=°C
DP=°F
°C
°F
m
mf
mmHg
) to display all sensor readings. Press Mode key (○
) to return to the main
J
I
) to enter the setup menu, then press Mode key (○
) to continue to the next
I
I
) / Down (○
) key to select options or change parameter values. And then press Enter key
H
) to return to the main screen.
I
01/01/2017 ~ 2099
12/24-hour time
Continue alarm
Yes (The buzzer sounds continuously.) / No (The buzzer sounds once.)
Screen auto rotate
Yes / No
Backlight level
1~32 (The larger the number, the brighter the screen)
Background color
□□□□
Backlight time
10~999 sec (Backlight display time)
Auto power off
1~999 min
Graph line 2
CO
/ O
2
About
CO2Meter – www.co2meter.com
The second page:
Analog for Zone 1
0-10V/4~20mA (Zone 1 voltage or current output)
Analog for 4-20mA
CO
/O
2
Data log time
Data logging interval time
Flow rate level
1~5 (The larger the number, the greater the pump output)
Exhaust time
10~240 sec
Ventilate time
0~120 sec
Gas compensation
Yes / No (CO
Sensor recalibrated time and pump life time (The time is displayed in the
countdown mode. If you need to calibrate the sensor or exceed the pump
life time, the time will be displayed in red.)
CO
: 99999999 days
2
Sensor times info
O
: 99999999 days
2
CO: 99999999 days
CH
: 99999999 days
4
Pump: 99999999 min
Reset to factory
Yes / No (Recover the factory setting to cancel customize setting.)
Sensor calibration (▲+▼)
Please refer to the calibration section.
5. Power off:
Switch the Power Key (○
) off.
A
Note: When the device is charged with USB, it cannot be turned off.
6. Calibration:
(1) Single-point calibration
After entering the setup menu, then press Mode key (○
select "Sensor calibration (▲+▼)" and press Enter key (○
Nothing
at the same time to proceed the calibration. Then, press Enter key (○
Target sensor
Raw reading
Offset
Perform reading
Unit 4
(2) Two-point calibration (calibrate by professional)
After entering the setup menu, then press Mode key (○
). "1" will appear next to "Sensor calibration (▲+▼)". Press Down
Up key (○
), then press Down key (○
E
H
hpa
key (○
) to select "Sensor calibration (▲+▼) 1" and press Enter key (○
H
Down (○
) keys at the same time to proceed the calibration. Then, press Enter key (○
H
settings.
Target sensor
Raw reading
Level 1 input
Level 2 input
Level 1 catch
Level 2 catch
Cali Set
Gain
Offset
Perform reading
(There are a total of 4 background colors.)
/CO/ CH
(The second trend line shows the gas measurements.)
2
4
/CO/CH
(Another analog output option)
2
4
and O
)
2
2
) to continue to the next page. Press Down key (○
) to
I
H
) to confirm it. Press Up (○
) and Down (○
) keys
E
H
J
) to enter the sensor settings.
J
CO
/O
/CO/CH
/AMB/RH
2
2
4
Unadjusted reading
The amount of offset
Adjusted reading
) to continue to the next page. Firstly, press and hold
I
) to confirm it. Press Up (○
) and
E
J
) to enter the sensor
J
CO
/O
/CO/CH
/AMB/RH
2
2
4
Unadjusted reading
Low concentration value of standard gas
High concentration value of standard gas
Connect the low concentration gas to the device, and
press Pump key (○
) to start the pump. (Set the data
K
log time to less than 1 min). After the reading is
stable, press Enter key (○
) to confirm it.
J
Connect the high concentration gas to the device,
and press Pump key (○
) to start the pump. (Set the
K
data log time to less than 1 min). After the reading is
stable, press Enter key (○
) to confirm it.
J
When "Start" turns red, press Enter key (○
) to
J
confirm the calibration.
The amount of Gain.
The amount of offset
Adjusted reading

Advertisement

loading

Summary of Contents for GasLab CM-1000

  • Page 1 High/low alarm Customize the range of color Bar About CO2Meter – www.co2meter.com alarm setting or higher, the data display and alarm functions are activated. The CM-1000 is equipped with The second page: Battery capacity Battery capacity without segment display. an RS485 interface for connecting to a computer for remote monitoring. The Multi Gas Sampling Data...
  • Page 2 Cause an impact on the accuracy of temperature when charging, please use a fan to blow toward the Temperature & Humidity Sensor (○ ) directly CM-1000 Multi Gas Sampling Logger in order to get a compensated balance of temperature and humidity between temperature sensor and Manual surrounding area.