Safety instructions for potential hazard areas Safety information for the individual operating phases Design and function Identification and labelling Set-up of PrograPrint Clean Magnetic stirrer Transporting and setting up the device Choosing the site of operation Transporting the device to the place of operation and unpacking...
Page 4
Table of contents Connecting the unit to the power supply Operating the magnetic stirrer Switching the magnetic stirrer on Setting the cleaning duration Setting the speed Using the quick set Switching the magnetic stirrer off Cleaning the dental object Labelling the cleaning containers Preparing the magnetic stirrer and filling the cleaning containers Inserting the build platform into the unit Conducting rough cleaning...
Page 5
Table of contents Technical data 12.1 Classifications 12.2 Measurements, weight 12.3 Ambient conditions 12.4 Electrical data 12.5 Power supply Appendix 13.1 Scope of delivery 13.2 Ordering spare parts 13.3 Recommended additional accessories 13.4 Approved cleaning fluid 13.5 CE certification / declaration of conformity...
About this document These Operating Instructions will familiarize you with the PrograPrint Clean features. PrograPrint Clean is a cleaning device for the further processing of dental objects produced with PrograPrint PR5. This device has been developed and produced according to the latest state-of-the-art technology.
About this document Signs and symbols in this document 1.2.1 Warnings Warnings are used in these Operating Instructions to warn you of a risk of injury to persons and/or damage to property. 1. Please always read and observe these warnings. 2.
Page 8
About this document 1.2.2 Signs and symbols Symbol Meaning Additional information, e.g. for better understanding, for simplifying workflows or for further information. Prerequisite which must be followed before the subsequent actions can be performed. Individual steps which you must perform. 1., 2.
About this document Warning symbols and command signs on the device Warning symbols and command signs are attached to the device in order to warn you of damage to property and personal injury. 1. Always observe the warning symbols and command signs on the device. 2.
Short instructions on the entire workflow of the PrograPrint system Instructions for Use for Manual for the CAM software "CAMbridge" CAMbridge Short Instructions for: Excerpt of operation-relevant topics from the respective Operating Instructions • PrograPrint PR5 • PrograPrint Clean • PrograPrint Cure...
Contents of the supplement sheet “Important information on the handling of isopropanol” If you operate the PrograPrint Clean cleaning device with ispropanol as the cleaning fluid, the following points must be observed: 1. The PrograPrint Clean cleaning device must only be operated in an explosion-proof fume hood / explosion-proof extraction cabinet due to the isopropanol vapours (explosion hazard).
Page 12
About this document Wear protective gloves / protective gear / eye protection / face protection. UPON EYE CONTACT: Carefully rinse with water for several minutes. Remove any contact lenses, if possible. Continue rinsing. Dispose of the contents / container according to the local / regional / national / international regulations.
Nevertheless, operation bears the risk of injury to the user or third parties. Moreover, impairments of the device and other property are possible. Intended use PrograPrint Clean has been designed for cleaning the dental objects produced by the PrograPrint PR5 in the following materials: •...
Page 14
Safety 2.1.1 Potential improper use The device is not suitable for cleaning 3D printed objects from outside of the dental field. The following is considered as misuse of the system: • Non-compliance with the specified intended use, intended user specifications or intended environment •...
Safety Improper use Consequences • Leaking or spilling the cleaning fluid Transportation of the device with full cleaning containers • Skin irritations and allergic reactions Removing the cleaning container when it is • Danger of fire or explosion full • Environmental damage Filling the cleaning containers beyond the •...
Safety Personnel qualifications Ensure that the personnel assigned to work with the device have read and understood these instructions and all documents of equal applicability, in particular safety, maintenance and repair information, before starting work. Ensure that personnel is aware of hazards and safety equipment (“Working areas, potential hazards and safety measures on the device”, page 17).
Safety Working areas, potential hazards and safety measures on the device As an operator, you are working at the areas described below. The device is built according to state-of-the-art technology and recognized technical safety regulations. Hazardous areas, which cannot be avoided by design, are equipped with appropriate protective devices.
Page 18
Safety CAUTION! Evaporation of cleaning fluid due to the lid having been open for too long. Skin irritations and allergic reactions. Danger of fire or explosion. Do not block the self-closing lid with objects which keep it open. CAUTION! Evaporation of cleaning fluid due to damaged seals around lid and cleaning container.
Page 19
Safety Type of Safety Protective Working area Tasks / Work hazard equipment effect Cleaning • Inserting Leaking, Build platform Safe transfer containers the build overflowing holder of the build platform material platform from spillage due to one container • Cleaning the build to the other dental...
Page 20
Safety 2.6.1 Fume hood / extraction cabinet at the site of use CAUTION! Evaporation of cleaning fluid during use without fume hood / extraction cabinet. Respiratory disorders, skin irritations and allergic reactions. Danger of fire or explosion. Place and operate the device filled with isopropanol exclusively under an explosion-proof fume hood / explosion-proof extraction cabinet.
Safety Safety instructions for potential hazard areas 2.7.1 Magnetism The inductive drive of the magnetic stirrer works with magnetism. Keep pacemakers, data carriers, magnetic cards and other devices that may be affected by magnetic fields away from the device. 2.7.2 Electrical safety If covers are opened or parts are removed which are only accessible with tools, voltage- carrying components may be exposed.
Page 22
Safety 2.7.3 Hazardous, inflammable materials Cleaning fluid (isopropanol) Depending on the cleaning fluid used, there is a risk of skin irritations, allergic reactions or respiratory diseases without the corresponding safety precautions. Due to the cleaning fluid vapours, there is a risk of fire or explosion if the device and cleaning fluid are handled incorrectly.
Safety Safety information for the individual operating phases 2.8.1 Transport Only transport the device according to the description in these Operating Instructions (“Transporting and setting up the device”, page 32). In order to avoid corrosion on the device and subsequent damage to the device, transport and store the device only within the permissible temperature range and ambient conditions (“Ambient conditions”, page 71).
Page 24
Safety Unsupervised operation: The device may be operated unsupervised, provided the national and local laws and provisions allow for such action and provided that they are observed. Furthermore, the requirements of the respective insurance company must be met. Never use the device if the cleaning containers are heavily soiled. ...
When the build platform with the dental objects is removed from the 3D printer PrograPrint PR5, the surfaces of the objects will be covered with unpolymerized material, which must be removed. The PrograPrint Clean unit is designed to clean these dental objects. PrograPrint Clean consists at least of the following components: •...
Page 26
Material bottle with valve Provision of the material in the (Pos. c, bottle content 1000 ml) cartridge (b) PrograPrint Clean Cleaning of the dental object after the printing process PrograPrint Cure (to be ordered Final polymerization of the dental...
Design and function Identification and labelling PrograPrint Clean has the product name “PrograPrint Clean” and the Ivoclar Digital logo imprinted on the front side. The back of the device and the magnetic stirrer are both labelled with a type plate:...
Page 28
Design and function Example for PrograPrint Clean: PrograPrint Clean ® 48 V 20 W Ivoclar Vivadent AG 9494 Schaan / Liechtenstein Made in Austria Symbols on the type plates: Symbol Meaning Manufacturer Serial number Manufacturer / Production date This product meets the requirements of the applicable EU directives...
Design and function Set-up of PrograPrint Clean Description Function Handle is in the lid Open and close the lid Self-closing lid Cover for the cleaning container Cleaning container for rough or fine Cleaning container for the first cleaning cleaning sequence (rough cleaning) or the second...
Page 30
Design and function Description Function Sliding device with guide rail for the Moving the magnetic stirrer from one magnetic stirrer cleaning container to another Hand pump with suction fitting Disposal of contaminated cleaning fluid Standing feet (4 x at the base of the Secure position of the device unit) Mains connection / power supply of the...
Design and function Magnetic stirrer PrograPrint Clean must only be used with the “2-mag Magnetic stirrer MIX-1 Ivoclar” illustrated here. Element Function Reference Mixing plate In the PrograPrint “Cleaning and construction, without direct maintenance table”, function page 59 Key [+] According to mode: •...
Transporting and setting up the device Transporting and setting up the device WARNING! Risk of injury from the device tipping over or falling down. Crushing of extremities. Damage to the device. Note the weight of the unit (> 18 kg) and if necessary, carry with two persons. ...
Risk of slipping due to dripping uncured material or skin contact with dripping, uncured material: a. In order to create an efficient working environment, set up the PrograPrint PR5 and PrograPrint Clean in close proximity to each other. b. Keep cloths and collection containers ready and use in case of contamination. ...
Transporting and setting up the device Transporting the device to the place of operation and unpacking WARNING! Risk of injury from the device tipping over or falling down. Crushing of extremities. Damage to the device. Note the weight of the unit (> 18 kg) and if necessary, carry with two persons. ...
Connecting the unit to the power supply Connecting the unit to the power supply Only use the supplied original power adapter or an equivalent replacement power adapter from Ivoclar Vivadent. The requirements for the power supply are met (“Electrical data”, page 71, “Electrical safety”, page 21 and type plate, see “Identification and labelling”, page 27).
Operating the magnetic stirrer Operating the magnetic stirrer In the following sections, the operation of the magnetic stirrer is described in general and without reference to the types of material and their requirements in terms of speed and cleaning time. Note that the unit will start stirring directly as soon as the button [On / Off] (1) is activated.
Operating the magnetic stirrer Setting the cleaning duration The pre-set cleaning duration of the magnetic stirrer is 20 min. The cleaning time can be set to values between 1 and 90 min. In setting <1 min, the magnetic stirrer is in continual operation. To set the cleaning duration, proceed as follows: ...
Operating the magnetic stirrer Using the quick set With this function, you can set the unit directly to the minimum or maximum cleaning time or speed. 6.4.1 Setting the minimal cleaning duration To set the cleaning duration to the a minimum of 1 min, proceed as follows: ...
Page 39
Operating the magnetic stirrer 6.4.3 Setting the minimal speed To set the speed to the a minimum of 100 rpm, proceed as follows: The device is switched on. 1. Press the [Mode] key (2). The speed mode is activated. 2.
Operating the magnetic stirrer Switching the magnetic stirrer off The device is switched on. Activate the [On / Off] key (1) at the magnetic stirrer. The magnetic stirrer is switched off.
The patient absorbs the residues. Nausea or allergic reactions. Poor precision and material properties. As described, clean the dental object with the PrograPrint Clean before further processing. WARNING! Skin or eye contact with the cleaning fluid. Skin irritation, allergic reactions or eye irritation.
Cleaning the dental object Labelling the cleaning containers PrograPrint Clean is equipped with two cleaning containers. One cleaning container is for rough cleaning and the other is for fine cleaning. Ivoclar Vivadent recommends that the cleaning containers are marked with the labels supplied in order to avoid any mix-up.
Cleaning the dental object Preparing the magnetic stirrer and filling the cleaning containers DANGER! Spilling, leaking or overflowing of cleaning fluid when filling the cleaning containers. Leaking of material harmful to health and the environment. Skin irritations and allergic reactions. Danger of fire or explosion.
Page 44
Cleaning the dental object The cleaning containers are labelled (“Labelling the cleaning containers”, page 42). The cleaning containers are inserted into the unit. 1. Magnetic stirrer is inserted into the guide rail. WARNING! The lid falls shut unintentionally. Risk of crushing.
Page 45
Cleaning the dental object 2. Keep the lid open and insert one magnetic stir bar (1) into each cleaning container. 3. Keep the lid open and fill both cleaning containers with cleaning fluid up to at least the lower fill level mark and at the most to the upper fill level mark (2) (min.
Cleaning the dental object Inserting the build platform into the unit CAUTION! Contamination from uncured material dripping down when attaching the build platform to the build platform holder. Skin irritation, allergic reactions or eye irritation. Always wear suitable protective equipment when working with the build platform (“Personal protective equipment”, page 16).
Page 47
Cleaning the dental object CAUTION! Contact with the material upon removal of the build platform. Spilling or slopping of material harmful to health and the environment. Skin irritations and allergic reactions. Danger to the environment. Damage to the device and the cartridge. ...
Page 48
Cleaning the dental object 5. Keep the lid open. 6. Place the build platform holder with the inserted build platform into the cleaning container for rough cleaning. 7. Close the lid. The build platform is inserted into the unit.
Cleaning the dental object Conducting rough cleaning WARNING! Skin or eye contact with the cleaning fluid. Skin irritation, allergic reactions or eye irritation. Always wear suitable protective equipment when working with the cleaning fluid (“Personal protective equipment”, page 16). NOTICE! Incorrect positioning of the magnetic stirrer under the cleaning container.
Page 50
Cleaning the dental object WARNING! Insufficient cleaning of the dental objects due to low speed or reduced cleaning duration. Residues of uncured material are still on the dental objects. The patient absorbs the residues. Nausea or allergic reactions. Poor precision and material properties. ...
Cleaning the dental object Conducting fine cleaning WARNING! Skin or eye contact with the cleaning fluid. Skin irritation, allergic reactions or eye irritation. Always wear suitable protective equipment when working with the cleaning fluid (“Personal protective equipment”, page 16). CAUTION! Evaporation of cleaning fluid due to the lid having been open for too long when inserting the build platform.
Page 52
Cleaning the dental object 2. Remove the build platform holder with the inserted platform from the rough cleaning container. Do not change the position of the rough and fine cleaning containers during different printing procedures. Ivoclar Vivadent recommends labelling with the supplied identification labels.
Page 53
Cleaning the dental object NOTICE! Incorrect positioning of the magnetic stirrer under the cleaning container. Reduced cleaning performance. Make sure that the magnetic stirrer is located in the centre, underneath the cleaning container. 5. Push the magnetic stirrer (1) under the cleaning container for fine cleaning (2).
Cleaning the dental object 8. Wait during the total cleaning time. 9. Switch the magnetic stirrer off (“Switching the magnetic stirrer off”, page 40). Fine cleaning has been completed. You can now remove the build platform with the dental objects from the unit and dry with compressed air (“Removing the build platform and drying the object”, page 54).
Page 55
Cleaning the dental object 2. Remove the build platform holder (1) with the inserted build platform from the cleaning container for fine cleaning. 3. Close the lid. CAUTION! Careless handling of the build platform. Damage to printed dental object. Handle the build platform with care.
Page 56
Cleaning the dental object CAUTION! Inhalation of harmful material particles stirred up by compressed air. Respiratory disorders and allergic reactions. Never point the compressed air in the direction of a person. Carry out drying procedure away from people. NOTICE! Point the compressed air directly at the dental object.
Device messages and troubleshooting Device messages and troubleshooting If you cannot solve possibly occurring problems with the help of the following table, please contact your authorized service partner. Error Possible reason(s) Solution Magnetic stirrer is not No magnetic stir bar •...
Cleaning and maintenance work for the user Cleaning and maintenance work for the user This chapter deals exclusively with the cleaning of the device. For information on cleaning the other system components, such as the PrograPrint PR5 or PrograPrint Cure, refer to the Operating Instructions of the relevant components (“Documents of equal applicability”, page 10).
Cleaning and maintenance work for the user Cleaning and maintenance table Compliance with the following cleaning and maintenance intervals is the sole responsibility of the operator! Duties Responsible person Interval Reference Cleaning the base User/operator Upon contamination • “Cleaning the and magnetic stirrer of the base or mixing base and the...
Cleaning and maintenance work for the user Cleaning the base and the magnetic stirrer DANGER! Cleaning work during operation. Contact with voltage-carrying components. Danger of electric shock. Damage to the device. Perform cleaning work only when the device is switched off and disconnected from the power supply.
Cleaning and maintenance work for the user Refilling the cleaning fluid DANGER! Spilling, leaking or overflowing of cleaning fluid when filling the cleaning containers. Leaking of material is harmful to health and the environment. Skin irritations and allergic reactions. Danger of fire or explosion. Danger to the environment.
Page 62
Cleaning and maintenance work for the user NOTICE! Level of cleaning fluid in the cleaning container is too low. Reduced cleaning performance. Ensure that the cleaning containers are filled with cleaning fluid according to the fill level marking in the cleaning containers (min. and max. level). ...
Cleaning and maintenance work for the user Replacing the cleaning fluid CAUTION! Evaporation of cleaning fluid due to the lid having been open for too long when changing the cleaning container. Skin irritations and allergic reactions. Danger of fire or explosion. ...
Page 64
Cleaning and maintenance work for the user DANGER! Spilling, leaking or overflowing of cleaning fluid when filling the cleaning containers. Leaking of material is harmful to health and the environment. Skin irritations and allergic reactions. Danger of fire or explosion. Danger to the environment.
Cleaning and maintenance work for the user Disposing of the cleaning fluid The unit comes with a hand pump, two hose ends and a suction nozzle. Using this equipment, you can pump the used, dirty cleaning fluid out of the cleaning containers and into empty cleaning canisters for appropriate disposal according to country-specific regulations.
Page 66
Cleaning and maintenance work for the user The magnetic stirrer is switched off (“Switching the magnetic stirrer off”, page 40). No build platform holders are in the cleaning containers. 1. Place the suction nozzle (1) over the edge of the cleaning container and the base as shown.
Decommissioning 10 Decommissioning WARNING! Tipping or falling of individual, filled cleaning containers or the filled unit. Leaking of material is harmful to health and the environment. Skin irritations and allergic reactions. Danger of fire or explosion. Danger to the environment. ...
Decommissioning 10.2 Disposal of the device and materials The operator is responsible for proper disposal. The product must not be disposed of in the household waste. It must be collected separately from the household waste and disposed of according to local regulations for the disposal of electronic equipment in an environ- mentally safe manner or returned to Ivoclar Vivadent for disposal.
Repair 11 Repair DANGER! Improper repairs/maintenance work. Danger to the user and to the patient. Repairs on the device must only be carried out by service partners authorized by Ivoclar Vivandent using the valid version of the Service Manual. 11.1 Authorized service partners For information about the authorized service partners in your country, contact your local distribution partner or Ivoclar Vivadent or go to: http://www.ivoclarvivadent.com.
Technical data 12 Technical data The following technical data refer exclusively to PrograPrint Clean. For information on the technical data of the accessories or other components, please refer to the respective Operating Instructions (“Documents of equal applicability”, page 10). 12.1 Classifications...
Technical data 12.3 Ambient conditions 12.3.1 Ambient conditions for storage and transportation without cleaning fluid Characteristics Value Acceptable ambient temperature -20 °C bis +65 °C 12.3.2 Ambient conditions during operation with cleaning fluid / when filled with cleaning fluid Characteristics Value Acceptable ambient temperature +20 °C to +30 °C...
Appendix 13 Appendix 13.1 Scope of delivery Item description Number Base (housing with self-closing lid) Cleaning container Magnetic stirrer incl. power adapter Magnetic stir bar Build platform holder Hand pump with suction fitting Short instructions 13.2 Ordering spare parts 1. When ordering spare parts, keep the following information ready: •...