ThermoFisher Scientific OdaLog LL User Manual
ThermoFisher Scientific OdaLog LL User Manual

ThermoFisher Scientific OdaLog LL User Manual

Gas logger
Table of Contents

Advertisement

Quick Links

Thank you for purchasing the OdaLog
Thermo Fisher Scientific Australia Pty Ltd.
In our quest to continually improve the product we manufacture, we look
forward to your feedback on your experience using the OdaLog
environments and applications, as well as the usefulness of the information
you collect using it's data-logging function.
If you have any suggestions or comments, please contact your distributor or
alternatively e-mail us direct on sales@odalog.com.
The sole purpose of this document is to assist in the operation of the OdaLog
We do not convey any right to use this information for any purpose other than in the operation or maintenance
®
of the OdaLog
portable gas logger.
OdaStat™ & OdaStat-G™ © Software Copyright 2008 (Thermo Fisher Scientific Australia Pty Ltd)
This Manual summarises our knowledge of the OdaLog
the information available at the time. You should read this Manual carefully and consider the information in the
context of how the instrument is to be used.
The OdaLog
®
portable gas logger is sold subject to our standard terms and conditions of sale and any written
warranty given by Thermo Fisher Scientific. Thermo Fisher Scientific does not accept any other liability either
directly or indirectly for any losses suffered in connection with the use and application of the OdaLog
gas logger, whether or not in accordance with any advice, specification, recommendation or information given
by it or as a result of any errors or omissions in this Manual.
P/N: 10-1000
®
portable gas logger manufactured by
Gas Logger
COPYRIGHT© 2018
GENERAL DISCLAIMER
®
portable gas logger including how to use it based on
Revision: F, Issued: October 2019
®
in different
®
portable gas logger.
®
portable
Page: i

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OdaLog LL and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ThermoFisher Scientific OdaLog LL

  • Page 1 ® Thank you for purchasing the OdaLog portable gas logger manufactured by Thermo Fisher Scientific Australia Pty Ltd. In our quest to continually improve the product we manufacture, we look forward to your feedback on your experience using the OdaLog ®...
  • Page 2: Table Of Contents

    ® OdaLog Gas Logger User Manual Table of Contents CERTIFICATION / CONFORMITY ......................1 & P .. 1 XPLOSION ROTECTED LECTRICAL QUIPMENT ERTIFICATION ERFORMANCE ONFORMITY ........................1 ONDITIONS OF (EMC) – C ..........1 LECTROMAGNETIC COMPATIBILITY OMPLIANCE TATEMENT FCC S ............................
  • Page 3 ® OdaLog Gas Logger User Manual MAINTENANCE ............................19 ......................19 AINTENANCE ROCEDURES 8.1.1 Battery Replacement ........................19 8.1.2 Filter Replacement ........................21 ..............................22 ERVICE ® CALIBRATING THE ODALOG ......................23 ......................23 REPARATION FOR ALIBRATION ....................23 DSORPTION OF ALIBRATION IXTURE .......................
  • Page 4: Certification / Conformity

    ® OdaLog Gas Logger User Manual 1 CERTIFICATION / CONFORMITY 1.1 Explosion-Protected Electrical Equipment - Certification & Performance Conformity ® The following Authorities have granted intrinsic safety approval to the OdaLog Logger (Type L2). Authority Certificate Number Code EX TC IECEx ITA 08.0010X Ex ia IIC T4 (T = -20...
  • Page 5: Warnings And Cautions

    ® OdaLog Gas Logger User Manual 2 WARNINGS AND CAUTIONS ®  D O NOT USE SOLVENTS TO CLEAN THE ODALOG OGGER AMP CLOTH SHALL BE USED TO CLEAN OR WIPE  D O NOT INSERT OBJECTS INCLUDING FINGER INTO GAS INLET AS THE FILTER MAY BE DAMAGED ...
  • Page 6: Description

    ® OdaLog Gas Logger User Manual 3 DESCRIPTION ® The OdaLog is a compact portable gas data-logger, designed by industry-proven manufacturer Thermo Fisher Scientific Australia – the gas detection specialists for harsh environments, such as the waste water industry. With superior sealing and use of heavy-duty materials, this instrument is corrosion-resistant, durable and reliable.
  • Page 7 ® OdaLog Gas Logger User Manual P/N: 10-1000 Revision: F, Issued: October 2019 Page: 4...
  • Page 8: General Specifications

    ® OdaLog Gas Logger User Manual 4 GENERAL SPECIFICATIONS Table 2 General Product Specifications Instrument Temp Range -20°C (-4 F) to 50°C (122 Sensor Temp Range -10°C to +40°C (14 F to 104 Instrument Relative 15-90% non-condensing Humidity Range Ingress Protection IP66/68 External Dimensions Diameter 62mm (2.44”) x Height 196mm (7.72”)
  • Page 9: Quick Operational Guide

    ® OdaLog Gas Logger User Manual 5 QUICK OPERATIONAL GUIDE ® This section describes a brief operating procedure for the OdaLog instrument, covering the basics of all major functionality. For more detail, refer to corresponding information in Section 6. 5.1 Switch On Press &...
  • Page 10: Null (Zero)

    ® OdaLog Gas Logger User Manual 5.3 Null (zero) With unit in fresh air press & hold button down until “NULL” is displayed and then release. Move magnet over wording “NULL” on LHS of LCD screen – “DONE” should appear briefly when completed. (See Section 9.5) 5.4 Pre-Use Span Check Apply a known level of gas over sensor (using supplied calibration fitting, [Part No.
  • Page 11: Operation

    ® OdaLog Gas Logger User Manual 6 OPERATION 6.1 Operation Overview P/N: 10-1000 Revision: F, Issued: October 2019 Page: 8...
  • Page 12: Switching On

    ® OdaLog Gas Logger User Manual 6.2 Switching on Press & hold button down for three seconds (“On” will count down from 3) then release. The LCD display will show (In Order):  Test Screen – All segments visible/black  “CLOCK…not Set”...
  • Page 13: Displaying Current Status On The Lcd

    ® OdaLog Gas Logger User Manual 6.4 Displaying current status on the LCD Press & hold button until “STATUS” is displayed then release button. The display will cycle through “STATUS” screens as shown. 6.5 Start Logging Function (*Manual Logging Mode only, [Default = Enabled]) Press &...
  • Page 14: Auto Log Start Function [Default = Disabled]

    ® OdaLog Gas Logger User Manual 6.7 Auto Log Start Function [Default = Disabled] This function is activated via the OdaStat-G™ software package as per the “Instrument Parameters” section, by selecting the “Auto Log Start” option. Once active, logging will automatically start and stop every time the OdaLog ®...
  • Page 15: Communications Set-Up

    ® OdaLog Gas Logger User Manual  Ensure that other devices with Infra Red ports (phones, other computers, organisers etc) that are within ® about 10 metres of the OdaLog do not have their Infra Red ports activated.  Ensure that strong light sources are not directed onto the OdaLog ®...
  • Page 16: Bt (Bluetooth Data Communication Mode)

    ® OdaLog Gas Logger User Manual 6.9 BT (Bluetooth Data Communication Mode) 6.9.1 Enter “BT” mode (Bluetooth Data Communication Mode) From ON Mode: From the ‘INST’ screen, press & hold button down until “BT” is displayed then release button (BT will appear on screen after ‘irDA’, see diagram in Section 6.1). From OFF Mode: From the ‘OFF’...
  • Page 17: Switching Off

    ® OdaLog Gas Logger User Manual 6.10 Switching off Press & hold button down and wait for display to cycle through screens until “OFF2” is displayed. Keep button depressed until the display counts down from “OFF2” to “OFF” then release button. When the button is correctly released to turn the instrument off, the OFF title will shift 1 character to the right to confirm that the instrument is now in the off state.
  • Page 18: Low Battery Indication & Control

    ® OdaLog Gas Logger User Manual 6.12 Low Battery Indication & Control Battery power levels are indicated on-screen by means of the multi-segment battery symbol, as shown above with the transitions with remaining life expectancy. When the battery is above 50% capacity all segments of the will be displayed, as in stage 1 shown above.
  • Page 19: Negative Suppress Function (Default = Enabled)

    ® OdaLog Gas Logger User Manual ® To Un-Lock the OdaLog instrument: Place the magnetic base of Allen Tool provided on the left-hand side of the LCD display. Press the top button and the display will return to “INST” reading screen. The instrument is now unlocked. 6.14 Negative Suppress Function (Default = Enabled) This function is activated via the OdaStat-G™...
  • Page 20: Sensor Operation

    ® OdaLog Gas Logger User Manual 7 SENSOR OPERATION ® The OdaLog uses an electrochemical sensor to detect levels of a specific gas in ppm (parts per million). The sensor consists of electrolyte and electrodes packaged in a small container with a diffusion barrier which the gas passes through.
  • Page 21: Humidity

    ® OdaLog Gas Logger User Manual 7.2.2 Humidity IMPORTANT NOTE ON HUMIDITY: The Sensor fitted to the OdaLog Instrument is susceptible to reduced response in continuous high-humidity environments. With the wastewater industry being a common application for the OdaLog, it is IMPORTANT that users understand the effect humidity has on the sensor.
  • Page 22: Maintenance

    ® OdaLog Gas Logger User Manual 8 MAINTENANCE ® Routine maintenance of the OdaLog instrument is an extremely important factor affecting the reliability of the instrument. Optimum instrument performance and reliable operation will only be achieved if a dedicated maintenance programme is followed (Refer to Section 7 for sensor limitations). To ensure reliable measurements, the OdaLog ®...
  • Page 23: Figure 2 - Remove Top Cap Screw, Remove Top Cap And Removing Body

    ® OdaLog Gas Logger User Manual Figure 2 - Remove Top Cap Screw, remove Top Cap and Removing Body  Remove and replace battery, observing correct polarity (see Figure 3 below). Take care when inserting the new battery that the positive terminal insulator is not damaged – this insulator may cause the battery to break contact if the insulator tab damaged.
  • Page 24: Filter Replacement

    ® OdaLog Gas Logger User Manual  Place Dry Pack (moisture absorption pack) [Part No. 12-0129] on top of -ve battery clip  Plug top cap back into circuit board and insert top cap (carefully aligning the Top Cap and the body NOTE: Ensure the Top Cap is aligned correctly - See Figure 4 ...
  • Page 25: Service

    ® OdaLog Gas Logger User Manual NOTE: Ensure hands are free of dirt and grease prior to handling new filter  Remove filter  Remove, clean and replace stainless filter grill (if fitted)  Replace filter with ‘shiny’ side facing outwards ...
  • Page 26: Calibrating The Odalog

    ® OdaLog Gas Logger User Manual ® CALIBRATING THE OdaLog CAUTION OdaLog Instruments should be calibrated by authorised, competent personnel using the correct equipment. If unsure, or the correct equipment is not available, then return the instrument to Thermo Fisher Scientific or an authorised distributor / technical workshop for calibration.
  • Page 27: Calibration Mixture Application Times & Rates

    ® OdaLog Gas Logger User Manual Figure 6 - Calibration Mixture Connections Figure 7 - Magnetic switches (RMS/LMS, SPAN/ NULL) 9.4 Calibration Mixture Application Times & Rates This table shows the minimum calibration mixture flow time & rate necessary for calibration. Be sure that the correct flow rate and time is used.
  • Page 28: Null (Fresh Air Zero)

    ® OdaLog Gas Logger User Manual 9.5 NULL (Fresh Air Zero) NOTE It is recommended that logging sessions are stopped prior to a calibration  ® Ensuring the OdaLog is in Clean Air, press & hold down button until display progresses until “NULL” appears on screen, then release button.
  • Page 29: Span Set

    ® OdaLog Gas Logger User Manual 9.6 Span Set  Set the calibration level via the OdaStat-G™ software package “Instrument Parameters” section to match the level stated on the calibration mixture cylinder that is being used.  Insert the calibration fitting [Part No 11-0001] over the sensor in the bottom of the OdaLog ®...
  • Page 30: Calibration Errors

    ® OdaLog Gas Logger User Manual  Ensure that instantaneous display reads the same as shown on cylinder before turning off gas regulator and removing calibration fitting.  Ensure that reading returns to close to zero within 5 minutes in clean air and re-zero if necessary. Note: The instrument will TIMEOUT and return to the ‘INST’...
  • Page 31: Span Gas Check - Pre-Use And Post-Use (If Required)

    ® OdaLog Gas Logger User Manual SENSOR bAd1: The sensor response/reliability has been calculated to be LOW, recheck the calibration and review Section 7 for sensor information. Span value WILL still be saved, however readings may be inaccurate (> ±10%). SENSOR bAd2: The sensor response/reliability has been calculated to be VERY LOW, recheck the calibration and review Section 7 for sensor information.
  • Page 32: Decay Compensation

    ® OdaLog Gas Logger User Manual After running in fresh air for around 48 hours, the sensor should recover and again read close to correct level when exposed to calibration mixture, however, if it is not reading accurately it should be re-calibrated prior to the next “logging session”.
  • Page 33: Optional Accessories And Spares

    ® OdaLog Gas Logger User Manual 10 OPTIONAL ACCESSORIES AND SPARES 10.1 Accessories Part Number Description 11-0000 Allen Tool with Magnetic Base 11-0001 Calibration Fitting ® 11-0085 OdaLog Software Kit: OdaStat-G™, IrDA adapter & Manual 11-0008 Plastic Carry Case for Gas 25-0391 Thermo Fisher Scientific Dry Dock (Sensor Conditioning Kit) 10.2 Spare Parts...
  • Page 34: Gas Test Kits

    ® OdaLog Gas Logger User Manual 10.3 Gas Test Kits Gas Type Gas Concentration All gas Test Kits Include Test Gas, Regulator, Calibration Fitting, Plastic Carry Case & Tubing Hydrogen sulphide 50ppm Hydrogen sulphide 25ppm Carbon monoxide 500ppm Sulphur dioxide 20ppm Nitric oxide 50ppm...
  • Page 35: Warranty

    ® OdaLog Gas Logger User Manual 11 WARRANTY 11.1 Warranty Thermo Fisher Scientific Australia Pty. Ltd. warrants, for a period of 12 months from the date of despatch from Thermo Fisher Scientific’s premises, that the OdaLog ® Gas Logger instrument including sensors and all internal components, supplied to you will be free from defects arising in the manufacture.
  • Page 36: Appendix A - Glossary Of Terms And Abbreviations

    ® OdaLog Gas Logger User Manual 13 Appendix A - GLOSSARY OF TERMS AND ABBREVIATIONS ® The following abbreviations and terminology are used in this manual and / or appear on the OdaLog display screen, label and software. -OF- / -OS- / OFL Over display range –...
  • Page 37: Appendix B - Certificates

    ® OdaLog Gas Logger User Manual 14 Appendix B - CERTIFICATES P/N: 10-1000 Revision: F, Issued: October 2019 Page: 34...
  • Page 38 ® OdaLog Gas Logger User Manual P/N: 10-1000 Revision: F, Issued: October 2019 Page: 35...