Digital Apgar Timer Service Sheet
Hospital/Organisation:
Location (department, room no.):
Engineer (print):
Service Date:
Action
OK
Further action required
Serial Number:
Batteries replaced
Comments:
Check for overall signs of damage
Comments:
Check START button enters Counting Mode
Comments:
Check FREEZE button enters Freeze Mode
Comments:
Check FREEZE button resumes Counting Mode
Comments:
Re-enter Freeze Mode, check START button resumes Counting Mode
Comments:
Check RESET button enters Stand-by Mode
Comments:
Start a stopwatch whilst simultaneously pressing START
Check single beep audible indication at 1 minute
Comments:
Check two beeps audible indication at 5 minutes
Comments
Check three beeps audible indication at 10 minutes
Comments:
Check count is synchronized with the stopwatch at 10 minutes, ± 1 sec
Comments:
Check mounting clamp and bracket are securely fastened
Comments:
Clean unit as detailed in the Operator's Manual
Comments: