Section 2:Xstrahl 100 System Description
100_OPMAN_GB Version 9 ©2017 Xstrahl Ltd. All rights reserved
Inserting the Applicator
To insert the applicator
1. Lift latch from the locked position.
Figure 2-4: Fiting the Applicator
2.
Insert the applicator into the sub-tube assembly (the applicator will
only fit in one direction).
Figure 2-5: cont'd..
3.
Push the applicator fully into the sub-tube assembly. When the
applicator reaches the home position, the latch will spring back
down into its locked position. It is not possible to rotate the
applicator within the sub-tube assembly.
Figure 2-6: cont'd..
4.
Ensure the locking latch is in the home position before exiting the
treatment room.
Figure 2-7: cont'd
Xstrahl 100 Operator Manual
06/04/2018 | 2-5