3 hp automatic euro top post mount doctor’s system with flex arm

Advertisement

Installation

1. Install the chair adapter as per instructions provided with the chair adapter.
2. Place the shipping carton in the desired location. Remove the unit with arm system
3. Insert the unit post into the chair adapter hub. Level, then tighten the set screws to
4. Place the adapter in the unit post and insert the umbilical into the unit post, pulling
5. Cut the safety tie strap and carefully release the arm system.
6. Rotate the flex arm 180
7. Remove the rigid arm end cap and tighten the cap screw to set the stop limits for
8. Install master valve assemblies, with filter and regulator, onto the manual shut-off
9. Feed the supply tubings from the umbilical through the 3/4" hole in the utility center
10. Connect the supply tubings as shown in the Utility Center Template & Guide.
Final
Check air pressure, adjust to 80 psi if necessary.
Adjustments
Check water pressure, adjust to 40 psi if necessary.
Test handpieces for proper function.
Test syringe for proper pressure and function.
Adjust flex arm tension, if necessary.
800.237.2303 ◆ 503.538.8756 ◆ 503.538.2845 Fax ◆ www.beaverstatedental.com
and unit from the shipping carton.
secure the post in place.
it out through the bottom. Slide the arm knuckle into the adapter.
The arm system is spring loaded. Before removing
the safety tie strap, firmly grasp the flex arm and hold
0
do not become crimped or twisted.
the flex arm. Replace the end cap.
valves. (Pressures are preset at the factory: 80 psi air , 40 psi water).
housing. Screw in the spiral tubing to act as a strain relief.
WARNING
in a counter-clockwise direction so that the umbilical tubings
S-6307
Arm Stop Cap Screw
1