Content Product Application Mode ......................1 1.1 Stranger Temperature Measurement – Standalone Version ..........1 1.2 Fixed Personnel Temperature Measurement – Standalone Version ........1 1.3 Online Temperature Measurement ..................1 Operating Instructions ......................2 2.1 Power On ..........................2 2.2 Main Interface ........................
1. Product Application Mode 1.1 Stranger Temperature Measurement – Standalone Version Once power on, WD-C7808 Facial Recognition Thermometer (referred to as terminal) will automatically enters into stranger temperature measurement application mode. When someone walks into the detection zone, the terminal will automatically capture the human face, the temperature will be displayed on the screen and broadcasted by audio.
2. Operating Instructions 2.1 Power On Insert charger in charging port at back side of terminal, the temperature measurement software will be started automatically. It takes about 10 seconds for the boot process, please wait patiently. After power on of terminal, calibration with ambient temperature is required. It is recommended to wait 10 minutes after power on before use terminal for temperature measurement.
Sound control - click this icon the volume control bar will pop up, which can be used to adjust system volume. Server connection status – if the dot in the icon is red, it indicates that server is not connected; otherwise, if the dot in the icon is green, it indicates that server is connected.
Page 6
Enter password "6116" (default password) in the input box, click "OK" button, it should enter the system management interface, which is displayed as below. Click the icon in the upper right corner to quickly enter the android settings interface for related operations.
After entering system management interface, select the "communication settings" as shown in the following figure. URL address - management platform address, the default is Watchdata management platform address: 101.201.181.195:12080 Reg Code - the organization code is used to identify the company to which the device belongs. The default value is wd-test.
Device name - user-defined device name, if user need to have multiple devices connect to the server, device name is used for distinction. Device gate - the default value is "Entrance", it can be set according to the actual installation environment as Entrance, Exit or Entrance /Exit.
Calibrate the temperature value by sliding up and down with finger for temperature compensation. 2.3.5 List of People After entering system management interface, select the "List of people" to view the list of people in the terminal, including white list, guest and blacklist.
2.3.6 Historical Record Review After entering system management interface, select the "History" to view the history recognition data in the terminal, which is shown as follows. The history comparison record is displayed on the interface. Double-click to view the details of the record.
2.3.8 Networking Settings Swipe up on the lower side of the screen to display the navigation bar and enter the android desktop. Use android desktop Settings button to go to WIFI Settings or Ethernet Settings. This setting is the same as for a universal android device.
3. Data Import/Export with USB Device For WD-C7808 to work at Fixed Personnel Temperature Measurement – Standalone Version as mentioned in section 1.2, data of pre-set personnel should be imported and exported by USB port, as illustrated as follow. 3.1 Data Import To support the function of Fixed Personnel Temperature Measurement –...
Column Description Necessity Remark Serial Number Enter the serial number in column A of the import file (add up from 1) Personnel Name Enter the name of the person in column B of the import file Employee ID Enter the Employee ID in column D of the import file Reserved for use Personnel Type...
Page 14
Select "Yes" to continue saving. Then the import file shall be transformed into UTF-8 encoded format by following operations. Open the import file with Notepad. Open the Import File with Notepad.
Page 15
After that, please save the import file as UTF-8 encoded file.
Copy the .cvs import file and photos to the same directory of USB flash disk. Then insert USB flash disk to WD-C7808 to import data. After inserting USB flash disk into WD-C7808, the following prompts are displayed in order on the touching screen of terminal.
3.2.2 Export Historical Temperature Measurement Data To export historical temperature measurement data, first please enter the facial recognition and temperature measurement application, click the time button five times in a row, and enter "6116" on the verification password interface. Click History in System Management interface, and then click the "EXPORT" button in History interface to complete the export process.
Page 19
The saving path of export file is as follow in the USB flash disk. Here is an example of file exported from WD-C7808, which can be opened by EXCEL.
Figure 3.1 Desktop, On the Wall and On the Gate With additional base in the package, terminal of WD-C7808 can be placed on table, desk and similar positions. The additional base is suitable to place a contactless reader inside.
Page 21
Step 3: As the hexagon screws at the bottom of the host can be seen from the through hole on the back of the base, use the hexagon tool with the equipment to screw out the screws 3 to 4 times counterclockwise to fix the base and the host in front and back positions Hexagon Screw Position...