LS-320 0482 Infrared Thermometer Instruction Manual Genesis Biotech Pty Ltd (LifeSmart) Manual Version: 1.0 Date of Issue: 2019/02...
Page 2
Introduction Thank you for choosing the LifeSmart Dual mode Digital Infrared Thermometer. Thanks to unique technology, this Dual mode infrared thermometer (LS- 320) can give consistently accurate results without interference from ambient temperatures. The dual mode infrared thermometer can scan either the forehead or the ear within one second.
In case of any damage or operation problem, please contact the dealer or contact LifeSmart directly. You will need the following information when making your claim: device model, serial number, purchase date, and your contact information and address.
Page 5
◆ Do not subject the thermometer to vibration or impact. ◆ The normal body temperature varies from person to person. Tracking a person's body temperature will help determine whether they have fever. ◆ Do not take body temperature readings within 20 minutes after you do physical exercises or get excited.
Warning Warning ⚫ Do not force the temperature probe of the thermometer into an ear canal. Otherwise, the ear canal may get injured. ⚫ Keep the thermometer out of the reach of children. ⚫ The result may be inaccurate if you use the overdue thermometer. ⚫...
Symbols Symbol Description Type BF applied part. Attention must be paid. The action is prohibited. Information about the manufacturer. Date of manufacture. Consult the instructions for use. This product complies with the MDD93/42/EEC requirements. Waste electrical materials should be sent to a dedicated collection point for recycling.
Page 8
from different sites should not be compared directly. To determine if an individual is experiencing an elevated body temperature and/or having a fever, it is critical to know the individual’s normal body temperature when he/she is well. Take multiple readings to obtain the normal body temperature range and note the specific body site measured, for example: forehead or eardrum temperature.
Intended use The LS-320 Dual Mode Digital Infrared Thermometer is intended for the measurement of human body temperatures. The forehead mode is indicated for use by people of all ages and the eardrum mode is indicated for use by people above three months old.
Features Good safety ● Passive infrared receiving technology. Easy operation ● Ergonomic design ● One button measurement Quick measurement ● 1-second measurement High accuracy ● Advanced infrared temperature sensor, with high sensitivity ● High accuracy with automatic temperature calibration Diverse functions ●...
Product Structure (1) LCD display screen (2) Power button / Measure button (3) Unit switch button (˚C / ˚F) (4) Mode button (Adult mode / Child mode/ Object) (5) Probe (take off the cover when measuring the ear temperature) (6) Probe cover (Automatic switch among object, forehead and ear modes, by removing the probe cover to measure ear temperature and placing back the probe cover for object and forehead temperature measurement.) (7) Memory button / Sound switch...
(8) Battery cove Display Description 1. Object temperature mode 2. Forehead temperature mode 3. Child mode 4. Ear temperature mode 5. Mute / un-mute 6. Temperature unit (˚F/˚C) 7. Low battery 8. Memory recall 9.Temperature value Sounds and Backlight Color Instructions Range Sounds Backlight...
indicates that your body temperature is a little high. You may have a fever. Please consult your doctor if you are not sure. Display and Operating Instructions Operating Instructions Sound and Screen Display Backlight Color Displayed State Measuring Ear temperature (Adult / Child) Take off the probe cover, press and release the Power button for 1 second to power on the...
Page 14
Operating Instructions Sound and Screen Display Backlight Color Displayed State Put the cover on the probe, press and release the Power button for 1 second to power on the thermometer The “Head” symbol is displayed on the screen. See the table in the Switch to the adult or child “Sounds and measurement mode by pressing the...
Page 15
Operating Instructions Sound and Screen Display Backlight Color Displayed State Put the cover on the probe, switch to the object measurement mode by pressing the Mode button.The See the table in the thermometer enters the Object “Sounds and mode. The “House” symbol backlight color displayed on the screen.
Page 16
Operating Instructions Sound and Screen Display Backlight Color Displayed State In a power-on state, press the Memory button enter the memory mode. When the Memory button is released, 01 will be shown, followed by the recorded reading. Press the Memory button again for the next recorded data.
Page 17
Operating Instructions Sound and Screen Display Backlight Color Displayed State The display is as shown, when there When the power is is no more data to be checked while turned back on, recalling memories. Remove 2 dry there are a long batteries and after 10 seconds re- beep and a green install the power to clear all...
Page 18
Operating Instructions Sound and Screen Display Backlight Color Displayed State When putting on the probe cover, press the Mode button to switch between object and forehead Silent, the modes. Ear mode (adult/child) will backlight is green. be activated by removing the probe cover.
Operating Instructions Sound and Screen Display Backlight Color Displayed State Error information & low battery The ambient temperature is higher A long beep,the than 40.0ºC (104.0ºF) or lower than backlight is red. 10.0ºC (50.0ºF). An error occurs when data is being read from or written to the memory, A long beep,the or the temperature correction is not...
Page 20
ear temperature. 3. Press the Power button to power on the thermometer. 4. The “Ear” symbol is displayed on the screen. 5. Switch between “adult” and “child” modes by pressing Mode button. symbol shows up in “child” mode. 6. Insert the temperature probe into the ear canal. 7.
Be careful when taking temperature on a child, whose ear canal is small. Measuring Forehead Temperature Put the cover on the probe of thermometer. Press the Power button to power on the thermometer. the “ ” symbol is displayed on the screen. 3.Switch between “adult”...
Page 22
After a measurement (1) After each measurement, you can enter the recall mode and query earlier temperature readings. For more details, see "Recall 20 memories" in the preceding table. (2) After each measurement, clean the temperature probe with a soft cloth, and put the thermometer in a dry and well-ventilated place.
(6) No intense emotion or strenuous exercises before measuring; (7) After measuring the data once, you should wait for the backlight to turn off before measuring again. Replacing Batteries 1、Slide the battery cover off along the marked direction and take it off. 2、Insert the two AAA batteries into the compartment according to the stated polarities.
Cleaning and Disinfection Cleaning Recommended detergents: * Medical detergents; * Home use mild detergents; Cleaning steps: (1) Take the batteries out before cleaning. (2) Clean the temperature probe with a soft cloth. Clean the lens of the temperature probe with a cotton swab. (3) Wipe the thermometer body with a slightly damp soft cloth.
Wet the clean soft cloth with a small quantity of disinfectant, wipe the thermometer and quickly dry it. Disinfect the thermometer body and the area around the temperature probe with a cloth slightly moistened with 75% medical alcohol. Do not use hot steam or ultraviolet radiation for disinfection. Otherwise, the thermometer may be damaged or quickly aged.
requirements. Check regularly whether safety risks exist. Remove the batteries if the thermometer will not be used for more than two months. Troubleshooting Problem Possible Cause Solution Low battery Change the batteries. Polarities of the Make sure that the The thermometer batteries are reversed.
Specifications Product Name Infrared Thermometer Product Model LS-320 Power Supply Internal power supply Mode Operating Voltage DC 3V Battery Model AAA x 2 Operating Mode Continuous operating Display...
Outer dimensions 155.9*40.2*49.2mm (mm) Weight (g) Thermometer (with batteries): 90g Temperature: 10°C~ 40°C (50°F–104°F) Operating Humidity: 15%–95% RH, non-condensing Environment Atmospheric pressure: 86–106 kPa The infrared thermometer has been tested and conforms to the standard ASTM E1965-98. ASTM laboratory accuracy requirements in the display range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is ±0.4°F (±0.2°C).
Guidance and manufacturer´s declaration – Electromagnetic emission The Infrared Thermometer LS-320 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer LS-320 should assure that it is used in such an environment. Emissions...
Page 30
Guidance and manufacturer´s declaration – Electromagnetic immunity The Infrared Thermometer LS-320 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer LS-320 should assure that it is used in such an environment. Immunity...
Page 31
Guidance and manufacturer´s declaration – Electromagnetic immunity The Infrared Thermometer LS-320 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer LS-320 should ensure that it is used in such an environment. Immunity...
Page 32
according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed transmitters, determined by an Electromagnetic site survey,a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:...
Page 33
EQUIPMENT or SYSTEM -for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING The Infrared Thermometer LS-320 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Infrared Thermometer LS-...
0.01 0.12 0.23 0.38 0.73 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Page 35
Wellkang Ltd Suite B, 29 Harley Street, London W1G 9QR,UK Shenzhen Jumper Medical Equipment Co, Ltd. Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China Fax:+86-755-26852025...
Page 36
Manufactured For: Genesis Biotech Pty Ltd. Level 1, 16 McDougall Street, MILTON QLD 4064 www.mylifesmart.net.au CONTACT US For Australia & New Zealand customers support@mylifesmart.net.au Australian customers 1300 110 155 (AEST - Mon-Fri, 11 am - 4 pm)