Bindicator VRF-1000R Series Quick Start Manual

Variable radio frequency point level sensors remote electronics with opti-sense

Advertisement

Quick Links

Start-up & Operation
The VRF-1000R Remote model will automatically begin operation when
powered up. It should be initially powered after it is installed in the application
and with material below the probe. When initially powered up, the VRF-1000R
Remote model will automatically begin a 3 second calibration operation. If the
unit is powered on the bench prior to installation, or moved from one installation
to another, recalibration is required. Momentarily press the recalibration
pushbutton. The green calibration indication LED will go off, and after 3 seconds
illuminate to indicate calibration is complete. The VRF-1000R Remote model is
shipped from the manufacturer with the automatic calibration/recalibration
feature enabled. If only manual pushbutton calibration/recalibration is desired,
set SW1 DIP switch position 6 to off.
Please see VRF180002 for complete installation guidelines.
Dimensional and electrical drawings of the VRF-1000R models are available in
various formats from Bindicator's website; www.Bindicator.com
From the home page, click on the link Installation Manuals/VRF180002 to obtain
the applicable drawing.
If access to the internet is not available, please contact Bindicator
Care Department at 1-800-778-9242 for assistance in obtaining these drawings.
150 Venture Boulevard
Spartanburg, SC 29306
Phone: (864) 574-8060, Fax: (864) 574-8063
Customer Care: (800) 778-9242
Internet: http://www.bindicator.com
email: sales@bindicator.com
© 2006 All rights reserved.
All data subject to change without notice.
Quick-Start Guide
SAFETY INFORMATION
Before installing the VRF-1000R Remote model, please
read these instructions and familiarize yourself with the
requirements and functions. If any questions or problems
arise during the installation, please contact Bindicator
Applications at 1-800-778-9242.
The VRF-1000R Remote model must only be installed and
operated as described in this operating instruction. Please
note that other action can cause damage for which
Bindicator
installed correctly or used in approved applications,
dangers may arise such as product overflow.
Ensure that all personnel installing, wiring, and calibrating this device are
suitably qualified.
®
Customer
Observe all local and national electrical codes for the wiring of this device.
IMPORTANT!
Either AC or DC power is to be connected to the power input depending on
the model. Consult the nameplate of the unit. Models beginning with
VRF1REA are AC powered units and accept any voltage between 85 VAC
and 265 VAC. Models beginning with VRF1RED are DC powered units and
accept any voltage between 9 VDC and 36 VDC.
IMPORTANT!
An electrical earth ground connection must be made to the housing input
ground screw. See Figure 1 for location. This connection is needed for both
electrical safety and proper operation of the unit. This is required for both
AC and DC units.
External ground wire must be connected to tank wall or top of tank.
VRF 180402 Rev. B 11/06
of Variable Radio Frequency
Remote Electronics with Opti-Sense
®
does not take responsibility. If the model is not
VRF-1000R Series
Point Level Sensors
®
TM

Advertisement

loading

Summary of Contents for Bindicator VRF-1000R Series

  • Page 1 From the home page, click on the link Installation Manuals/VRF180002 to obtain the applicable drawing. Ensure that all personnel installing, wiring, and calibrating this device are suitably qualified. If access to the internet is not available, please contact Bindicator ® Customer Care Department at 1-800-778-9242 for assistance in obtaining these drawings.
  • Page 2 Figure 2. High/Low Level Failsafe The output relay contacts are labeled in the un-powered alarm state. The relay is energized when the VRF-1000R is not alarmed. This status may be opposite that of other Bindicator ® brand units. If the VRF-1000R is replacing an older Model RF unit, connections will need to be opposite.