ALWAYS disengage the torque
tester IMMEDIATELY upon
releasing the torque wrench.
The maximum torque value is
shown in the display.
3.8

Electronic overload notification

An out-of-calibration torque tester can result in the failure of
the screw connections, the torque wrench, or the accessories.
This could cause SERIOUS INJURY or DEATH. (Section 4.1)
For model 8612-012
The electronic torque tester emits a constant beeping sound if
too high a torque (> 13 Nm) is applied.
Immediately release the torque electronic tester. Horizontal
lines appear at the upper and lower edge of the display.
Disconnect the electronic torque tester to turn off the beeping
sound.
NOTICE.
To make sure, that the torsion shaft has not been damaged
as a result of an excessive overload having occurred, you
should have the tester checked (calibrated). For this purpose,
please contact our service: www.gedore.com
4.

Maintenance

4.1

Testing and Calibrating

An out-of-calibration torque tester can result in the failure of
the screw connections, the torque wrench, or the accessories.
This could cause SERIOUS INJURY or DEATH.
An out-of-calibration torque tester may cause the torque
wrench settings to be over- or under- torque, resulting in
SERIOUS INJURY or DEATH.
EN
WARNING
INACCURATE TORQUE SETTINGS
WARNING
INACCURATE TORQUE SETTINGS