The FGD14 is fitted with two relays, which are linked to user definable alarm set points. The FGD14 may be connected to an interface unit that will allow direct connection to a Sensors The unit can be fitted with a range of sensor modules allowing it to be tailored to individual needs.
Avoid mounting the instrument where it may be subjected to sudden transients in ambient temperature (e.g. above a heater/radiator). Power Supply Options The FGD14 is available with one of two power supply options. 2.2.1 Mains Powered A universal power supply is fitted that operates in the range of 85-265Vac, 47-400 Hz.
Status Scientific Controls module Remote keypad connector Note: When a DC Power Supply is fitted within the FGD14, the terminals marked L and N on the above diagram will be identified as +Ve and –Ve. TD06/026 Issue: Change note: 1882...
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 2.3.1 Remote Indicator Connection E N L Relay 2 Relay 1 To gain access to the terminals within the remote indicator assembly, remove the screw located at each corner of the assembly when viewed from below (4 in total).
To exit the calibration mode press MENU. While the instrument is in a calibration mode – any data displayed on the screen will alternate between the cal number and the reading. The following features are available via the ‘FGD14’ menu system: - Cal number Function...
Page 9
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor Set Relay 2 Mode 3.3.13 Set Alarm Level 1 3.3.14 Set Alarm Level 2 3.3.15 Sounder enable 3.3.16 Alarm hysteresis 3.3.17 Note: The MENU numbers have gaps, this is to maintain compatibility with other products that have different features.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor Alarm Test The FGD14 is equipped with an alarm test function. During alarm the relays, Leds and sounder will operate, see section 3.4 In order to gain access to the alarm test switch see section 3.1.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.2 Zero Sensor This is a calibration feature. It allows the instrument to determine the sensor output under zero gas conditions. • Apply target free gas to the sensor inlet and allow enough time for the sensor to respond and all gas to be purged (typically 2 minutes dependent upon flow rate).
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.3 Span Sensor This is a calibration feature. It allows the instrument to determine the sensor output when it is exposed to a know concentration of gas. • Apply a known concentration of Hydrocarbon gas (typically 20% LEL pentane) to the sensor inlet and allow enough time for the sensor to respond.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.5 Set Sensor FSD This option is used to match the sensor’s operating range with the electronics so that over-range conditions can be detected. • Select calibration mode C: 3 (refer to section 3.2) and press ENTER.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.7 Display Firmware Version Menu mode C: 7 This option is not available in this version – it is used for compatibility with other products. 3.3.8 Restore This feature is used to restore all the settings to the factory default settings.
View Engineer/Diagnostics Data This feature is a view-only feature. No configuration changes are possible from within this menu. This information is for the use of Status Scientific Controls. • Select calibration mode C: 9 (refer to section 3.2) and press ENTER.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.11 Cross Reference This option is used to allow the user to calibrate the sensor with a commonly available gas (e.g. methane or propane) but use the unit to detect a different gas (e.g. methanol or acetone etc.).
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.12 Set Relay 1 Mode The unit is fitted with a relay that is operated in conjunction with the alarm level. The user can select if the relay is normally energised, E’ or normally de-energised, ‘d’ when the unit is not in an alarm condition.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.14 Set Alarm Level 1 • Select calibration mode C: 15 (refer to section 3.2) and press ENTER. • Using the UP and DOWN buttons, adjust the displayed reading so that it matches the desired alarm set point.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor 3.3.17 Set Alarm Hysteresis Hysteresis is applied during alarm checking to stop the alarm relays chattering when the gas level is at the alarm set point. The hysteresis level is expressed as a percentage of the alarm setting. For example if the alarm setting is 10% vol. CO the hysteresis setting is 1% then the alarm relay would be activated at 10% gas and de-activate at 9.9% gas.
The GREEN power led on the remote indicator is continuously lit. Alarm Condition If the FGD14 detects gas levels in excess of the preset alarm level, an alarm is raised. This is indicated by: The RED LED on the front panel of the FGD14 flashes and the sounder will be heard.
Page 21
If the instrument detects a fault with the sensor or the configuration data, a fault is raised. This is indicated by: The RED LED on the front panel of the FGD14 flashes and the display flashes F xxx, where xxx is a fault code as follows...
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor Sensor Replacement Status Scientific Controls field service personnel, or a trained engineer should only perform sensor replacement. Consult Status Scientific Controls for further details. Following sensor replacement: - • A full calibration must be performed.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor ROUBLE HOOTING The ‘FGD14’ Infrared detectors have been extensively tested and have been found to provide reliable monitoring of hydrocarbon gas levels. Any problems encountered by customers are usually the result of: • Failure to adhere to recommended calibration procedures.
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor PECIFICATION Hydrocarbon Size : 122 x 142 x 75mm nominal Weight : 0.9Kg Gas Types : Hydrocarbons Operating Voltage : 85-265V, 47-400 Hz 2 user adjustable set point between 10%LEL Alarm set points and 60% LEL...
STATUS SCIENTIFIC CONTROLS FGD14 Environmental Monitor Carbon Dioxide Size : 122 x 142 x 75mm nominal Weight : 0.9Kg Gas Types : Carbon Dioxide Operating Voltage : 85-265V, 47-400 Hz 2 user adjustable set points between 0.3%Vol Alarm set points...
Need help?
Do you have a question about the FGD14 and is the answer not in the manual?
Questions and answers