Operating Parameter List —
Parameter
Meaning
Counting
mode
Decimal places 1 / 2 / 3 / 4 / 5 / 6 (up to 8 with inch display)
Decimal point
Linear error compensation
Compensation – 99 999,9 < P41 < + 99 999,9 [µm/m]
Reference
marks
Reference mark
evaluation
Encoder
Encoder
monitoring
Baud rate
Additional line feeds
Value for
Preset
datum
Reset/Preset
Display after
Message
switch-on
Inhibit PRINT
Mode
function
1)
Determine entry value for P41
Example: Displayed measuring length L a = 620.000 mm
Actual length (determined with, for example, the VM 101
comparator system from HEIDENHAIN) L t = 619.876 mm
Length difference ∆L = L t – L a = –124 µm
Compensation factor k:
k = ∆L / L a = –124 µm / 0.62 m = –200 [µm/m]
continued
Function / Effect
0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - 0
0 - 2 - 4 - 6 - 8 - 0
0 - 5 - 0
1)
One reference mark
Distance-coded with 500 • SP
(SP = signal period)
Distance-coded with 1000 • SP
(e.g. for LS 303 C / LS 603 C)
Distance-coded with 2000 • SP
Distance-coded with 5000 • SP
Evaluation
No evaluation
No monitoring ( Alarm Off )
Contamination
Frequency
Contamination and frequency
110, 150, 300, 600, 1200, 2400, 4800, 9600
0 to 99 ( Linefeed )
Enter numerical value for datum setting with
the ENT key
No zero reset/preset with CL/ENT
Zero reset with CL ( Set Zero ),
no preset with ENT
Zero reset with CL and
preset with ENT to value in P79
message displayed
message not displayed
PRINT inhibited
PRINT not inhibited
Setting