TESTSETUP
SETTING UP A MULTIPLE STEP TEST IN PROGRAM MODE:
PROGRAM Mode will run your test routine according to the parameters that have
been entered within the Test Parameters screen. In most cases there will be a test
time associated with PROGRAM Mode unless the time has been selected to ∞ or the
cycle mode is ON.
Turn the POWER switch to ON. From the Set screen, press the "<more>" soft key
and in the next screen press the "System" soft key.
In the System Parameters menu, use the up and down arrow soft keys to highlight the
AUTO RUN parameter which changes the test mode. If the AUTO RUN parameter is
not set to PROGRAM, you will need to change it by pressing the "Edit" soft key.
In the next screen, the AUTO RUN parameter will be highlighted. Press the
"Change" soft key to toggle the highlighted parameter to PROGRAM. Press the
"Enter" soft key to save changes and then press the "Exit" soft key two times to return
to the Set screen. The display will be as follows: (See Figure 3)
(Figure 3. Set Screen)
From the Set screen press the "Step" soft key. In the next screen, press the "Edit" soft
key and use the up or down arrow soft keys to highlight the "Connect" parameter.
Now, press the "Edit" soft key and in the next screen press the "Change" soft key to
toggle the connect mode to ON. Press the "Enter" soft key to save changes made
and then press the "Exit" soft key two times to return to the Set screen.
Connecting a second step:
From the Set screen, press the "Step" soft key. Now, press the "Step +" soft key one
time and that will activate step 2. In the next screen, press the "Edit" soft key and
use the up or down arrow soft keys to highlight the "Connect" parameter. Press
the "Edit" soft key and in the next screen press the "Change" soft key to toggle the
connect mode to ON. Press the "Enter" soft key to save changes made.
Press the "Exit" soft key two times to return to the Set screen. You have now set up
a 2-step test. Note: For each additional step you want to add, you need to turn the
"Connect" parameter ON. There are total 9 step locations.
To initialize the test, press the Test/Reset key and the LED for the key will illuminate.
The text "Set" on the Set screen will turn to Dwell. In addition, your soft keys will
change to include: Meter, Cycle, Keypad, and Trig.