Download Print this page

Test Setup - Cal Power APT 7000 Series Quick Start Manual

Automated ac power sources

Advertisement

TESTSETUP
SETTING UP A TEST IN MANUAL MODE:
When the AUTO RUN parameter in the System Parameters menu is set to MANUAL
Mode and you are in the set screen the display will be as follows: (See Figure 1)
(Figure 1. Set Screen)
To initialize the test, press the Test/Reset key and the LED for the key will illuminate.
The text "Set" on the set screen will turn to Dwell. In addition, your soft keys will
change to include Meter, AUTO, Keypad, and Trig (see figure 2). In MANUAL
Mode, output will run continuously until the Test/Reset key is pressed again or there
is a failure condition.
(Figure 2. Set Screen)
In MANUAL Mode, the Rotary Knob on the front panel also becomes active. It can
be used to adjust voltage or frequency. When the instrument is in an idle state, it
can also be used to edit the Hi-Lmt. To adjust the frequency with the Rotary Knob
the meter selection must be on F. To adjust the voltage with the Rotary Knob the
meter can be on any selection except F. To adjust the Hi-Lmt with the Rotary Knob
the meter selection must be on Hi-Lmt. To adjust the output, rotate the knob either
clockwise or counterclockwise. Clockwise rotations will increase the output, whereas
counterclockwise rotations will decrease the output. If LOCK is set to ON, the Rotary
Knob is disabled.
4.

Advertisement

loading

This manual is also suitable for:

Apt 7004Apt 7008Apt 7016Apt 7040