S C A L E I N D I C A T O R G5102
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G5102 and is the answer not in the manual?

Questions and answers

Summary of Contents for GINEERS G5102

  • Page 1 S C A L E I N D I C A T O R G5102 USER MANUAL...
  • Page 2: Table Of Contents

    IV. Serial interface RS-232 Physical connection Data protocol V. Advanced options VI. Setting dosing levels in normal weighing mode VII. Troubleshooting VIII. Safety measures when working with G5102 IX. Contacts Scale indicator G5102, v.1.1 – USER MANUAL 2 / 27...
  • Page 3 Fig.4. Selection of particular menu Fig.5. Algorithm for changing parameter value in menu „PArAS” Fig.6. Algorithm for changing parameter value in menu „Adc” Fig.7. Real weight calibration of G5102 scale Table 1. Characteristics of weigh unit G5102 Scale indicator G5102, v.1.1 – USER MANUAL...
  • Page 4: Used Terminology And Descriptions

    - Full scale – nominal capacity of build weigh scale Descriptions - keyboard button of G5102 unit - it is necessary to press showed button - Attention, important notice - Note Scale indicator G5102, v.1.1 – USER MANUAL 4 / 27...
  • Page 5: General Description

    General look of G5102 is shown on Fig. 1: Fig. 1. Base look of G5102 unit On the picture can be seen most important parts and details of G5102: 1. Display – 5-digit seven-segment display with digit height of 25.4mm. Display lights in super-bright red (λ= 660nm).
  • Page 6 10. *Button for entering in service mode – accessible only if front ring is removed G5102 fulfils requirements of precision class III for measuring weight, but can be used also for more precise measurements. Basic sub-devices and functions (according to EN45501:2001 terminology) are: initial zero setting device;...
  • Page 7: Technical And Metrological Parameters

    -10 °C ÷ +40 °C Power supply, V 220V AC +10%/-15% Power consumed, W < 1,5 W Interfaces RS-232C serial Precision class Dimensions, LxWxH, mm 170x190x70 Protection class, IP IP33 Own weight, kg 1.1kg Scale indicator G5102, v.1.1 – USER MANUAL 7 / 27...
  • Page 8 220V AC Fractal error part pi: Maximum length of connecting cable between Indicator and junction box: 5m, cable LIYCY 4x0,5mm Serial interface: RS-232 Serial interface maximum cable length: 15m / 3x0,5mm Scale indicator G5102, v.1.1 – USER MANUAL 8 / 27...
  • Page 9: Connections

    Every G5102 comes with power supply cable for 220V AC, suited for indicator. From the one side this cable should be plugged in G5102 (connector 6 from I. General description), from the other – in power supply network. It can not be plugged wrong.
  • Page 10: Connecting Serial Interface Rs-232

    This interface can be used for sending weigh values to external cash register or personal computer. If working with PC special software from Gineers can be used or we can supply data protocol to who is interested in reading weight from G5102 with its own software.
  • Page 11 For a wall mounting 4 mounting holes are provided on the back of G5102 indicator. They are placed on 4 corners of the device and are visible when front ring is removed. Through these holes device can be mounted with the help of screws to the wall (not included in the package).
  • Page 12: Functions And Parameters

    Pressing button will change signs on the display – for 1 second use will see „code”, after that G5102 will wait for the user to enter valid code for entering service menu: Fig. 3. Entering service code Scale indicator G5102, v.1.1 – USER MANUAL...
  • Page 13 - used to exit from current menu, sub-menu or parameter change (“Esc”). Generally this is always a step back in menu hierarchy. G5102 is offered with two types of firmware – basic and extended. Here we will make full description of basic type and brief show of extended type.
  • Page 14: Menu „Paras

    Indication that user is in parameter value change mode are LEDs for “Tare” and “Zero” position – they both are lit. Graphically process of setting parameters in menu „PArAS” is showed on Fig. 5: Scale indicator G5102, v.1.1 – USER MANUAL 14 / 27...
  • Page 15: Menu „Adc

    To exit from parameter value change, respectively menu „Adc” user should press button If user wants to change parameter value, he can do that with buttons while he is in mode of changing parameter. Parameter value is automatically Scale indicator G5102, v.1.1 – USER MANUAL 15 / 27...
  • Page 16: Menu „Cal

    Weight calibration can be done with any known mass – what is put on scale platform is entered during the process by G5102 keyboard. Calibration is carried in two Scale indicator G5102, v.1.1 – USER MANUAL...
  • Page 17: Example Of Calibration With Real Weight

    When the calibrating step is step 2 - „Cal n”, after confirm entering user will see “ 0” on display. G5102 now expects user to enter real weight that is put on load platform. Enter correct weight with keyboard buttons (if error in entered number –...
  • Page 18 “0”, LED „Tare” will light up to show user he is in sub-menu „Cal n”. Now G5102 expects user to enter calibration weight. This in fact should be the load that is intended to calibrate scale with. With buttons 0..9 from the keyboard user must enter correct weight.
  • Page 19: Factory Default Values

    Fig.7. Real weight calibration of G5102 scale Note: Fine correction does not work in zero position, because there is no point adding or subtracting from zero (no way to calculate final correction) Scale indicator G5102, v.1.1 – USER MANUAL 19 / 27...
  • Page 20: Menu „Other

    It should not be used often – only when the person setting the scale is in “dead end” and is wondering what to do with G5102. Confirmation of erasing memory is on two levels - first button should be pressed to enter in this function.
  • Page 21: Serial Interface Rs-232

    Start bit: 1 Stop bit: 1 Data bits: 8 Parity check: None When a PC is connected to G5102, serial data transmission can occur in two modes: Weight data is transmitted always after a command is received from PC/external device;...
  • Page 22: Data Protocol

    XX are three bytes current weight in grams, in hexadecimal format (MSB first), CRC is control checksum – ‘W’+Weight. If button is pressed scale sends data to external device in the same way as it responds to a command. Scale indicator G5102, v.1.1 – USER MANUAL 22 / 27...
  • Page 23: Advanced Options

    0 – 24 Correction for latitude 0 - 12 Counting function On/Off Measure in percents function Perc On/Off Avoid zero before sending data On/Off serial Working with label printer Pr-LP On/Off Scale indicator G5102, v.1.1 – USER MANUAL 23 / 27...
  • Page 24: Setting Dosing Levels In Normal Weighing Mode

    . To confirm new value press button . If this parameter value is valid, a sign „Saved” will be written for 1 second, after that G5102 will go back one step in menu hierarchy. Exit without saving/changing parameter is made by pressing button „Print”.
  • Page 25: Troubleshooting

    If the scale doesn’t light at all when powered check the following: Check if power supply from the grid is present; Check if power cable is placed good in G5102 plug Check the fuse. If it is blown – replaced it, parameters are (0.5 ÷ 1.25А)/220V Scale indicator G5102, v.1.1 –...
  • Page 26: Safety Measures When Working With G5102

    However, buttons should be pressed only by hand and with sharp objects or instruments If G5102 should be cleaned outside be sure first to remove power supply cable! Cleaning is made with soft towel, (not wet). Do not use detergents with benzyl or acetone components Scale indicator G5102, v.1.1 –...
  • Page 27: Contacts

    +359 2 975 81 05 office@gineers.com All principal schemes, board layouts, firmware and software are developed by “Gineers” Ltd. Changes are possible only from Gineers, and are negotiable. If you have any questions, suggestions or problems, please contact us at support@gineers.com...