NOTICE
Install a separator (e.g. Woulff bottle) infront of the intake moun-
ting to protect against the ingress of liquid.
In the case of induced solvent vapours, a vapour condenser (acces-
sory of pump) installed downstream of the pump helps to conden-
se the vapours and prevents release into the atmosphere.
Liquid in the pump chambers deteriorates the pump characteris-
tics.
Attach the IKA
®
VC 10 vacuum controller to a stand (d=16 mm)
close to the pump.
Secure the controller with the stand locking screw (N).
To do so, use a stand with diameter 16 mm, for example, IKA
®
stand VC 10.400, IKA
VSS 1 vacuum safety set, the stand on
®
IKA
MVP 10 digital vacuum pump or attach the controller to
®
the IKA
RV 8 rotary evaporator, or at IKA
reactor.
Fig. 6
VC 10 at stand
VC 10.400
Observe the general information,
always
connect
the
(load/vacuum vessel/glass cooler)
at their highest points to the suc-
tion line. This will prevent the risk
of fluid entering into the vacuum
controller or pump.
®
LR 1000 laboratory
VC 10 to VSS 1 with MVP 10 basic

Setting up

recipients
®
Fig. 7
VC 10 to MVP 10 digital
d=16 mm
Stand locking screw (N)
Fig. 8
VC 10 to RV 8
Fig. 5
Fig. 9
33